头部旋转降低i-gel与LMA Supreme口咽部漏气压
这项随机试验在肌松全麻患者中比较i-gel与LMA Supreme在头部0°、30°、60°旋转时的口咽部漏气压,发现两种装置漏气压均随旋转角度增加而显著下降,二者间无显著差异。
要点
- 头部旋转30°和60°均显著降低i-gel与LMA Supreme的口咽部漏气压
- 两种装置在三个头位下的口咽部漏气压无显著差异
- 头部旋转时需关注通气效率,尤其i-gel在0°与60°间呼气潮气量有差异
核心结论
- 头部旋转至30°和60°会降低i-gel与LMA Supreme的口咽部漏气压
- 两种装置在三个头位下的口咽部漏气压无差异
研究设计与样本
- 麻醉诱导并肌松后按计算机随机置入装置,主要结局为0°、30°、60°头位下的口咽部漏气压
主要发现
- i-gel组口咽部漏气压随头位旋转显著下降:0° vs 30°均值差3.5(95%CI 2.2-4.8),P<.001;30° vs 60°均值差2.0(95%CI 0.6-3.5),P=.002;0° vs 60°均值差5.5(95%CI 3.3-7.8),P<.001
- LMA Supreme组口咽部漏气压随头位旋转显著下降:0° vs 30°均值差4.1(95%CI 2.6-5.5),P<.001;30° vs 60°均值差2.4(95%CI 1.1-3.7),P<.001;0° vs 60°均值差6.5(95%CI 5.1-8.0),P<.001
- i-gel组0°与60°间呼气潮气量和通气评分有统计学差异,LMA Supreme组30°与60°间通气评分有统计学差异
- 两种装置在所有结局指标上无统计学显著差异,不良事件发生率无显著差异
关键数值与推荐
- 主要结局为0°、30°、60°头位旋转时的口咽部漏气压
- 机械通气采用容量控制模式,潮气量10 mL/kg(理想体重)
临床怎么用
- 使用i-gel或LMA Supreme时,头部旋转30°和60°会降低口咽部漏气压,需注意通气效率可能下降
- 两种装置在头部旋转时的口咽部漏气压表现相似,选择可依据其他临床因素
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Second-generation supraglottic airway (SGA) devices are useful for airway management during positive pressure ventilation in general anesthesia and emergency medicine. In some clinical settings, such as the anesthetic management of awake craniotomy, SGAs are used in the head-rotated position, which is required for exposure of the surgical field, although this position sometimes worsens the efficiency of mechanical ventilation with SGAs. In this study, we investigated and compared the influence of head rotation on oropharyngeal leak pressures (OPLP) of the i-gel and LMA® Supreme™, which are second-generation SGA devices. METHODS: Patients who underwent elective surgery under general anesthesia were enrolled in this study and randomly divided into i-gel or LMA Supreme groups. After induction of anesthesia with muscle relaxation, the i-gel or LMA Supreme was inserted according to computerized randomization. The primary outcome was the OPLP at 0°, 30°, and 60° head rotation. The secondary outcomes were the maximum airway pressure and expiratory tidal volume when patients were mechanically ventilated using a volume-controlled ventilation mode with a tidal volume of 10 mL/kg (ideal body weight), ventilation score, and fiber-optic views of vocal cords. RESULTS: Thirty-four and 36 participants were included in the i-gel and LMA Supreme groups, respectively. The OPLPs of the i-gel and LMA Supreme significantly decreased as the head rotation angle increased (mean difference [95% confidence interval], P value: i-gel; 0° vs 30°: 3.5 [2.2-4.8], P < .001; 30° vs 60°: 2.0 [0.6-3.5], P = .002; 0° vs 60°: 5.5 [3.3-7.8], P < .001, LMA Supreme; 0° vs 30°: 4.1 [2.6-5.5], P < .001; 30° vs 60°: 2.4 [1.1-3.7], P < .001; 0° vs 60°: 6.5 [5.1-8.0], P < .001). There were statistically significant differences in expiratory tidal volume and ventilation score between 0° and 60° in the i-gel group and in ventilation score between 30° and 60° in the LMA Supreme group. There was no statistically significant difference between the 2 devices in all outcome measures. The incidences of adverse events, such as hoarseness or sore throat, were not significantly different between i-gel and LMA Supreme. CONCLUSIONS: Head rotation to 30° and 60° reduces OPLP with both i-gel and LMA Supreme. There is no difference in OPLP between i-gel and LMA Supreme in the 3 head rotation positions.