文献速递区域麻醉与超声

超声激活针尖追踪针在软防腐Thiel尸体坐骨神经阻滞中的表现

The effect of an ultrasound-activated needle tip tracker needle on the performance of sciatic nerve block on a soft embalmed Thiel cadaver.
PubMed 文献 · Anaesthesia · 2020-08-14

本研究在软防腐Thiel尸体上模拟坐骨神经阻滞,比较医学生和麻醉医师使用针尖追踪针开启与关闭时的操作表现。追踪针通过压电元件在超声束中位置变化显示针尖,主要指标为成功步骤数和错误数。结果显示,开启追踪后10名(25%)受试者步骤改善、6名(15%)错误减少,针尖识别相关步骤和错误的风险比显著。

要点

  • 针尖追踪技术使25%受试者步骤改善、15%错误减少。
  • 注射前识别针尖和进针前识别针尖的OR分别为2.12和1.80。
  • 未识别针尖即注射和丢失后未快速恢复针尖位置的OR分别为2.40和2.03。

核心结论

  • 针尖追踪技术在一部分受试者中改善了坐骨神经阻滞模拟操作表现,使10名(25%)受试者步骤改善、6名(15%)错误减少。

研究设计与样本

  • 在软防腐Thiel尸体上进行模拟坐骨神经阻滞的前瞻性研究,受试者为医学生和麻醉医师,采用自身对照设计,追踪针开启与关闭顺序随机等量分配。
  • 主要结局指标为每次阻滞中成功完成的步骤数和所犯错误数,视频由培训专家使用经验证的核查表评分。

主要发现

  • 开启追踪针后,10名(25%)受试者步骤改善,6名(15%)受试者错误减少。
  • 注射前识别针尖的OR为2.12(95%CI 1.61-2.80;p < 0.001)。
  • 进针前识别针尖的OR为1.80(95%CI 1.36-2.39;p < 0.001)。
  • 注射前未识别针尖的OR为2.40(95%CI 1.78-3.24;p < 0.001)。
  • 丢失针尖可见性后未快速恢复针尖位置的OR为2.03(95%CI 1.5-2.75;p < 0.001)。

临床怎么用

  • 在超声引导坐骨神经阻滞模拟训练中,可考虑使用针尖追踪针辅助识别针尖位置,以减少注射前未识别针尖等关键错误。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

In this study, we measured the performance of medical students and anaesthetists using a new tracker needle during simulated sciatic nerve block on soft embalmed cadavers. The tracker needle incorporates a piezo element near its tip that generates an electrical signal in response to insonation. A circle, superimposed on the ultrasound image surrounding the needle tip, changes size and colour according to the position of the piezo element within the ultrasound beam. Our primary objective was to compare sciatic block performance with the tracker switched on and off. Our secondary objectives were to record psychometrics, procedure efficiency, participant self-regulation and focused attention using eye-tracking technology. Our primary outcome measures were the number of steps successfully performed and the number of errors committed during each block. Videos were scored by trained experts using validated checklists. Sequential tracker activation and deactivation was randomised equally within subjects. With needle activation, steps improved in 10 (25%) subjects and errors reduced in six (15%) subjects. The most important steps were: needle tip identification before injection, OR (95%CI) 2.12 (1.61-2.80; p < 0.001); and needle tip identification before advance of the needle, 1.80 (1.36-2.39; p < 0.001). The most important errors were: failure to identify the needle tip before injection, 2.40 (1.78-3.24; p < 0.001); and failure to quickly regain needle tip position when tip visibility was lost, 2.03 (1.5-2.75; p < 0.001). In conclusion, needle-tracking technology improved performance in a quarter of subjects.

来源:Anaesthesia,查看原文。版权归原作者所有。

引用

  1. McKendrick M, Sadler A, Taylor A, et al. The effect of an ultrasound-activated needle tip tracker needle on the performance of sciatic nerve block on a soft embalmed Thiel cadaver. Anaesthesia. 2020. PMID: 32797700. doi:10.1111/anae.15211

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。