缝合式 versus 针穿式导管用于连续腘窝坐骨神经阻滞:一项随机临床试验
该随机临床试验比较了新型缝合式神经周围导管与传统针穿式导管在足踝术后连续腘窝坐骨神经阻滞中的镇痛效果。结果显示,缝合式导管组术后前两天平均疼痛评分不劣于传统导管组,且导管完全脱位率为0%,而传统导管组有9%的导管尖端在术后第3天计划拔除前已脱出皮肤。
要点
- 缝合式导管镇痛效果不劣于传统针穿式导管。
- 缝合式导管未发生完全脱位,传统导管脱位率为9%。
- 新型导管设计可作为传统设计的可行替代方案。
核心结论
- 缝合式神经周围导管在足踝术后连续腘窝坐骨神经阻滞中提供的镇痛效果不劣于传统针穿式导管。
主要发现
- 术后前两天平均疼痛评分:缝合式导管组为2.7 ± 2.4,针穿式导管组为3.4 ± 2.4,差异有统计学非劣效性(95% CI, -1.9 to 0.6; P < 0.001)。
- 缝合式导管完全脱位率为0%,传统导管有3例(9%)在术后第3天计划拔除前导管尖端脱出皮肤(P = 0.239)。
关键数值与推荐
- 术后3天输注0.2%罗哌卡因,基础速率6 ml/h,单次追加4 ml,锁定时间30分钟。
- 主要结局为术后前两天平均疼痛评分(数字评定量表)。
临床怎么用
- 对于足踝术后需要连续腘窝坐骨神经阻滞的患者,可考虑使用缝合式导管作为传统针穿式导管的替代。
局限性
- 样本量较小(每组35例),导管脱位率的比较未达到统计学显著性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The basic perineural catheter design has changed minimally since inception, with the catheter introduced through or over a straight needle. The U.S. Food and Drug Administration recently cleared a novel perineural catheter design comprising a catheter attached to the back of a suture-shaped needle that is inserted, advanced along the arc of its curvature pulling the catheter past the target nerve, and then exited through the skin in a second location. The authors hypothesized that analgesia would be noninferior using the new versus traditional catheter design in the first two days after painful foot/ankle surgery with a primary outcome of average pain measured with the Numeric Rating Scale. METHODS: Subjects undergoing painful foot or ankle surgery with a continuous supraparaneural popliteal-sciatic nerve block 5 cm proximal to the bifurcation were randomized to either a suture-type or through-the-needle catheter and subsequent 3-day 0.2% ropivacaine infusion (basal 6 ml/h, bolus 4 ml, lockout 30 min). Subjects received daily follow-up for the first four days after surgery, including assessment for evidence of malfunction or dislodgement of the catheters. RESULTS: During the first two postoperative days the mean ± SD average pain scores were lower in subjects with the suture-catheter (n = 35) compared with the through-the-needle (n = 35) group (2.7 ± 2.4 vs. 3.4 ± 2.4) and found to be statistically noninferior (95% CI, -1.9 to 0.6; P < 0.001). No suture-style catheter was completely dislodged (0%), whereas the tips of three (9%) traditional catheters were found outside of the skin before purposeful removal on postoperative day 3 (P = 0.239). CONCLUSIONS: Suture-type perineural catheters provided noninferior analgesia compared with traditional catheters for continuous popliteal-sciatic blocks after painful foot and ankle surgery. The new catheter design appears to be a viable alternative to traditional designs used for the past seven decades.
引用
- Finneran JJ, Swisher MW, Gabriel RA, et al. Suture-method versus Through-the-needle Catheters for Continuous Popliteal-sciatic Nerve Blocks: A Randomized Clinical Trial. Anesthesiology. 2020. PMID: 31996555. doi:10.1097/ALN.0000000000003145
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。