文献速递气道管理

Tritube与显微喉镜管对喉部手术声门可见性的随机对照试验

Glottic visibility for laryngeal surgery: Tritube vs. microlaryngeal tube: A randomised controlled trial.
PubMed 文献 · Eur J Anaesthesiol · 2019-01-01

这项随机对照试验纳入40例择期喉部手术患者,比较Tritube联合FCV与MLT 6.0联合VCV对喉部结构遮挡程度的影响。结果显示Tritube组遮挡中位数7% vs 22%(P<0.001),手术条件评分无差异,但FCV组呼吸系统顺应性更高、平台压更低。

要点

  • Tritube-FCV较MLT-VCV显著减少喉部结构遮挡(7% vs 22%,P<0.001)。
  • 两组手术条件评分无显著差异(P=0.06),但住院医师亚组认为Tritube更优(P=0.006)。
  • Tritube-FCV的呼吸系统顺应性更高、平台压更低、呼气末肺容积变化更大。

核心结论

  • 喉部手术中Tritube较MLT 6.0改善手术部位可见性,但未改善手术条件。
  • FCV较VCV改善肺通气与呼吸系统顺应性。

研究设计与样本

  • 随机对照试验,在大学医学中心进行。
  • 55例连续患者评估,40例可评估数据,每组20例。

主要发现

  • Tritube组喉部结构遮挡中位数7%[6-9] vs MLT组22%[18-27],P<0.001。
  • 手术条件评分两组相当,P=0.06;住院医师亚组认为Tritube组更优,P=0.006。
  • Tritube-FCV组呼吸系统顺应性61[52-71] vs 46[41-51] ml cmH2O,P<0.001。
  • Tritube-FCV组平台压14[13-15] vs 17[16-18] cmH2O,P<0.001。
  • Tritube-FCV组呼气末肺容积变化681[463-849] vs 414[194-604] ml,P=0.023。

关键数值与推荐

  • 潮气量标准化为7 ml kg预测体重,呼气末正压7 cmH2O。
  • 主要终点为管相关喉部结构遮挡,通过视频喉镜照片软件测量。
  • 次要终点包括手术条件四分评分、呼吸变量及呼气末肺容积变化。

临床怎么用

  • 喉部手术可选用Tritube联合FCV以改善术野可见性和呼吸力学。
  • 手术条件整体相当,但住院医师操作时Tritube可能更有优势。
  • 需注意FCV对肺复张和顺应性的改善效应。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Good visibility is essential for successful laryngeal surgery. A Tritube with outer diameter 4.4 mm, combined with flow-controlled ventilation (FCV), enables ventilation by active expiration with a sealed trachea and may improve laryngeal visibility. OBJECTIVES: We hypothesised that a Tritube with FCV would provide better laryngeal visibility and surgical conditions for laryngeal surgery than a conventional microlaryngeal tube (MLT) with volume-controlled ventilation (VCV). DESIGN: Randomised, controlled trial. SETTING: University Medical Centre. PATIENTS: A total of 55 consecutive patients (>18 years) undergoing elective laryngeal surgery were assessed for participation, providing 40 evaluable data sets with 20 per group. INTERVENTIONS: Random allocation to intubation with Tritube and ventilation with FCV (Tritube-FCV group) or intubation with MLT 6.0 and ventilation with VCV (MLT-VCV) as control. Tidal volumes of 7 ml kg predicted body weight, and positive end-expiratory pressure of 7 cmH2O were standardised between groups. MAIN OUTCOME MEASURES: Primary endpoint was the tube-related concealment of laryngeal structures, measured on videolaryngoscopic photographs by appropriate software. Secondary endpoints were surgical conditions (categorical four-point rating scale), respiratory variables and change of end-expiratory lung volume from atmospheric airway pressure to ventilation with positive end-expiratory pressure. Data are presented as median [IQR]. RESULTS: There was less concealment of laryngeal structures with the Tritube than with the MLT; 7 [6 to 9] vs. 22 [18 to 27] %, (P < 0.001). Surgical conditions were rated comparably (P = 0.06). A subgroup of residents in training perceived surgical conditions to be better with the Tritube compared with the MLT (P = 0.006). Respiratory system compliance with the Tritube was higher at 61 [52 to 71] vs. 46 [41 to 51] ml cmH2O (P < 0.001), plateau pressure was lower at 14 [13 to 15] vs. 17 [16 to 18] cmH2O (P < 0.001), and change of end-expiratory lung volume was higher at 681 [463 to 849] vs. 414 [194 to 604] ml, (P = 0.023) for Tritube-FCV compared with MLT-VCV. CONCLUSION: During laryngeal surgery a Tritube improves visibility of the surgical site but not surgical conditions when compared with a MLT 6.0. FCV improves lung aeration and respiratory system compliance compared with VCV. TRIAL REGISTRY NUMBER: DRKS00013097.

来源:Eur J Anaesthesiol,查看原文。版权归原作者所有。

引用

  1. Schmidt J, Günther F, Weber J, et al. Glottic visibility for laryngeal surgery: Tritube vs. microlaryngeal tube: A randomised controlled trial. Eur J Anaesthesiol. 2019. PMID: 31644514. doi:10.1097/EJA.0000000000001110

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。