文献速递区域麻醉与超声气道管理

带针尖追踪技术的新型区域阻滞针在尸体坐骨神经阻滞新手操作中的初步评估

An initial evaluation of the effect of a novel regional block needle with tip-tracking technology on the novice performance of cadaveric ultrasound-guided sciatic nerve block.
PubMed 文献 · Anaesthesia · 2019-09-10

研究评估一种带压电元件(距针尖2-2.3mm)的追踪针在超声引导坐骨神经阻滞中对新手操作的影响。8名新手在Thiel软固定尸体上各完成20次平面内坐骨神经阻滞,随机分组交替开启或关闭追踪器,共分析160段视频。结果显示追踪针开启时,5个正确步骤改善、1个错误减少,包括进针前识别针尖、针与探头对齐及针尖可见性均更优。

要点

  • 追踪针可改善新手在尸体坐骨神经阻滞中的操作表现。
  • 进针前识别针尖的OR为3.4(95%CI 1.6-7.7,p<0.001)。
  • 针与探头对齐及针尖可见性亦显著改善。

核心结论

  • 使用带追踪技术的穿刺针可改善新手在软固定尸体模型上实施坐骨神经阻滞的操作表现。

研究设计与样本

  • 前瞻性随机对照模拟研究,8名新手麻醉医生在Thiel软固定尸体上各完成20次平面内坐骨神经阻滞,按追踪器开/关交替序列随机分组,共分析160段视频,由6名评分者评估。

主要发现

  • 使用追踪针时,5个正确步骤改善、1个错误减少。
  • 进针前更好识别针尖的OR为3.4(95%CI 1.6-7.7,p<0.001)。
  • 针与探头更好对齐的OR为3.1(95%CI 1.3-8.7,p=0.009)。
  • 针尖更好可见性的OR为3.0(95%CI 1.4-7.3,p=0.005)。

临床怎么用

  • 该追踪针技术有望辅助新手在超声引导神经阻滞中实时确认针尖位置,减少操作错误。
  • 目前证据仅来自尸体模拟环境,临床转化需进一步验证。

局限性

  • 研究样本量小(8名受试者),仅评估新手在尸体模型上的模拟操作,结论外推至临床患者需谨慎。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Visibility of the needle tip is difficult to maintain during ultrasound-guided nerve block. A new needle has been developed that incorporates a piezo element 2-2.3 mm from the tip, activated by ultrasound. The electrical signal manifests as a coloured circle surrounding the needle tip, and allows real-time tracking. We hypothesised that novice regional anaesthetists would perform nerve block better with the tracker turned on rather than off. Our primary objective was to evaluate the new needle by measuring the performance of novice anaesthetists conducting simulated sciatic block on the soft embalmed Thiel cadaver. Training consisted of a lecture, scanning in volunteers and practice on cadavers. Testing entailed scanning the sciatic nerve of a cadaver and conducting 20 in-plane sciatic blocks in the mid-to-upper thigh region. Subjects were randomised equally, in groups of five, according to the sequence: tracker on/off/on/off; or tracker off/on/off/on. Video recordings were assessed by six raters for steps performed correctly and errors committed. Eight subjects were recruited and 160 videos were analysed. Using the tracking needle, five correct steps improved and one error reduced. The benefits included: better identification of the needle tip before advancing the needle, OR (95%CI) 3.4 (1.6-7.7; p < 0.001); better alignment of the needle to the transducer, 3.1 (1.3-8.7; p = 0.009); and better visibility of the needle tip 3.0 (1.4-7.3; p = 0.005). In conclusion, use of the tracker needle improved the sciatic block performance of novices on the soft embalmed cadaver.