患者自控镇痛装置相关错误:某三级医院82,698例患者分析

Device-Related Error in Patient-Controlled Analgesia: Analysis of 82,698 Patients in a Tertiary Hospital.
PubMed 文献 · Anesth Analg · 2019-01-01

该研究分析某三级医院2011至2014年82,698例术后使用PCA患者的装置相关错误,发生率为0.19%,其中电子可编程泵错误发生率最高(0.47%,95%CI 0.36-0.59),63%的错误患者出现不良临床结局但无死亡。

要点

  • PCA装置相关错误总体发生率低于0.2%,但电子可编程泵错误率最高。
  • 装置错误中输注不足最常见,占30.3%。
  • 63%的装置错误患者出现不良结局,从轻微症状到呼吸骤停,无死亡。

核心结论

  • PCA装置相关错误发生率为0.19%,且因输注泵类型不同而存在显著差异。
  • 63%的装置错误患者出现不良临床结局,但无死亡。

研究设计与样本

  • 共纳入82,698例使用PCA(静脉、硬膜外、神经阻滞)的手术患者,涉及4种输注泵。

主要发现

  • 610例(0.74%)报告为人因错误,155例(0.19%)为装置相关错误。
  • 装置相关错误中最常见类型为输注不足,共47例(30.3%)。
  • 电子可编程泵的错误发生率在4种装置中最高,为15,052例中70例(0.47%,95%CI 0.36-0.59)。
  • 152例装置错误患者中96例(63%)出现某种不良结局,范围从轻微症状到呼吸骤停。

临床怎么用

  • 临床应关注PCA装置相关错误,尤其是电子可编程泵的错误风险。
  • 需加强PCA装置使用监测与患者安全措施,以减少错误及不良结局。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Patient-controlled analgesia (PCA) is one of the most popular and effective methods for managing postoperative pain. Various types of continuous infusion pumps are available for the safe and accurate administration of analgesic drugs. Here we report the causes and clinical outcomes of device-related errors in PCA. METHODS: Clinical records from January 1, 2011 to December 31, 2014 were collected by acute pain service team nurses in a 2715-bed tertiary hospital. Devices for all types of PCA, including intravenous PCA, epidural PCA, and nerve block PCA, were included for analysis. The following 4 types of infusion pumps were used during the study period: elastomeric balloon infusers, carbon dioxide-driven infusers, semielectronic disposable pumps, and electronic programmable pumps. We categorized PCA device-related errors based on the error mechanism and clinical features. RESULTS: Among 82,698 surgical patients using PCA, 610 cases (0.74%) were reported as human error, and 155 cases (0.19%) of device-related errors were noted during the 4-year study period. The most common type of device-related error was underflow, which was observed in 47 cases (30.3%). The electronic programmable pump exhibited the high incidence of errors in PCA (70 of 15,052 patients; 0.47%; 95% confidence interval, 0.36-0.59) among the 4 types of devices, and 96 of 152 (63%) patients experienced some type of adverse outcome, ranging from minor symptoms to respiratory arrest. CONCLUSIONS: The incidence of PCA device-related errors was <0.2% and significantly differed according to the infusion pump type. A total of 63% of patients with PCA device-related errors suffered from adverse clinical outcomes, with no mortality. Recent technological advances may contribute to reducing the incidence and severity of PCA errors. Nonetheless, the results of this study can be used to improve patient safety and ensure quality care.

来源:Anesth Analg,查看原文。版权归原作者所有。

引用

  1. Son HJ, Kim SH, Ryu JO, et al. Device-Related Error in Patient-Controlled Analgesia: Analysis of 82,698 Patients in a Tertiary Hospital. Anesth Analg. 2019. PMID: 31425211. doi:10.1213/ANE.0000000000003397

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。