成人鼻咽距离与商用鼻气管导管选择评估
本研究基于200例成人头颈胸部CT图像,测量鼻至甲状软骨下缘距离(NLD),并对比多厂商鼻气管导管尺寸,发现NLD与性别、身高、体重相关,仅用身高的分位数回归模型即可准确预测NLD,验证集92.8%预测值误差在±20 mm内;同尺寸导管长度差异显著。
要点
- 成人NLD与性别、身高、体重相关,仅用身高即可较准确预测。
- 同型号不同厂商鼻气管导管长度及弯曲至套囊距离差异大,选管需谨慎。
- 基于身高的预测模型有助于选择合适导管,避免导管过深或套囊位于声带水平。
核心结论
- 仅用身高的分位数回归模型可较准确预测成人鼻咽距离,同尺寸鼻气管导管长度差异大,需参考弯曲至套囊距离选管。
研究设计与样本
- 回顾性分析388例成人头颈上胸部CT图像,最终分析200例数据,测量鼻至甲状软骨下缘距离并对比多厂商鼻气管导管尺寸。
主要发现
- NLD与性别、身高、体重相关,仅用身高的分位数回归模型预测准确,验证集92.8%的预测值误差在±20 mm内。
- 同尺寸鼻气管导管的外径、比例、鼻咽部分及引导标记差异明显,弯曲至套囊距离在218至270 mm之间。
关键数值与推荐
- 验证集92.8%的NLD预测值误差在±20 mm内。
- 同尺寸导管弯曲至套囊距离范围为218至270 mm。
临床怎么用
- 临床可基于患者身高使用分位数回归模型预测鼻咽距离,结合导管弯曲至套囊距离选择合适鼻气管导管。
局限性
- 原文未报告
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Preformed nasal endotracheal tubes (NETs) come with a predefined insertion depth due to their curved design. While size indication refers to internal diameter, there is a considerable variability in the corresponding lengths and proportions of same-sized tubes of different manufacturers which is probably based on the lack of data of nasolaryngeal distances (NLDs) in the adult population. Choosing the best-fitting NET is therefore difficult and carries the risk of endobronchial intubation or, on the contrary, cuff inflation at the vocal cord level. The aim of this study was to develop a prediction model for NLD and a selection guide to choose the appropriate NET based on a radiographic description of NLD in comparison to the measurements of available NETs of several manufacturers. METHODS: After institutional ethics board review, 388 computed tomography (CT) scan images of head, neck, and upper thorax in a heterogeneous adult cohort were included. Mean distances from the nares to the lower border of the thyroid cartilage were measured. NETs from different manufacturers were measured and compared to the NLD derived from the radiographic analysis. The patients' sex, body height, and weight were considered as possible covariates in quantile regression models for predicting the NLD. RESULTS: Data from 200 patients were analyzed. NLD was associated with sex, body height, and weight. A simple quantile regression model using the body height as the only covariate sufficed to achieve accurate predictions of NLD. Validation on independent test data showed that 92.8% of the NLD predictions were closer than ±20 mm to the observed NLD values. Measurements of equal-sized NETs varied considerably in outer diameter, proportion, the nasopharyngeal part, and guide marks. Length differences of the bend-to-cuff distance, containing the anatomically NLD, ranged between 218 and 270 mm at same sizes. CONCLUSIONS: A reliable prediction of NLD can be obtained simply by body height, using the formula (Equation is included in full-text article.). As manufacturers' tube lengths vary substantially, additional information about the bend-to-cuff distance as corresponding tube section would allow for more accurate tube selection.
引用
- Massoth C, Schülke C, Köppe J, et al. Nasolaryngeal Distances in the Adult Population and an Evaluation of Commercially Available Nasotracheal Tubes. Anesth Analg. 2020. PMID: 31162158. doi:10.1213/ANE.0000000000004241
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。