困难喉镜预测患者中Macintosh直接喉镜与GlideScope视频喉镜施力对比的随机对照试验
结果显示GlideScope可降低峰值力和平均力,但喉镜操作时间延长,导致冲量力两组相似。
要点
- GlideScope视频喉镜较Macintosh直接喉镜降低峰值力(17 vs 21 N)和平均力(6 vs 11 N)。
- GlideScope操作时间更长(30 vs 18 s),冲量力两组相似(206 vs 175 N,P=0.92)。
- 临床需权衡视频喉镜降低组织受力与操作时间延长带来的冲量累积。
核心结论
- 在具有困难直接喉镜预测特征的患者中,GlideScope视频喉镜较Macintosh直接喉镜降低峰值力和平均力,但因操作时间延长,冲量力两组相似。
研究设计与样本
- 通过将三个FlexiForce传感器(A201-25)粘贴于Macintosh和GlideScope喉镜刀片凹面测量对口咽组织的施力,麻醉医师或经验丰富的麻醉住院医师按随机顺序使用两种设备。
- 主要结局为峰值力,次要结局为平均力和冲量力(力对作用时间的积分)。
主要发现
- 40例患者获得完整数据。
- GlideScope组峰值力降低(17 vs 21 N,P=0.03)。
- GlideScope组平均力降低(6 vs 11 N,P<0.001)。
- GlideScope组喉镜操作时间延长(30 vs 18 s,P<0.001)。
- 冲量力两组相似(中位数206 vs 175 N,P=0.92)。
临床怎么用
- 对具有困难直接喉镜预测特征的患者,选用GlideScope视频喉镜可降低口咽组织所受峰值力和平均力。
- 需注意视频喉镜操作时间延长可能抵消力降低的优势,冲量力(力与时间的乘积)两组无差异。
局限性
- 最终仅40例患者获得完整数据。
- 研究在单中心进行,样本量较小。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: In patients with predictive features associated with easy direct laryngoscopy, videolaryngoscoy with the GlideScope has been shown to require less force when compared with Macintosh direct laryngoscopy. OBJECTIVE: The aim of this study was to compare forces applied with Glidescope vs. Macintosh laryngoscopes in patients with predictive features associated with difficult direct laryngoscopy. DESIGN: A randomised study. SETTING: Toronto General Hospital, a university tertiary centre in Canada. PATIENTS: Forty-four patients aged over 18 years, with one or more features of difficult intubation, undergoing elective surgery requiring single-lumen tracheal intubation. INTERVENTION: We measured the force applied to oropharyngeal tissues by attaching three FlexiForce Sensors (A201-25) to the concave surface of Macintosh and GlideScope laryngoscope blades.Anaesthetists or experienced anaesthesia residents performed laryngoscopies with both devices in a randomised sequence. MAIN OUTCOME MEASURES: The primary outcome was peak force. The secondary outcomes were average force and impulse force. The latter is the integral of the force over the time during which the force acted. RESULTS: Complete data were available for 40 individuals. Peak and average forces decreased with GlideScope (17 vs. 21 N, P = 0.03, and 6 vs. 11 N, P < 0.001, respectively). Laryngoscopy time increased with the GlideScope (30 vs. 18 s, P < 0.001), resulting in similar median impulse forces (206 vs. 175 N, P = 0.92). CONCLUSION: GlideScope laryngoscopy resulted in reduced peak and average forces, but as the laryngoscopy duration increased, the product of force and time (impulse force) was similar with both devices. TRIAL REGISTRATION: Clinicaltrials.gov Identifier: NCT01814176.
引用
- Cordovani D, Russell T, Wee W, et al. Measurement of forces applied using a Macintosh direct laryngoscope compared with a Glidescope video laryngoscope in patients with predictors of difficult laryngoscopy: A randomised controlled trial. Eur J Anaesthesiol. 2019. PMID: 30308524. doi:10.1097/EJA.0000000000000901
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。