注射速率对在线与针尖传感器测量注射压力曲线的影响:体外研究
该体外研究用注射泵以5、10、15 ml·min⁻¹三种速率向硅胶凝胶注射5 ml生理盐水,用光纤压力传感器同步记录在线和针尖压力曲线。共完成15次注射、30条压力曲线,结果显示在线峰值压力随注射速率增加而升高,针尖压力保持恒定;在线测量无法准确反映针尖实时压力变化。
要点
- 在线测压峰值随注射速率增加而升高,针尖测压不随速率变化。
- 在线测压不能准确记录针尖实时压力变化,差异显著。
- 准确监测注射压力应选择针尖测压而非在线测压。
核心结论
- 在线测量的注射压力是不可靠的替代指标,准确监测真实注射压力需在针尖处测量。
研究设计与样本
- 体外研究,向硅胶凝胶注射5 ml生理盐水,三种速率各5次,共15次注射、30条压力曲线。
主要发现
- 在线峰值压力随注射速率增加而升高:5 ml·min⁻¹时为30.76 (3.45) kPa,15 ml·min⁻¹时为72.25 (1.55) kPa。
- 针尖压力在5 ml·min⁻¹和15 ml·min⁻¹时分别为19.92 (1.22) kPa和20.93 (2.66) kPa,保持恒定。
- 在线测压与针尖测压的平均(95%CI)压力差分别为10.8 (6.98-14.70) kPa和51.2 (47.52-54.89) kPa。
临床怎么用
- 外周神经阻滞中为准确监测注射压力,应使用针尖压力传感器而非在线测压。
局限性
- 研究为体外硅胶凝胶模型,结论外推到临床人体需谨慎。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Alongside ultrasonic visualisation, measurement of injection pressure is an effective tool for reducing the risk of intraneural injection during peripheral nerve block. The aim of this study was to compare injection pressure profiles when measured along the injection line with the pressure measured directly at the needle tip using different rates of injection. A syringe pump delivered a 5-ml injection of saline into silicone gel at three different speeds (5 ml.min-1 , 10 ml.min-1 and 15 ml.min-1 ). Fibreoptic pressure sensors recorded real-time pressure profiles of the injection pressure directly at the needle tip and along the injection line. A total of 15 injections were successfully performed, five for each injection rate, totalling 30 recorded pressure profiles. More rapid rates of injection caused peak pressure measured in-line to increase, whereas pressure measured at the needle tip remained constant (mean (SD) pressure in-line 30.76 (3.45) kPa vs. 72.25 (1.55) kPa and mean (SD) pressure at needle tip 19.92 (1.22) kPa vs. 20.93 (2.66) kPa at 5 ml.min-1 and 15 ml.min-1 , respectively). Injection pressure profiles showed that in-line pressure measurement failed to record precise real-time pressure changes occurring at the needle tip (mean (95%CI) pressure difference 10.8 (6.98-14.70) kPa vs. 51.2 (47.52-54.89) kPa for in-line and needle-tip measures, respectively). We conclude that, in order to accurately monitor the true injection pressure generated, independent from operator and injection parameters, measurement at the needle tip is necessary, as injection pressure measured along the injection line is an unreliable surrogate.
引用
- Saporito A, Quadri C, Kloth N, et al. The effect of rate of injection on injection pressure profiles measured using in-line and needle-tip sensors: an in-vitro study. Anaesthesia. 2018. PMID: 30133695. doi:10.1111/anae.14415
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。