新型环甲膜切开导引器在模拟肥胖猪模型中的随机交叉对照评价
该研究在模拟肥胖猪喉模型中对新型环甲膜切开导引器与手术刀环甲膜切开术进行随机交叉比较,共26名参与者。导引器组插入时间更短(85秒 vs 84秒,p=0.030),成功率100% vs 92%,假道发生率31% vs 65%(p=0.022),气管损伤、易用性和设备偏好无统计学差异。
要点
- 导引器组环甲膜切开成功率100%,手术刀组92%。
- 导引器组假道发生率31%,显著低于手术刀组的65%。
- 两组气管损伤、易用性和设备偏好无统计学差异。
核心结论
- 新型环甲膜切开导引器在模拟肥胖猪模型中显示出作为紧急颈前入路辅助工具的潜力。
- 经充分培训后,该导引器可能补充手术刀-探条环甲膜切开技术。
研究设计与样本
- 随机交叉对照研究,在模拟肥胖猪喉模型中进行。
- 共26名参与者比较新型环甲膜切开导引器与手术刀环甲膜切开术。
主要发现
- 导引器组插入时间中位数(IQR[范围])为85(65-123[48-224])秒,手术刀组为84(72-184[46-377])秒,p=0.030。
- 导引器组26名(100%)参与者成功完成环甲膜切开,手术刀组24名(92%)成功。
- 导引器组进入气管所需尝试次数少于手术刀组(p=0.046)。
- 导引器组假道发生8次(31%),手术刀组17次(65%),p=0.022。
- 气管损伤(p=0.490)、易用性(p=0.220)和设备偏好(p=0.240)无统计学差异。
临床怎么用
- 在模拟肥胖模型中,导引器可提高环甲膜切开成功率并减少假道形成。
- 该导引器可作为手术刀-探条技术的补充,但需经过充分培训。
局限性
- 研究使用模拟肥胖猪喉模型,结果不能直接外推至人体。
- 样本量为26名参与者,规模有限。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
The Difficult Airway Society 2015 guidelines for management of unanticipated difficulties in tracheal intubation in adults have generated much discussion regarding Plan D: emergency front-of-neck access with a scalpel-bougie cricothyroidotomy technique. There is concern that this technique may not provide an adequate pathway for the bougie and subsequently the tracheal tube, especially in obese patients with deeper airway structures. This could lead to the formation of a false passage, trauma and failure. A novel cricothyroidotomy introducer, 8 mm wide and 170 mm long, with a sharp leading edge and guiding channel to pass a bougie into the trachea, has been designed to complement the scalpel cricothyroidotomy technique. A comparison study of the use of this novel introducer with the scalpel technique in a simulated obese porcine laryngeal model demonstrated shorter insertion times (median (IQR [range]) 85 (65-123 [48-224]) s vs. 84 (72-184 [46-377]) s, p = 0.030). All 26 (100%) participants successfully performed cricothyroidotomy in the introducer group, whereas only 24 (92%) participants were successful in the scalpel group. The introducer group required fewer attempts to access the trachea compared with the scalpel group (p = 0.046). False passages occurred eight (31%) times in the introducer group compared with 17 (65%) times in the scalpel group (p = 0.022). There were no statistical differences in tracheal trauma (p = 0.490), ease of use (p = 0.220) and device preference (p = 0.240). This novel cricothyroidotomy introducer has shown promising results in securing the airway in an emergency front-of-neck access situation. With robust training, this introducer could potentially be complementary to the scalpel-bougie cricothyroidotomy technique.
引用
- Yeow C, Greaney L, Foy C, et al. Evaluation of a novel cricothyroidotomy introducer in a simulated obese porcine model: a randomised crossover comparison with scalpel cricothyroidotomy. Anaesthesia. 2018. PMID: 30006929. doi:10.1111/anae.14321
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。