文献速递小儿麻醉气道管理

现有小儿气管导管置管策略下导管尖端与套囊位置研究

Tube tip and cuff position using different strategies for placement of currently available paediatric tracheal tubes.
PubMed 文献 · Br J Anaesth · 2018-06-12

该研究汇总三项临床研究数据,纳入422名出生至16岁儿童,计算七种带套囊小儿气管导管品牌在三种置管策略下的导管尖端与套囊位置。结果显示所有研究的带套囊导管均存在重大设计缺陷,可导致气道并发症,现有替代置管策略仅能安全放置无套囊导管。

要点

  • 基于深度标记置管可致高达18%的意外支气管插管和高达91%的套囊位于声门下。
  • 导管尖端置于计算安全距离可使高达54%的套囊位置过高。
  • 气管中段置管导致100%套囊位于声门下或声门上。

核心结论

  • 所有研究的带套囊小儿气管导管存在重大设计缺陷,可能致气道并发症。
  • 替代置管策略可安全放置无套囊导管,但不能安全放置带套囊导管。

研究设计与样本

  • 汇总三项临床研究数据,共纳入422名出生至16岁儿童,记录七种带套囊小儿气管导管品牌(ID 3.0-7.0 mm)的尺寸数据。

主要发现

  • 基于深度标记置管致意外支气管插管高达18%,套囊位于声门下高达91%。
  • 导管尖端置于计算安全距离致高达54%的套囊位置过高。
  • 气管中段置管致100%套囊位于声门下或声门上。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Tracheal tube tip and cuff positions of different cuffed paediatric tracheal tube brands in the trachea can vary with design. METHODS: Data from three clinical studies with measured tracheal lengths were pooled in a database including 422 children aged from birth to 16 yr. Dimensional data of seven cuffed paediatric tracheal tube brands (ID 3.0-7.0 mm) were recorded. Positions of tracheal tube tip and upper cuff border were calculated for each of the 422 tracheas using depth mark based tracheal tube placement, placement of the tracheal tube tip at a calculated safety distance above the carina, and mid-tracheal tube placement. Percentage of accidental bronchial intubations and tracheal tube cuff positions in the subglottic or supraglottic region were calculated. RESULTS: Depth mark based tracheal tube placement resulted in accidental bronchial intubation of up to 18% of cases and tracheal tube cuffs being placed in the subglottic region in up to 91%. Tracheal tube tip placement at a calculated safety distance resulted in up to 54% of tube cuffs placed too high. Mid-tracheal tube placement led to 100% subglottic or supraglottic tracheal tube cuff positions. CONCLUSIONS: All studied cuffed paediatric tracheal tubes have major design flaws potentially leading to airway complications. Tracheal tube manufacturers are urgently asked to improve the design of cuffed paediatric tracheal tubes. Alternative strategies for tracheal tube placement can allow safe tracheal tube placement of uncuffed but not of cuffed tracheal tubes.

来源:Br J Anaesth,查看原文。版权归原作者所有。

引用

  1. Kemper M, Imach S, Buehler PK, et al. Tube tip and cuff position using different strategies for placement of currently available paediatric tracheal tubes. Br J Anaesth. 2018. PMID: 30032890. doi:10.1016/j.bja.2018.05.002

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。