温度对i-gelTM套囊硬度和回弹性的影响:台架研究

Changes in hardness and resilience of i-gelTM cuffs with temperature: a benchtop study.
PubMed 文献 · Anaesthesia · 2018-05-03

本研究用食品工业质构仪在10–60°C宽温范围内测试2号和3号i-gel套囊的硬度和回弹性。结果显示套囊硬度和回弹性总体随温度升高而降低,但在临床可能遇到的室温至体温范围(21–37.4°C)内变化幅度很小。

要点

  • i-gel套囊硬度与回弹性随升温总体降低,但临床温区内变化极小。
  • 室温至体温范围内,2号和3号i-gel硬度分别变化-3.15%和+0.47%。
  • 该温区内回弹性分别下降1.85%和2.68%,提示软化假说不能完全解释密封改善。

核心结论

  • 套囊硬度和回弹性随升温总体降低,但在临床使用温度范围内变化幅度很小。

研究设计与样本

  • 台架研究:将2号和3号i-gel固定在温控水浴平台上,用质构仪球形探头以1 mm·s-1速率压入4 mm后回撤,记录向上压力。
  • 在10°C至60°C(5°C递增)共11个温度下各运行3次,获得105,864个数据点,计算硬度和回弹性。

主要发现

  • 10–60°C范围内,2号和3号i-gel的最小硬度值分别占最大值的88.2%和89.8%,回弹性最小值分别占最大值的92.8%和86.2%。
  • 室温至体温范围(21–37.4°C)内,2号和3号i-gel硬度分别下降3.15%和上升0.47%,回弹性分别下降1.85%和2.68%。

临床怎么用

  • 临床体温范围内i-gel套囊硬度与回弹性变化极小,提示置入后密封改善可能不完全由加温软化解释。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

The i-gelTM is a supraglottic airway with a gel-like thermoplastic cuff. It has been suggested that the seal around the larynx improves following insertion. Perhaps the most intuitive hypothesis proposed for this is that cuff softening occurs during warming from ambient to body temperature. We investigated this using a food industry texture analyser over a wide temperature range. Size 2 and 3 i-gels were secured to a platform within a temperature-controlled water bath, which was in turn mounted on a texture analyser test stand. Both water and i-gel cuff temperatures were recorded. A spherical probe was advanced 4 mm into the surface of each i-gel at a rate of 1 mm.s-1 , then retracted at the same rate while the upward pressure on the probe was recorded. Three runs made at each of the 11 temperatures (10 °C to 60 °C, 5 °C increments) gave 105,864 data points, from which values for hardness (the peak force on the probe at maximum indentation), and resilience (the rate at which the material recovers its original shape) were calculated. Over 10 to 60 °C, the smallest hardness value expressed as a proportion of the largest was 88.2% and 89.8% for size 2 and 3 i-gels, respectively, and for resilience these were 92.8% and 86.2%, respectively. Over room temperature to body temperature range (21-37.4 °C), hardness decreased by 3.15% and increased by 0.47% for i-gel sizes 2 and 3, respectively, whereas resilience values decreased by 1.85% and 2.68%, respectively. Cuff hardness and resilience did generally reduce with warming, but the effect was minimal over temperature ranges that may be encountered during clinical use.

来源:Anaesthesia,查看原文。版权归原作者所有。