iThermonitor WT701腋温在非心脏手术成人中良好代表核心体温
本研究纳入80例气管插管上腹部手术成人,比较无线腋温计iThermonitor WT701与远端食管核心体温,共3339对数据。腋温与食管温差平均0.14°C±0.26°C,91%测量绝对差在0.5°C内,Bland-Altman 95%一致性界限为±0.52°C,提示腋温可良好代表核心体温。
要点
- iThermonitor腋温与食管核心体温平均差仅0.14°C。
- 91%的配对测量绝对差在0.5°C以内。
- 该设备适用于非插管患者体温监测。
核心结论
- iThermonitor WT701记录的腋温在非心脏手术成人中良好代表核心体温,适合临床使用。
研究设计与样本
- 前瞻性研究纳入80例气管插管上腹部手术成人,每5分钟同步测量腋温与远端食管温度,共3339对数据。
主要发现
- 食管减腋温平均差0.14°C±0.26°C。
- Bland-Altman 95%一致性界限上限0.66°C、下限-0.38°C,即±0.52°C。
- 绝对差在0.5°C内的测量占91%(95%CI 88%-93%)。
关键数值与推荐
- 临床可接受准确性预设为多数估计在食管参考值±0.5°C内。
- 精密度合适预设为一致性界限在±0.5°C内。
临床怎么用
- 在椎管内麻醉等无法使用食管或鼻咽温监测时,可用iThermonitor腋温替代核心体温监测。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Core temperature can be accurately measured from the esophagus or nasopharynx during general anesthesia, but neither site is suitable for neuraxial anesthesia. We therefore determined the precision and accuracy of a novel wireless axillary thermometer, the iThermonitor, to determine its suitability for use during neuraxial anesthesia and in other patients who are not intubated. METHODS: We enrolled 80 adults having upper abdominal surgery with endotracheal intubation. Intraoperative core temperature was measured in distal esophagus and was estimated at the axilla with a wireless iThermonitor WT701 (Raiing Medical, Boston MA) at 5-minute intervals. Pairs of axillary and reference distal esophageal temperatures were compared and summarized using linear regression and repeated-measured Bland-Altman methods. We a priori determined that the iThermonitor would have clinically acceptable accuracy if most estimates were within ±0.5°C of the esophageal reference, and suitable precision if the limits of agreement were within ±0.5°C. RESULTS: There were 3339 sets of paired temperatures. Axillary and esophageal temperatures were similar, with a mean difference (esophageal minus axillary) of only 0.14°C ± 0.26°C (standard deviation). The Bland-Altman 95% limits of agreement were reasonably narrow, with the estimated upper limit at 0.66°C and the lower limit at -0.38°C, thus ±0.52°C, indicating good agreement across the range of mean temperatures from 34.9°C to 38.1°C. The absolute difference was within 0.5°C in 91% of the measurements (95% confidence interval, 88%-93%). CONCLUSIONS: Axillary temperature, as recorded by the iThermonitor WT701, well represents core temperature in adults having noncardiac surgery and thus appears suitable for clinical use.
引用
- Pei L, Huang Y, Mao G, et al. Axillary Temperature, as Recorded by the iThermonitor WT701, Well Represents Core Temperature in Adults Having Noncardiac Surgery. Anesth Analg. 2018. PMID: 29293176. doi:10.1213/ANE.0000000000002706
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。