文献速递气道管理

不同实验条件下纤维支气管镜尖端旋转的有效性:一项体外交叉研究

Effectiveness of tip rotation in fibreoptic bronchoscopy under different experimental conditions: an in vitro crossover study.
PubMed 文献 · Br J Anaesth · 2017-01-01

该体外交叉研究将五种不同尺寸(外径2.2、2.4、3.5、4.2、5.2 mm)的纤维支气管镜经合适尺寸喉罩置入小儿气道模型,在镜体或线缆处顺时针或逆时针旋转45、90、180°,线缆保持直或弯曲,评估尖端旋转与操作旋转的比值。结果显示线缆弯曲时镜体旋转效率显著降低,而线缆旋转可减轻该差异;外径2.2和2.4 mm的细镜镜体旋转传递至尖端的效果较差。结论建议伸直线缆并联合镜体与线缆旋转。

要点

  • 线缆旋转比单纯镜体旋转更有效地传递尖端旋转。
  • 线缆弯曲时镜体旋转效率显著下降,线缆旋转可减轻差异。
  • 外径2.2和2.4 mm细镜的镜体旋转传递效果较差。

核心结论

  • 线缆旋转比单纯镜体旋转更有效地旋转纤维支气管镜尖端。
  • 建议伸直纤维支气管镜线缆并联合镜体与线缆旋转。

研究设计与样本

  • 体外交叉研究,使用五种不同外径(2.2、2.4、3.5、4.2、5.2 mm)的纤维支气管镜对。
  • 经合适尺寸喉罩置入小儿气道模型,在镜体或线缆处旋转45、90、180°,线缆保持直或弯曲。

主要发现

  • 总体而言,线缆弯曲时镜体旋转效率显著降低,平均差异范围为29.8%(95%CI 8.8-50.9)至117.4%(95%CI 93.6-141.2)。
  • 线缆旋转时该差异减小。
  • 外径2.2和2.4 mm的细镜在镜体旋转传递至尖端方面较差。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Proper manipulation of fibreoptic bronchoscopes is essential for successful tracheal intubation or diagnostic bronchoscopy. Failure of proper navigation and rotation of the fibrescope may lead to difficulties in advancing the fibrescope and might also be responsible for (unnecessary) difficulties and delays in fibreoptic tracheal intubation, with subsequent hypoxaemia. The present study, therefore, aimed to assess the effectiveness of tip rotation in flexible bronchoscopes in different experimental conditions. METHODS: Five differently sized pairs of fibrescopes (outer diameters of 2.2, 2.4, 3.5, 4.2, and 5.2 mm) were inserted into paediatric airway manikins via an appropriately sized laryngeal mask and were turned clockwise or anticlockwise at the fibrescope body or cord to 45, 90, and 180°, with the cord held either straight or bent. The primary outcome measure was the ratio of rotation measured at the tip over the rotation performed with the fibrescope body or cord. RESULTS: Overall, the 'body' turn was significantly less effective when a bent cord was present (mean difference ranging from 29.8% (95% confidence interval 8.8-50.9) to 117.4% (93.6-141.2). This difference was diminished when the 'cord' turn was performed. Smaller fibrescopes, with outer diameters of 2.2 and 2.4 mm, were inferior with respect to the transmission of 'body' rotation to the tip. CONCLUSIONS: 'Cord' turning of the fibrescope appears to be more effective in rotating the tip than a turn of the fibrescope 'body' only. Straightening the fibrescope cord and combined 'body' and 'cord' turning are recommended.

来源:Br J Anaesth,查看原文。版权归原作者所有。

引用

  1. Piegeler T, Clausen NG, Weiss M, et al. Effectiveness of tip rotation in fibreoptic bronchoscopy under different experimental conditions: an in vitro crossover study. Br J Anaesth. 2017. PMID: 29028928. doi:10.1093/bja/aex322

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。