外周神经阻滞针尖注射压力实时连续监测新方法的描述

Real-time continuous monitoring of injection pressure at the needle tip for peripheral nerve blocks: description of a new method.
PubMed 文献 · Anaesthesia · 2017-09-14

该可行性研究在4个猪下肢解剖模型上以10 ml·min⁻¹分次注射2 ml生理盐水,验证一种新型神经阻滞针实时连续监测针尖注射压力的原理,并与注射管路压力比较。

要点

  • 新型神经阻滞针可实时连续监测针尖注射压力。
  • 注射管路压力不能可靠代表针尖注射压力。
  • 针尖与管路压力差随注射量增加而增大。

核心结论

  • 所描述系统可实现针尖注射压力的实时连续监测。
  • 注射管路测得的压力值不能假定为针尖注射压力的可靠指标。

研究设计与样本

  • 可行性研究,在4个猪下肢解剖模型上进行外周神经阻滞注射。

主要发现

  • 针尖与注射管路压力差的均值(95%CI)在注射0.5 ml时为14.33(12.58–16.08)kPa。
  • 针尖与注射管路压力差的均值(95%CI)在注射结束时为41.56(39.66–43.45)kPa。

临床怎么用

  • 临床行外周神经阻滞时,应认识到注射管路压力不能替代针尖注射压力监测。

局限性

  • 研究在猪下肢解剖模型上进行,非人体临床环境。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

The measurement of injection pressure during the performance of peripheral nerve blocks can be pivotal to detect intraneural placement of the needle tip and thus avoid intrafascicular injection. However, injection pressure can only be measured along the injection line (tubing), which is influenced by several factors. The primary aim of this feasibility study was to describe and validate the principle of a novel nerve-block needle conceived for real-time continuous monitoring of injection pressures at the needle tip. Our secondary aim was to provide measurements and compare injection pressure values at the needle tip and in the injection line. Four porcine lower limb anatomic models were prepared and extraneural injections were performed with fractioned boluses of 2 ml saline at a controlled infusion rate of 10 ml.min-1 (0.16 ml.s-1 ). Injection pressure at the needle tip was monitored and compared with the pressure in the injection line. The system proved to be reliable. Thirty injections were successfully performed without technical failures. The mean (95%CI) difference between pressures at the needle tip and the injection line varied substantially from 14.33 (12.58-16.08) kPa at 0.5 ml injected volume to 41.56 (39.66-43.45) kPa at the end of the injection. This study demonstrates that the described system allows for real-time continuous monitoring of injection pressure at the needle tip. Moreover, this study shows that injection pressure values measured in the injection line cannot be assumed to be a reliable indicator of the injection pressure at the needle tip.

来源:Anaesthesia,查看原文。版权归原作者所有。