无创呼吸容量监测仪在儿科全麻患者中的评估
这项前瞻性观察研究纳入72例ASA I-III级、气管插管全麻的儿科患者,比较无创呼吸容量监测仪(RVM)与在线监测 spirometer(NM3)在切皮前和术毕后10分钟内的分钟通气量(MV)、潮气量(TV)和呼吸频率(RR)测量。结果显示RVM与NM3测量值具有等效性(±10%以内),支持RVM在儿科机械通气患者中可接受的一致性。
要点
- RVM与NM3在儿科机械通气患者中测量MV、TV和RR具有等效性(±10%以内)。
- RVM测量偏差小:MV为-3.8%,TV为-4.9%,RR为1.1%。
- RVM可为非插管儿科患者提供客观呼吸参数监测的潜在工具。
核心结论
- RVM与NM3在儿科机械通气患者中的测量具有可接受的一致性。
研究设计与样本
- 前瞻性观察研究,纳入72例ASA I-III级、气管插管全麻的儿科患者(27女,45男)。
主要发现
- RVM与NM3测量偏差:MV为-3.8%(95%一致性界限-19.9%至12.2%),TV为-4.9%(-21.0%至11.3%),RR为1.1%(-4.1%至6.2%)。
- RVM平均测量准确度:MV为11.9%,TV为12.0%,RR为4.2%(0.6次/分)。
- 等效性检验以90%把握度认为RVM与NM3的MV、TV和RR测量在±10%内等效。
临床怎么用
- RVM可用于儿科机械通气患者中连续、无创监测MV、TV和RR。
- RVM可能有助于早期识别儿科患者呼吸功能恶化,但需进一步研究其他临床场景。
局限性
- 研究仅纳入气管插管全麻的儿科患者,结论是否适用于非插管或其他临床场景需进一步验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Pediatric patients following surgery are at risk for respiratory compromise such as hypoventilation and hypoxemia depending on their age, comorbidities, and type of surgery. Quantitative measurement of ventilation in nonintubated infants/children is a difficult and inexact undertaking. Current respiratory assessment in nonintubated patients relies on oximetry data, respiratory rate (RR) monitors, and subjective clinical assessment, but there is no objective measure of respiratory parameters that could be utilized to predict early respiratory compromise. New advances in technology and digital signal processing have led to the development of an impedance-based respiratory volume monitor (RVM, ExSpiron, Respiratory Motion, Inc, Waltham, MA). The RVM has been shown to provide accurate real-time, continuous, noninvasive measurements of tidal volume (TV), minute ventilation (MV), and RR in adult patients.In this prospective observational study, our primary aim was to determine whether the RVM accurately measures TV, RR, and MV in pediatric patients. METHODS: A total of 72 pediatric patients (27 females, 45 males), ASA I to III, undergoing general anesthesia with endotracheal intubation were enrolled. After endotracheal intubation, continuous data of MV, TV, and RR were recorded from the RVM and an in-line monitoring spirometer (NM3 monitor, Phillips Healthcare). RVM and NM3 measurements of MV, TV, and RR were compared during a 10-minute period prior to the incision ("Presurgery") and a 10-minute period after the end of surgery ("Postsurgery"). Relative errors were calculated over 1-minute segment within each 10-minute period. Bias, precision, and accuracy were calculated using Bland-Altman analyses and paired-difference equivalence tests were performed. RESULTS: Combined across the Presurgery and Postsurgery periods, the RVM's mean measurement bias (RVM - NM3 measurement) for MV was -3.8% (95% limits of agreement) (±1.96 SD): (-19.9% to 12.2%), for TV it was -4.9 (-21.0% to 11.3%), and for RR it was 1.1% (-4.1% to 6.2%). The mean measurement accuracies for MV, TV, and RR were 11.9%, 12.0%, and 4.2% (0.6 breaths/min), respectively. Note that lower accuracy numbers correspond to more accurate RVM measurements. The equivalence tests rejected the null hypothesis that the RVM and NM3 have different mean values and conclude with 90% power that the measurements of MV, TV, and RR from the RVM and NM3 are equivalent within ±10%. CONCLUSIONS: Our data indicate acceptable agreement between RVM and NM3 measurements in pediatric mechanically-ventilated patients. Future studies assessing the capability of the RVM to detect respiratory compromise in other clinical settings are needed.
引用
- Gomez-Morad AD, Cravero JP, Harvey BC, et al. The Evaluation of a Noninvasive Respiratory Volume Monitor in Pediatric Patients Undergoing General Anesthesia. Anesth Analg. 2017. PMID: 28759491. doi:10.1213/ANE.0000000000002029
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。