基于声门下CO2监测优化气管导管套囊压力的随机对照试验
这项多中心前瞻性随机对照非劣效性试验纳入ICU机械通气患者,比较基于声门下CO2自动闭环控制套囊压力与每日3次手动测压的效果。自动组套囊周围CO2泄漏量均值低于手动组(0.09 vs 0.22 mmHg AUC/h,P=0.01),且套囊压力处于预设安全范围的时间比例更高。结论为自动控制非劣效且优于手动控制,是优化套囊压力的有效方法。
要点
- 基于声门下CO2的自动闭环套囊压力控制可减少套囊周围CO2泄漏。
- 自动组套囊压力处于安全范围的时间比例显著高于手动组。
- 该方法安全且可适用于任何气管插管患者。
核心结论
- 自动套囊压力控制组不仅非劣效于手动组,而且优于手动组。
- 基于声门下CO2测量的自动套囊压力控制是优化气管导管套囊压力的有效方法,安全且易于使用。
研究设计与样本
- 多中心、前瞻性、随机对照、非劣效性试验,纳入ICU患者,最终分析64例。
- 所有患者使用带额外腔隙监测声门下CO2的AnapnoGuard气管导管。
- 自动组根据声门下CO2自动调整套囊压力,手动组每日3次使用测压计手动管理套囊压力。
主要发现
- 套囊周围CO2泄漏均值(mmHg AUC/h):手动组0.22±0.32,自动组0.09±0.04(P=0.01)。
- 单侧95%CI下限为0.05,证明非劣效(>-0.033);双侧95%CI为0.010至0.196,证明优效(>0.0)。
临床怎么用
- 对机械通气患者可采用基于声门下CO2监测的自动闭环套囊压力控制,以减少套囊周围泄漏并维持套囊压力在安全范围。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Many of the complications of mechanical ventilation are related to inappropriate endotracheal tube (ETT) cuff pressure. The aim of the current study was to evaluate the effectiveness of automatic cuff pressure closed-loop control in patients under prolonged intubation, where presence of carbon dioxide (CO2) in the subglottic space is used as an indicator for leaks. The primary outcome of the study is leakage around the cuff quantified using the area under the curve (AUC) of CO2 leakage over time. METHODS: This was a multicenter, prospective, randomized controlled, noninferiority trial including intensive care unit patients. All patients were intubated with the AnapnoGuard ETT, which has an extra lumen used to monitor CO2 levels in the subglottic space.The study group was connected to the AnapnoGuard system operating with cuff control adjusted automatically based on subglottic CO2 (automatic group). The control group was connected to the AnapnoGuard system, while cuff pressure was managed manually using a manometer 3 times/d (manual group). The system recorded around cuff CO2 leakage in both groups. RESULTS: Seventy-two patients were recruited and 64 included in the final analysis. The mean hourly around cuff CO2 leak (mm Hg AUC/h) was 0.22 ± 0.32 in the manual group and 0.09 ± 0.04 in the automatic group (P = .01) where the lower bound of the 1-sided 95% confidence interval was 0.05, demonstrating noninferiority (>-0.033). Additionally, the 2-sided 95% confidence interval was 0.010 to 0.196, showing superiority (>0.0) as well. Significant CO2 leakage (CO2 >2 mm Hg) was 0.027 ± 0.057 (mm Hg AUC/h) in the automatic group versus 0.296 ± 0.784 (mm Hg AUC/h) in the manual group (P = .025). In addition, cuff pressures were in the predefined safety range 97.6% of the time in the automatic group compared to 48.2% in the automatic group (P < .001). CONCLUSIONS: This study shows that the automatic cuff pressure group is not only noninferior but also superior compared to the manual cuff pressure group. Thus, the use of automatic cuff pressure control based on subglottic measurements of CO2 levels is an effective method for ETT cuff pressure optimization. The method is safe and can be easily utilized with any intubated patient.
引用
- Efrati S, Bolotin G, Levi L, et al. Optimization of Endotracheal Tube Cuff Pressure by Monitoring CO2 Levels in the Subglottic Space in Mechanically Ventilated Patients: A Randomized Controlled Trial. Anesth Analg. 2017. PMID: 28787340. doi:10.1213/ANE.0000000000002358
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。