文献速递气道管理

经自增压一次性声门上气道喉罩盲插管:国际多中心前瞻性队列研究

Blind Intubation through Self-pressurized, Disposable Supraglottic Airway Laryngeal Intubation Masks: An International, Multicenter, Prospective Cohort Study.
PubMed 文献 · Anesthesiology · 2017-01-01

该国际多中心前瞻性队列研究纳入1000例择期全麻气管插管成人,评估Air-Qsp声门上气道装置盲插管成功率。装置在99%病例中提供有效通气和氧合,总盲插管成功率为78%(95%CI 75-81%),但三中心成功率不一致(P<0.001)。结论认为该装置可可靠提供气道并在近80%患者中实现盲插管,是可考虑的初始气道控制方法。

要点

  • Air-Qsp声门上气道装置盲插管总成功率为78%(95%CI 75-81%)。
  • 三中心成功率差异显著(P<0.001),范围41%至84%。
  • 未肌松时漏气压与肌松后漏气压相关性良好,肌松可安全使用。

核心结论

  • 该声门上气道装置可靠提供良好气道,近80%患者可盲插管,是合理的初始气道控制方法。
  • 未肌松时漏气压适当时可安全使用肌松。

研究设计与样本

  • 国际多中心前瞻性队列研究,纳入1000例择期全麻气管插管成人。
  • 常规诱导后置入声门上气道装置并间断通气,测量肌松前后漏气压,最多行两次盲插管。

主要发现

  • 声门上气道装置在99%病例中提供有效通气和氧合。
  • 总盲插管成功率为78%(95%CI 75-81%)。
  • 三中心成功率不一致(P<0.001):瑞士Winterthur中心80%(95%CI 75-85%),波兰Lodz中心41%(95%CI 29-53%),瑞士Zurich中心84%(95%CI 80-88%)。
  • 肌松前漏气压与肌松后漏气压相关性良好。

临床怎么用

  • 可将该声门上气道装置作为气道控制的初始选择之一,盲插管成功率约80%。
  • 未肌松时漏气压适当时可安全给予肌松药。

局限性

  • 盲插管成功率在三中心间不一致(P<0.001),提示存在中心相关影响因素。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Supraglottic airway devices commonly are used for securing the airway during general anesthesia. Occasionally, intubation with an endotracheal tube through a supraglottic airway is indicated. Reported success rates for blind intubation range from 15 to 97%. The authors thus investigated as their primary outcome the fraction of patients who could be intubated blindly with an Air-Qsp supraglottic airway device (Mercury Medical, USA). Second, the authors investigated the influence of muscle relaxation on air leakage pressure, predictors for failed blind intubation, and associated complications of using the supraglottic airway device. METHODS: The authors enrolled 1,000 adults having elective surgery with endotracheal intubation. After routine induction of general anesthesia, a supraglottic airway device was inserted and patients were ventilated intermittently. Air leak pressure was measured before and after full muscle relaxation. Up to two blind intubation attempts were performed. RESULTS: The supraglottic airway provided adequate ventilation and oxygenation in 99% of cases. Blind intubation succeeded in 78% of all patients (95% CI, 75 to 81%). However, the success rate was inconsistent among the three centers (P < 0.001): 80% (95% CI, 75 to 85%) at the Institute of Anesthesia and Pain Therapy, Kantonsspital Winterthur, Winterthur, Switzerland; 41% (95% CI, 29 to 53%) at the Department of Anesthesiology and Intensive Therapy, Medical University of Lodz, Lodz, Poland; and 84% (95% CI, 80 to 88%) at the Institute of Anesthesiology, University Hospital Zurich, Zurich, Switzerland. Leak pressure before relaxation correlated reasonably well with air leak pressure after relaxation. CONCLUSIONS: The supraglottic airway device reliably provided a good airway and allowed blind intubation in nearly 80% of patients. It is thus a reasonable initial approach to airway control. Muscle relaxation can be used safely when unparalyzed leak pressure is adequate.

来源:Anesthesiology,查看原文。版权归原作者所有。

引用

  1. Ruetzler K, Guzzella SE, Tscholl DW, et al. Blind Intubation through Self-pressurized, Disposable Supraglottic Airway Laryngeal Intubation Masks: An International, Multicenter, Prospective Cohort Study. Anesthesiology. 2017. PMID: 28570294. doi:10.1097/ALN.0000000000001710

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。