GlideScope可视喉镜插管时导管通过困难与声门下气道损伤:三种气管导管的随机对照比较
本研究为随机对照临床试验,比较PVC、加强型和BlockBuster三种气管导管在GlideScope可视喉镜下经口气管插管时的通过阻力、插管时间及声门下气道损伤。PVC组插管时间中位数35秒,长于加强型25秒和BlockBuster组25秒(p<0.001)。拔管后纤维支气管镜评估显示PVC组和加强型组声门下气道损伤均高于BlockBuster组(p<0.001和p=0.012),结论认为BlockBuster导管优于PVC和加强型导管。
要点
- GlideScope插管时PVC导管通过困难及声门下损伤风险高于BlockBuster导管
- BlockBuster导管通过阻力最低(1.9%),插管时间中位数25秒
- 拔管后纤维支气管镜评估显示BlockBuster组声门下气道损伤最少
核心结论
- BlockBuster气管导管是GlideScope经口气管插管的更优选择,优于PVC和加强型导管。
主要发现
- PVC组导管通过困难(中度或重度阻力)15例(21.4%),加强型组4例(7.4%),BlockBuster组1例(1.9%),PVC vs BlockBuster p=0.003。
- 插管时间中位数(IQR[范围]):PVC组35(27-45[15-115])秒,加强型组25(20-27[15-110])秒,BlockBuster组25(22-30[16-90])秒,PVC vs 加强型及PVC vs BlockBuster均p<0.001。
- 拔管后纤维支气管镜评估声门下气道损伤:PVC组高于BlockBuster组(p<0.001),加强型组高于BlockBuster组(p=0.012)。
临床怎么用
- 使用GlideScope行经口气管插管时,优先选择BlockBuster导管以减少通过阻力和声门下气道损伤。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Difficulty during placement of the tracheal tube is a known problem when intubating with the GlideScope® , which may lead to subglottic airway injury. This randomised, controlled clinical trial was designed to compare the resistance to passage of PVC (polyvinyl chloride), reinforced or BlockBuster tracheal tubes during intubation with the GlideScope. Secondary outcomes included the time taken to intubate and assessment of subglottic airway injury. One-hundred and seventy-seven patients were included in the data analysis. There was difficult tracheal tube passage (moderate or severe resistance) in 15 (21.4%) patients using the PVC tube compared with 4 (7.4%) and 1 (1.9%) using the reinforced and BlockBuster tubes, respectively (p = 0.003 for PVC vs. BlockBuster). The median (IQR [range]) time taken to intubate was 35 (27-45 [15-115]) s, 25 (20-27 [15-110]) s and 25 (22-30 [16-90]) s, respectively, (p < 0.001 for PVC vs. reinforced as well as PVC vs. BlockBuster). Subglottic airway injury, assessed using a fibreoptic bronchoscope after extubation, was higher with the PVC tube (p < 0.001) and the reinforced tube (p = 0.012) compared with the BlockBuster tube. We conclude that the BlockBuster tracheal tube is a better choice for orotracheal intubation with the GlideScope than PVC or reinforced tubes.
引用
- Su K, Gao X, Xue FS, et al. Difficult tracheal tube passage and subglottic airway injury during intubation with the GlideScope® videolaryngoscope: a randomised, controlled comparison of three tracheal tubes. Anaesthesia. 2016. PMID: 27995626. doi:10.1111/anae.13755
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。