脊髓导管拉伸致尖端标记消失:一例病例报告
本文报告一例ASA4患者在全膝关节置换翻修术中接受连续脊髓麻醉,使用Pajunk IntraLong导管套件。拔管时预期尖端标记缺失,怀疑导管碎片残留,影像学确认无异物。体外拉伸新导管可复制标记消失现象,提示导管材料在张力下可拉伸致标记褪色,模仿残留碎片。
要点
- 脊髓导管在张力下可能拉伸致尖端标记消失,模仿导管残留。
- 拔管时未见尖端标记应先确认导管完整性,避免不必要的影像检查。
- 了解导管机械特性有助于减少患者和临床医生的焦虑。
核心结论
- 脊髓导管材料在张力下可拉伸致尖端标记褪色,可能被误认为导管碎片残留。
研究设计与样本
- 病例报告:一例ASA4患者在全膝关节置换翻修术中接受连续脊髓麻醉,使用Pajunk IntraLong导管套件。
主要发现
- 拔管时预期尖端标记缺失,怀疑导管碎片残留,影像学确认无残留异物。
- 体外拉伸新导管可复制尖端标记消失现象。
临床怎么用
- 拔管时若尖端标记缺失,应考虑到导管拉伸致标记消失的可能,避免不必要的影像检查。
局限性
- 单例病例报告,结论需进一步研究验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Continuous spinal anesthesia (CSA) offers precise, prolonged neuraxial anesthesia suitable for high-risk patients. This technique minimizes hemodynamic instability but comes with notable challenges. Vigilant catheter management is crucial to avoid complications, including the risk of catheter retention. We report a case where the mechanical properties of a spinal catheter led to diagnostic uncertainty. CASE: An ASA4 patient underwent a revision of total knee replacement surgery under CSA using a Pajunk IntraLong catheter set. Following successful insertion, the surgery proceeded uneventfully. During catheter removal, the expected tip markings were absent, raising concern for a retained fragment. Imaging confirmed no retained foreign body. To investigate further, stretching of a new catheter replicated the disappearance of tip markings. This finding suggested that the catheter's material had the potential to stretch under tension, leading to fading of the markings and mimicking a retained fragment. CONCLUSIONS: This case underscores the need for awareness of the mechanical properties of spinal catheters, especially their potential for stretching and loss of tip markings. Understanding these characteristics can prevent unnecessary imaging and reduce patient and clinician anxiety.
引用
- McGrath J, Hussein K. Disappearance of spinal catheter tip due to catheter stretching: a case report. Reg Anesth Pain Med. 2025. PMID: 39842941. doi:10.1136/rapm-2024-106235
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。