McGrath MAC与3D打印视频喉镜对比:医学生模拟人随机非劣效对照研究
在标准气道站,STVL组首次成功率60.7%,3DVL组36.3%,差异24.4%(95%CI 17.5%-31.3%),超过7.5%非劣效界值;困难气道站两组无差异。结论为3DVL在医学生模拟场景中首次插管成功率劣于STVL。
要点
- 在标准气道模拟中,McGrath MAC首次插管成功率显著高于3D打印视频喉镜。
- 3D打印视频喉镜未达到预设的7.5%非劣效界值,表现为劣效。
- 困难气道模拟中两种视频喉镜的首次插管成功率无差异。
核心结论
- 在医学生模拟人研究中,3D打印视频喉镜首次插管成功率劣于McGrath MAC标准视频喉镜,差异超过7.5%非劣效界值。
研究设计与样本
- 受试者随机分为先使用STVL(McGrath MAC)或3DVL组,在标准气道和困难气道站使用Laerdal气道管理训练模拟人进行插管。
- 非劣效界值设定为首次插管成功率的7.5%差异。
主要发现
- 标准气道站:STVL组首次插管成功率为60.7%(n=65),3DVL组为36.3%(n=37),差异24.4%(95%CI 17.5%-31.3%)。
- 困难气道站:两种视频喉镜的首次插管成功率无差异。
临床怎么用
- 对于医学生模拟训练,McGrath MAC视频喉镜在标准气道首次插管成功率上优于低成本3D打印视频喉镜。
局限性
- 研究为模拟人场景,非临床患者,结果外推需谨慎。
- 受试者为医学生,不能直接推广至经验丰富的麻醉医生。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Proficiency in endotracheal intubation (ETI) is essential for medical professionals and its training should start at medical schools; however, large caseload may be required before achieving an acceptable success rate with direct laryngoscopy. Video laryngoscopy has proven to be an easier alternative for intubation with a faster learning curve, but its availability in medical training may be an issue due to its high market prices. We devised a low-cost 3-dimensionally printed video laryngoscope (3DVL) and performed a randomized trial to evaluate if the intubation success rate on the first attempt with this device is noninferior to a standard commercially available video laryngoscope (STVL). METHODS: Two hundred and nine medical students from 5 medical schools were enrolled and randomized to start with the STVL (McGrath MAC) or the 3DVL. Four stations (standard airway using the STVL/3DVL and difficult airway using the STVL/3DVL) were set to simulate ETI with standard Airway Management Trainer manikins (Laerdal Medical Ltd.). The noninferiority margin of 7.5% was defined for the success rate on the first attempt, considering the difference in proportions between the STVL (expected to be higher) and 3DVL groups. RESULTS: Regarding the standard airway station, 60.7% (n = 65) of the students successfully performed TI on the first attempt with the STVL within the established timeframe, compared to 36.3% (n = 37) of the students using the 3DVL. This represented a difference of 24.4% (95% confidence interval, 17.5%-31.3%). Considering the difficult airway station, the success rates on the first intubation attempt with the 2 VLs did not differ. CONCLUSIONS: The 3DVL was inferior in achieving first-attempt intubation when compared with the STVL with a difference in success rate >7.5% margin in simulated scenarios with medical students. Tracheal intubation might require a set of psychomotor skills for which the McGrath MAC device is superior to the low-cost alternative.
引用
- Detoni PB, Nascimento JS, Araújo Azi LMT, et al. McGrath MAC Versus Three-Dimensional Printed Video Laryngoscopes: A Randomized, Manikin-Simulated Noninferiority Controlled Study with Medical Students. Anesth Analg. 2024. PMID: 39804596. doi:10.1213/ANE.0000000000007067
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。