新型环状软骨压迫传感器装置提高Sellick手法中食管封闭效果:一项随机对照试验
该前瞻性随机对照试验在65例ASA I/II成年全麻插管患者中比较新型力反馈装置引导环状软骨压迫与传统操作。反馈组食管开口封闭率显著高于对照组(94% vs 6%,P=0.001),施加力更接近目标且更稳定(22.65±7.81 N vs 14.57±11.13 N,P<0.001),超声测得食管前后径更小(0.49 [0.36-0.56] mm vs 0.57 [0.48-0.65] mm,P=0.006)。结论:该装置可显著提高环状软骨压迫时食管封闭率。
要点
- 力反馈装置引导下食管开口封闭率达94%,显著高于对照组的6%。
- 反馈组施加力更接近目标值且各时间点更稳定(22.65±7.81 N vs 14.57±11.13 N)。
- 反馈组超声测得食管前后径更小,提示压迫效果更佳。
核心结论
- 使用新型传感器装置引导环状软骨压迫可显著提高食管开口封闭率。
主要发现
- 反馈组食管开口封闭率显著高于对照组(94% vs 6%,P=0.001)。
- 反馈组平均施加力在各时间点均显著更优且更稳定(22.65±7.81 N vs 14.57±11.13 N,P<0.001)。
- 反馈组环状软骨压迫期间食管前后径中位数[IQR]小于对照组(0.49 [0.36-0.56] mm vs 0.57 [0.48-0.65] mm,P=0.006)。
临床怎么用
- 临床实施Sellick手法时可借助实时力反馈装置指导操作者施加适当压力,以提高食管封闭效果。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The inability to measure the force applied during cricoid pressure is an important limitation in clinical practice. We developed a novel device to measure this force and provide real-time feedback to the operator. OBJECTIVES: To test the hypothesis of superior oesophageal occlusion during cricoid pressure when guided by the novel device as compared with conventional practice. DESIGN: A prospective randomised controlled trial. SETTING: Tertiary University Hospital, May to December 2021. PATIENTS: Sixty-five ASA I/II adult patients of either sex scheduled for elective surgery under general anaesthesia requiring tracheal intubation. Patients with comorbidities involving the neck, those at risk of pulmonary aspiration, anticipated difficult airway and BMI at least 35 kg m -2 were excluded. INTERVENTION: Operators applied cricoid pressure with their fingers in both groups. Participants were randomised into 'feedback group' where the operator adjusted this force based on the force measured by the novel device, and 'control group' where the force measured by the device was shielded from the operator. MAIN OUTCOME MEASURES: The primary outcome was the rate of occlusion of oesophageal opening assessed by whether a 12 CH orogastric tube could be inserted into the oesophageal entrance during videolaryngoscopy. Secondary outcomes included the magnitudes of force measured during cricoid pressure, oesophageal diameter measured ultrasonographically during cricoid pressure and intubation parameters. RESULTS: Sixty-four patients completed the study. The oesophageal opening was occluded in significantly more patients in the feedback group compared with the control group (94 vs. 6%; P = 0.001). The mean ± SD force (N) applied was significantly better and consistent at all time points in the feedback group compared to the control group (22.65 ± 7.81 vs. 14.57 ± 11.13, P < 0.001). The median [IQR] anteroposterior diameter of the oesophagus during cricoid pressure was less in the feedback group compared to the control group (0.49 [0.36 to 0.56] vs. 0.57 [0.48 to 0.65], P = 0.006). CONCLUSIONS: The use of the novel sensor device achieved a significantly high rate of oesophageal occlusion during application of cricoid pressure. TRIAL REGISTRATION: Clinical Trial Registry of India (CTRI/2021/05/033484).
引用
- Mishra A, Ashok V, Puri GD, et al. A novel cricoid pressure sensor device enhances the efficacy of oesophageal occlusion during Sellick's manoeuvre: A randomised controlled trial. Eur J Anaesthesiol. 2024. PMID: 39512179. doi:10.1097/EJA.0000000000002097
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。