文献速递区域麻醉与超声

针头贯穿神经后束膜损伤罕见:人正中与尺神经离体研究

Fascicular injury is rare following needle transfixion: a study on median and ulnar isolated human nerves.
PubMed 文献 · Reg Anesth Pain Med · 2025-12-05

该尸体研究用4种针(30°斜面22G神经阻滞针及15°斜面22G、25G、27G Quincke腰麻针)对5条正中与5条尺神经各贯穿穿刺40次,共400次,分析144个组织切片。仅发现3处束膜损伤,均由15°斜面针所致,结论为针头贯穿后束膜损伤罕见,单纯针伤难以解释阻滞后神经功能障碍。

要点

  • 针头贯穿离体人正中与尺神经后束膜损伤罕见,400次穿刺仅3处。
  • 3处束膜损伤均与15°斜面针相关,30°斜面神经阻滞针未造成损伤。
  • 单纯针头损伤不太可能是阻滞后神经功能障碍的主要原因。

核心结论

  • 针头贯穿神经后束膜损伤罕见,单纯针伤难以解释阻滞后神经功能障碍。

研究设计与样本

  • 尸体研究,取5条正中神经与5条尺神经(离体),每种针在每条神经上贯穿穿刺10次,共400次穿刺。
  • 使用4种针:30°斜面22G神经阻滞针,以及15°斜面22G、25G、27G Quincke腰麻针。

主要发现

  • 共分析144个组织切片(正中74个、尺神经70个),每切片束膜中位数正中16(范围7-23)、尺神经17(范围8-27)。
  • 束膜/神经外膜组织比例正中45%(范围35%-52%)、尺神经44%(范围39%-54%)。
  • 发现3处束膜损伤:尺神经1处、正中神经2处,均由15°斜面针所致(尺神经为22G,正中神经为22G和27G各一)。

临床怎么用

  • 临床进行神经阻滞时,针头贯穿神经后束膜损伤概率低,不必因担心束膜损伤而过度回避必要的穿刺。
  • 阻滞后神经功能障碍若发生,应综合评估其他原因,而非单纯归因于针头贯穿损伤。

局限性

  • 离体尸体神经研究,无法完全反映活体组织反应与临床神经功能。
  • 样本量有限(5条正中、5条尺神经),且仅评估正中与尺神经。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Needle trauma has been associated with peripheral nerve injury and neurological dysfunction. However, inadvertent needle puncture is frequent while postblock dysfunction is rare. We conducted a cadaveric study to evaluate the association between needle puncture and fascicular injury. METHODS: Five median and five ulnar (isolated) nerves were obtained from fresh human cadavers. Four different needles were used for the transfixing punctures: A 30° beveled (22G) nerve block needle, and 15° beveled (22G, 25G and 27G) Quincke spinal block needles. 10 transfixing punctures were made with each needle type on each nerve (40 punctures per nerve). Samples were then immersed in 5% formaldehyde solution for 30 days. Perpendicular cross-sections of the punctured segments were obtained. Samples were embedded in paraffin and analyzed under light microscopy with H&E staining. On each slice, the following variables were obtained: ratio of fascicular/epineurial tissue, the number of fascicles per nerve and the number of injured fascicles. RESULTS: A total of 400 punctures were made (200 median and 200 ulnar) and 144 histological nerve sections analyzed (74 median and 70 ulnar). The median number of fascicles per section was 16 (range 7-23) and 17 (range 8-27) with a fascicular/epineural tissue ratio of 45% (range 35%-52%) and 44% (range 39%-54%) for median and ulnar, respectively. Three fascicular injuries were identified: one in ulnar and two in median. All injuries were caused by a 15° beveled needle, the ulnar with a 22G and the median with a 22G and a 27G. CONCLUSIONS: Fascicular injury is rare following needle transfixion. Needle injury alone is unlikely to explain postblock neurological dysfunction.

来源:Reg Anesth Pain Med,查看原文。版权归原作者所有。

引用

  1. Varela V, Mejia J, Ruíz C, et al. Fascicular injury is rare following needle transfixion: a study on median and ulnar isolated human nerves. Reg Anesth Pain Med. 2025. PMID: 39209435. doi:10.1136/rapm-2024-105803

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。