i-gel Plus 声门上气道装置在择期手术中的前瞻性国际多中心评估
该国际多中心前瞻性队列研究评估 i-gel Plus 在成人择期全麻手术中的性能,报告前 1000 例患者数据。总体置入成功率为 98.6%,首次置入成功率为 88.2%,女性总体成功率低于男性(97.4% vs. 99.6%),口咽漏气压平均 32 cmH2O,并发症发生率较低。
要点
- i-gel Plus 总体置入成功率高,首次置入成功率亦较高。
- 女性总体置入成功率低于男性,但首次置入成功率无性别差异。
- 低操作者经验是首次置入失败的独立危险因素。
核心结论
- i-gel Plus 是一种有效的声门上气道装置,置入成功率高且并发症发生率较低。
研究设计与样本
- 国际多中心前瞻性队列研究,纳入成人择期全麻手术患者,报告前 1000 例数据。
主要发现
- 总体置入成功率为 98.6%,首次置入成功率为 88.2%。
- 女性与男性总体成功率分别为 97.4% 与 99.6%,首次置入成功率无性别差异。
- 口咽漏气压均值(SD)为 32(7)cmH2O。
- 低操作者经验是首次置入失败的独立危险因素。
- 并发症包括血氧饱和度下降 <85% 占 0.6%,装置带血 7.4%,喉痉挛 0.5%,碗内胃内容物 0.2%。
临床怎么用
- 可用于成人择期全麻手术的声门上气道管理,需注意女性总体成功率略低及操作者经验的影响。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The i-gel® Plus is a modified version of the i-gel® supraglottic airway device. It contains a wider drainage port; a longer tip; ramps inside the breathing channel; and an additional port for oxygen delivery. There has been no prospective evaluation of this device in clinical practice. METHODS: This international, multicentre, prospective cohort study aimed to evaluate the performance of the i-gel Plus in adult patients undergoing elective procedures under general anaesthesia. The primary outcome was overall insertion success rate, defined as the ability to provide effective airway management through the device from insertion until the end of the surgical procedure. Secondary outcomes included device performance and incidence of postoperative adverse events. Data from the first 1000 patients are reported. RESULTS: In total, 1012 patients were enrolled; 12 forms were excluded from the final analysis due to incomplete data leaving 1000 included patients (545 female). Overall insertion success rate was 98.6%, with a first-attempt success rate of insertion of 88.2%. A significant difference between females and males was seen for the overall success rate (97.4% vs. 99.6% respectively) but not for first-attempt successful insertion. Mean (SD) oropharyngeal seal pressure was 32 (7) cmH2O. The only independent factor that increased the risk of first-attempt failure was low operator experience. Complications included desaturation < 85% in 0.6%; traces of blood on the device in 7.4%; laryngospasm in 0.5%; and gastric contents inside the bowl in 0.2% of patients. CONCLUSIONS: The i-gel Plus appears to be an effective supraglottic airway device that is associated with a high insertion success rate and a reasonably low incidence of complications.
引用
- Werner J, Klementova O, Bruthans J, et al. Evaluation of the i-gel® Plus supraglottic airway device in elective surgery: a prospective international multicentre study. Anaesthesia. 2024. PMID: 39110995. doi:10.1111/anae.16401
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。