水循环射频探针显著增加大鼠模型中损伤长度与轴突损害

Water-circulating probes significantly modify lesion length and axon damage in cooled radiofrequency ablations when compared with similar-sized standard radiofrequency probes in rats.
PubMed 文献 · Reg Anesth Pain Med · 2024-06-03

该临床前研究在大鼠坐骨神经上比较水冷射频(CRF)与标准射频(SRF)探针,发现CRF大探针能量输出更高、热损伤区更多、轴突破坏更重;关闭水循环后能量输出与损伤长度显著下降,提示水循环技术本身而非探针尺寸是差异关键。

要点

  • 水冷射频探针的能量沉积、损伤长度与轴突破坏均显著大于标准射频探针。
  • 关闭水循环后,CRF探针的能量输出与损伤长度显著降低,表现趋近SRF探针。
  • 射频模态的结构差异不能仅归因于探针尺寸,水循环设计是重要因素。

核心结论

  • 水循环技术对能量沉积、损伤长度和轴突损伤的影响大于标准射频,且差异不能仅归因于探针尺寸。

研究设计与样本

  • 在大鼠坐骨神经上使用多尺寸探针进行射频消融:CRF为17 Ga/2 mm和17 Ga/4 mm,SRF为22 Ga/5 mm、18 Ga/10 mm和16 Ga/10 mm,样本量n=51。
  • 评估指标包括发生器能量输出、损伤长度、组织学轴突损伤和肌电图神经功能分析。

主要发现

  • CRF大探针比SRF探针传递更多能量(p<0.01),并在坐骨神经中产生多个热损伤区。
  • 关闭水循环后,CRF探针的能量输出(p<0.001)和损伤长度(p<0.05)显著降低。
  • 带水循环的CRF探针比更大尺寸的SRF探针引起显著更多的轴突破坏(p<0.0001)。

临床怎么用

  • 临床选择射频模式时需考虑水循环设计对损伤范围和神经破坏程度的影响,而非仅比较探针尺寸。

局限性

  • 研究为大鼠临床前实验,样本量n=51,结果向临床转化需谨慎。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: Preclinical research demonstrated water-cooled radiofrequency (CRF) ablations have a significant impact on structural and functional changes compared to standard radiofrequency (SRF) ablations. Clinical procedures utilizing RF to treat chronic pain conditions also show sustained functional outcomes. We hypothesize that the design of the RF probes plays an important role in interventional procedure success, but it remains unclear which specific design features. METHODS: RF ablations were performed in male Lewis rats (n=51) using multiple-sized probes for CRF (17 Ga/2 mm and 17Ga/4 mm) and SRF (22Ga/5 mm, 18Ga/10 mm and 16Ga/10 mm) to evaluate generator energy output, lesion length, axon damage by histology and nerve function analysis via electromyography. To exclude probe design variables beyond size and remain objective, we tested cooled probes with and without water circulation, which resulted in the CRF probe performing like an SRF probe. RESULTS: Consistent with our previous findings in smaller probes, CRF large probes delivered more energy (p<0.01) and generated multiple zones of thermal damage in sciatic nerves. When the water-circulating feature was turned off, however, energy output (p<0.001) and lesion length (p<0.05) was significantly reduced. CRF probes with the water circulation also featured significantly more axonal disruption, than larger sized SRF probes (p<0.0001). CONCLUSIONS: Overall, this data confirms that CRF's water-circulating technology has a greater impact on energy deposition, lesion length and axon damage compared with SRF ablations. Moreover, results suggest that the structural differences between RF modalities cannot be solely attributed to probe size, and it may shed light on its differences in clinical outcomes.