Airtraq®可视喉镜与可弯曲鼻支气管镜用于清醒气管插管的随机对照非劣效性研究
本研究比较可弯曲鼻支气管镜与Airtraq®可视喉镜在预期困难气道患者清醒气管插管中的效果。78例患者随机分组,支气管镜组插管成功率97%,Airtraq组82%(p=0.032),Airtraq未达到预设8%的非劣效界值。Airtraq插管时间更短,但并发症与舒适度无显著差异。
要点
- Airtraq可视喉镜在清醒气管插管中未达到对可弯曲支气管镜的非劣效性。
- 可弯曲支气管镜插管成功率97%,Airtraq为82%,差异有统计学意义。
- Airtraq插管时间更短,但需个体化评估其替代可行性。
核心结论
- 在临床环境中,Airtraq可视喉镜对可弯曲支气管镜在清醒气管插管中不具非劣效性。
- 根据个案判断,Airtraq可能是一种合适的替代选择。
研究设计与样本
- 随机对照非劣效性研究,招募78例预期困难气道行清醒气管插管的患者,随机分配至可弯曲鼻支气管镜组或Airtraq可视喉镜组。
- 所有操作均采用上气道区域麻醉阻滞和瑞芬太尼靶控静脉输注。
- 主要结局为分配技术的插管成功率,预设非劣效界值为8%。
主要发现
- 插管成功率:可弯曲支气管镜组97%,Airtraq组82%,p=0.032。
- 气管插管时间中位数(IQR[范围]):Airtraq组163(105-332[40-1004])秒,可弯曲支气管镜组217(180-364[120-780])秒,p=0.030。
- 两组并发症无显著差异。
- 插管容易度视觉模拟评分中位数(IQR[范围]):Airtraq组8(7-9[0-10]),可弯曲支气管镜组8(7-9[0-10]),p=0.710。
- 患者舒适度视觉模拟评分中位数(IQR[范围]):Airtraq组8(6-9[2-10]),可弯曲支气管镜组8(7-9[3-10]),p=0.370。
关键数值与推荐
- 非劣效性界值为8%,Airtraq组成功率82%与可弯曲支气管镜组97%的差异未满足非劣效标准。
临床怎么用
- 在清醒气管插管中,可弯曲支气管镜仍可作为首选技术。
- Airtraq可视喉镜可根据个案情况考虑作为替代方案。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Videolaryngoscopy is a suitable alternative to flexible bronchoscopy to facilitate awake tracheal intubation. The relative effectiveness of these techniques in clinical practice is unknown. We compared flexible nasal bronchoscopy with Airtraq® videolaryngoscopy in patients with an anticipated difficult airway scheduled for awake tracheal intubation. Patients were allocated randomly to flexible nasal bronchoscopy or videolaryngoscopy. All procedures were performed with upper airway regional anaesthesia blockade and a target-controlled intravenous infusion of remifentanil. The success rate with the allocated technique was the primary outcome. A non-inferiority analysis with a predefined limit of 8% was planned. Seventy-eight patients were recruited, allocated randomly and analysed. The rate of successful intubation was 97% and 82% in the flexible bronchoscopy and videolaryngoscopy groups, respectively, p = 0.032. The median (IQR [range]) time to tracheal intubation was shorter with the Airtraq, 163 (105-332 [40-1004]) vs. 217 (180-364 [120-780]) s, p = 0.030. There were no significant differences for complications found between the groups. The median visual analogue scale for ease of intubation was 8 (7-9 [0-10]) for Airtraq vs. 8 (7-9 [0-10]) for flexible bronchoscopy, p = 0.710. The median visual analogue scale for patient comfort for Airtraq was 8 (6-9 [2-10]) vs. 8 (7-9 [3-10]) for flexible bronchoscopy, p = 0.370. The Airtraq videolaryngoscope is not non-inferior to flexible bronchoscopy for awake tracheal intubation in a clinical setting when awake tracheal intubation is indicted. It may be a suitable alternative when judged on a case-by-case basis.
引用
- Kamga H, Frugier A, Boutros M, et al. Flexible nasal bronchoscopy vs. Airtraq® videolaryngoscopy for awake tracheal intubation: a randomised controlled non-inferiority study. Anaesthesia. 2023. PMID: 37188387. doi:10.1111/anae.16042
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。