程序化间歇硬膜外推注与手动硬膜外推注在闭端多孔导管中的峰值流量率直接比较:实验研究
该实验室研究比较了程序化间歇硬膜外推注(PIEB)与手动硬膜外推注通过多孔硬膜外导管的流量。使用3、5、10和20 mL四种注射器容量,手动推注无论是否连接过滤器均被研究。结果显示手动推注的峰值流量率比CADD-Solis泵给药高数倍,3 mL时差异最大达12倍,5 mL和10 mL时最多7倍;不带过滤器的手动推注平均峰值流量率高于带过滤器者。
要点
- 手动硬膜外推注的峰值流量率显著高于PIEB泵给药,3 mL时差异最大达12倍。
- 硬膜外过滤器会降低手动推注产生的流量率。
- 高推注压力可能改善局麻药分布和镇痛效果,但需进一步临床研究。
核心结论
- 手动硬膜外推注产生的流量率高于CADD-Solis硬膜外泵(Smiths Medical)给药。
- 放置颗粒过滤器会降低推注时产生的流量率。
研究设计与样本
- 实验室研究,使用多孔硬膜外导管比较PIEB设置与手动推注的流量。
- 选用3、5、10和20 mL四种注射器容量,并研究手动推注有无硬膜外过滤器的情况。
- 采用广义估计方程分析比较组间数据。
主要发现
- 无论注射器大小,手动推注的流量比泵给药高数倍。
- 3 mL推注时峰值流量率差异最大达12倍,5 mL和10 mL时差异最多7倍。
- 不带过滤器的手动推注平均峰值流量率高于带过滤器的手动推注。
临床怎么用
- 手动推注可产生更高峰值流量率,可能用于PIEB效果不佳时的爆发痛处理。
- 使用过滤器会降低推注流量率,临床操作中需考虑此影响。
局限性
- 本研究为实验室研究,结论需 future 临床研究验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The programmable intermittent epidural bolus (PIEB) has been popularized as the optimal delivery technique for labor analgesia. Suggested advantages of this method are less local anesthetic consumption, improved maternal satisfaction, potentially shorter duration of labor, and decreased workload requirements for the anesthesia providers. However, a manual bolus is still routinely used for breakthrough pain when the PIEB is underperforming. METHODS: We conducted a laboratory-based study to quantify the flow through a multiorifice epidural catheter using the PIEB setting on an epidural pump compared to the manual epidural bolus. Four syringe volumes, 3, 5, 10, and 20 mL, were selected for this experiment. The flow in a manual bolus was also studied with and without the presence of an epidural catheter filter. A generalized estimating equation analysis was done to compare data between the groups. RESULTS: Regardless of the syringe size, there was a several-fold increase in flow when a manual bolus was used compared to a pump-administered dose, with the highest difference in the peak flow rate observed in 3-mL boluses with up to a 12-fold difference, while the difference was, at most, 7-fold in 5-mL and 10-mL boluses. Manual boluses without a filter achieve a mean peak flow rate higher than manual boluses with a filter. CONCLUSIONS: Our study found that manual boluses produced a higher flow rate compared to the CADD-Solis epidural pump (Smiths Medical). This study also found that the placement of a particulate filter reduces the flow rates generated while bolusing. Bulk flow rate is directly correlated with induced pressure and solution spread. Because higher bolus pressure has been shown to provide a more efficient distribution of local anesthetic and more efficient pain relief, these results may have impactful clinical significance and will pave the way for future studies.
引用
- Younger JD, Faryami A, Prasad M, et al. Direct Comparison of Peak Bulk Flow Rate of Programmable Intermittent Epidural Bolus and Manual Epidural Bolus Using a Closed-End Multiorifice Catheter: An Experimental Study. Anesth Analg. 2022. PMID: 36730916. doi:10.1213/ANE.0000000000006268
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。