竖脊肌平面阻滞两种入路的解剖学研究
该人体尸体研究比较超声引导下胸段竖脊肌平面(ESP)阻滞在横突内侧与横突间两个注射位点的染料扩散范围。结果显示横突间注射的扩散更广泛,硬膜外扩散层次更多(中位数5层),且硬膜内扩散更常见;横突内侧注射硬膜外扩散较少(中位数1层)。
要点
- 横突间注射的染料头尾扩散及硬膜外扩散范围大于横突内侧注射。
- 横突间注射硬膜外扩散中位数5层,横突内侧仅1层。
- 横突间注射有5例硬膜内扩散,横突内侧无硬膜内扩散报告。
核心结论
- 在人体尸体模型中,横突间注射的竖脊肌平面阻滞扩散范围比横突内侧注射更广泛。
研究设计与样本
- 超声引导下于T5横突内侧(MED组)或T4-T5横突间(BTWN组)注射20 mL 0.1%亚甲蓝。
主要发现
- 染料头尾扩散:MED组C4-T12,BTWN组C5-T11。
- 向外侧扩散至髂肋肌:MED组5例,BTWN组全部7例。
- 背根神经节和背根多数注射有染料,BTWN组更广泛。
- 腹根染色:MED组4例,BTWN组6例。
- 硬膜外扩散:BTWN组3-12层(中位数5层),MED组中位数1层(范围0-3层)。
临床怎么用
- 选择横突间入路时需考虑更广泛的硬膜外和硬膜内扩散风险。
- 横突内侧入路硬膜外扩散较少,但可能扩散不足。
局限性
- 尸体研究,染料扩散不能完全代表活体药液扩散。
- 样本量小,每组仅7例。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND AND OBJECTIVES: Studies show variable spread with thoracic erector spinae plane (ESP) injections. Injection sites vary from lateral end of the transverse process (TP) to 3 cm from the spinous process, with many not describing the precise site of injection. This human cadaveric study examined dye spread of ultrasound-guided thoracic ESP block at two needle locations. METHODS: Ultrasound-guided ESP blocks were performed on unembalmed cadavers. Methylene blue (20 mL, 0.1%) was injected in the ESP at the medial TP at level T5 (medial transverse process injection (MED), n=7) and the lateral end of the TP between T4 and T5 (injection between transverse processes (BTWN), n=7). The back muscles were dissected, and the cephalocaudal and medial-lateral dye spread documented. RESULTS: Dye spread cephalocaudally from C4-T12 in the MED group and C5-T11 in the BTWN group, and laterally to the iliocostalis muscle in five MED injections and all BTWN injections. One MED injection reached serratus anterior. Dorsal rami were dyed in five MED and all BTWN injections. Dye spread to the dorsal root ganglion and dorsal root in most injections, though more extensively in the BTWN group. The ventral root was dyed in 4 MED and 6 BTWN injections. Epidural spread in BTWN injections ranged from 3 to 12 levels (median: 5 levels), with contralateral spread in two cases and intrathecal spread in five injections. Epidural spread in MED injections was less extensive (median (range): 1 (0-3) levels); two MED injections did not enter the epidural space. CONCLUSION: An ESP injection administered between TPs exhibits more extensive spread than a medial TP injection in a human cadaveric model.
引用
- Harbell MW, Langley NR, Seamans DP, et al. Evaluating two approaches to the erector spinae plane block: an anatomical study. Reg Anesth Pain Med. 2023. PMID: 36797037. doi:10.1136/rapm-2022-104132
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。