文献速递小儿麻醉

婴幼儿及儿童鼻咽温探头最佳置入深度:前瞻性队列研究

Optimal Positioning of Nasopharyngeal Temperature Probes in Infants and Children: A Prospective Cohort Study.
PubMed 文献 · Anesth Analg · 2022-11-29

该前瞻性队列研究纳入157例非心脏手术气管插管患儿,按年龄分5组,以食管温为参考,通过一致性界限(LOA)<0.5°C定义鼻咽温探头最佳置入距离。结果显示最佳深度随年龄变化:≤6月6–10 cm,7–12月7–8 cm,13–23月7.5–12 cm,≥6岁10–12 cm;除2–5岁组9 cm处LOA为-0.67至0.52°C外,其余年龄组LOA均<0.5°C,最佳位置偏差≤0.1°C。结论强调鼻咽温探头需按年龄调整置入深度才能准确测量核心体温。

要点

  • 鼻咽温探头置入深度应随年龄调整,过浅或过深均可能低估核心体温。
  • 各年龄组最佳置入范围:≤6月6–10 cm,7–12月7–8 cm,13–23月7.5–12 cm,≥6岁10–12 cm。
  • 除2–5岁组9 cm处LOA略超0.5°C外,其余年龄组最佳位置与食管温偏差≤0.1°C。

核心结论

  • 鼻咽温探头仅在按年龄调整至合适置入深度时才能准确测量核心体温,最佳深度随年龄变化。

研究设计与样本

  • 前瞻性队列研究,纳入157例非心脏手术气管插管患儿,按年龄分5组:新生儿至6月、7–12月、13–23月、2–5岁、6–12岁。
  • 以按身高置入的食管温探头为参考,鼻咽温探头初始置入10 cm(新生儿及婴儿)、15 cm(1–5岁)、20 cm(≥6岁),随后按年龄每次退出1、2.5或2 cm,间隔5分钟共10次。
  • 以鼻咽温与食管温的一致性界限(LOA)<0.5°C定义最佳置入距离。

主要发现

  • 最佳鼻咽温探头位置:≤6月6–10 cm,7–12月7–8 cm,13–23月7.5–12 cm,≥6岁10–12 cm。
  • 除2–5岁组9 cm处95% LOA为-0.67°C至0.52°C外,其余年龄组95% LOA均<0.5°C。
  • 各年龄组最佳位置处偏差(鼻咽温与食管温平均差)≤0.1°C。

关键数值与推荐

  • 最佳置入距离以鼻咽温与食管温的95% LOA<0.5°C为标准。
  • ≤6月:6–10 cm;7–12月:7–8 cm;13–23月:7.5–12 cm;≥6岁:10–12 cm。
  • 2–5岁组在9 cm处95% LOA为-0.67°C至0.52°C,超出0.5°C标准。

临床怎么用

  • 临床置入鼻咽温探头时应按患儿年龄选择相应深度,避免统一深度导致核心体温低估。
  • 对2–5岁患儿需注意9 cm处一致性略差,应结合临床综合判断核心体温。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The nasopharynx is an easily accessible core-temperature monitoring site, but insufficient or excessive nasopharyngeal probe insertion can underestimate core temperature. Our goal was to estimate optimal nasopharyngeal probe insertion depth as a function of age. METHODS: We enrolled 157 pediatric patients who had noncardiac surgery with endotracheal intubation in 5 groups: (1) newborn to 6 months old, (2) infants 7 months to 1 year old, (3) children 13 to 23 months old, (4) children 2 to 5 years old, and (5) children 6 to 12 years old. A reference esophageal temperature probe was inserted at an appropriate depth based on each patient's height. A nasopharyngeal temperature probe was inserted from the naris at 10 cm in newborn and infants, 15 cm in children aged 1 to 5 years old, and 20 cm in children who were 6 years or older. The study nasopharyngeal probes were withdrawn 1, 2.5, or 2 cm (depending on age) 10 times at 5-minute intervals. Optimal probe insertion distances were defined by limits of agreement (LOAs) between nasopharyngeal and esophageal temperatures <0.5 °C. RESULTS: Optimal nasopharyngeal temperature probe position ranged from 6 to 10 cm in infants up to 6 months old, 7 to 8 cm in infants 7 to 12 months old, 7.5 to 12 cm in children 13 to 23 months old, and 10 to 12 cm in children aged 6 years and older. The 95% LOAs were <0.5 °C for all age categories except the 2- to 5-year-old group where the limits extended from -0.67 °C to 0.52 °C at 9 cm. At the optimal position within each age range, the bias (average nasopharyngeal-to-esophageal temperature difference) was ≤0.1 °C. CONCLUSIONS: Nasopharyngeal thermometers accurately measure core temperature, but only when probes are inserted a proper distance, which varies with age. As with much in pediatrics, nasopharyngeal thermometer insertion depths should be age appropriate.

来源:Anesth Analg,查看原文。版权归原作者所有。

引用

  1. Zhong JW, Sessler DI, Mao G, et al. Optimal Positioning of Nasopharyngeal Temperature Probes in Infants and Children: A Prospective Cohort Study. Anesth Analg. 2022. PMID: 36730063. doi:10.1213/ANE.0000000000006312

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。