清醒气管插管操作中气溶胶产生的定量评估

A quantitative evaluation of aerosol generation during awake tracheal intubation.
PubMed 文献 · Anaesthesia · 2023-01-29

研究在超净手术室中对健康受试者进行清醒气管插管和鼻内镜检查,连续监测气溶胶浓度。结果显示,支气管镜通过声带产生的气溶胶峰值与咳嗽相似,而利多卡因喷洒声带诱发的咳嗽气溶胶浓度显著高于自主咳嗽;鼻内镜操作气溶胶浓度较低,但部分操作峰值超过自主咳嗽。提示清醒上气道内镜操作可产生高浓度气溶胶,需采取空气传播感染控制措施。

要点

  • 支气管镜通过声带产生的气溶胶峰值与咳嗽相似。
  • 利多卡因喷洒声带诱发的咳嗽气溶胶浓度显著高于自主咳嗽。
  • 清醒上气道内镜操作应使用空气传播感染控制措施。

核心结论

  • 清醒气管插管和鼻内镜检查可产生高浓度呼吸道气溶胶。

研究设计与样本

  • 在超净手术室中对健康受试者进行清醒气管插管和鼻内镜检查,连续采样气溶胶,操作由12名麻醉受训者完成并接受。

主要发现

  • 支气管镜通过声带产生的气溶胶峰值中位数(IQR[范围])为1020(645-1245[120-48948])粒子/升,与咳嗽的1460(390-2506[40-12280])粒子/升相似,p=0.266。
  • 利多卡因喷洒声带诱发的咳嗽气溶胶浓度为91700(41907-166774[390-557817])粒子/升,显著高于自主咳嗽(p<0.001)。
  • 38次鼻内镜检查的气溶胶浓度中位数为180(120-525[0-9552])粒子/升,但5次鼻内镜操作的峰值气溶胶浓度超过自主咳嗽。

临床怎么用

  • 进行清醒上气道内镜操作(包括清醒气管插管、鼻内镜和支气管镜)时,若不能排除可吸入病原体,应采取空气传播感染控制措施。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Aerosol-generating procedures are medical interventions considered high risk for transmission of airborne pathogens. Tracheal intubation of anaesthetised patients is not high risk for aerosol generation; however, patients often perform respiratory manoeuvres during awake tracheal intubation which may generate aerosol. To assess the risk, we undertook aerosol monitoring during a series of awake tracheal intubations and nasendoscopies in healthy participants. Sampling was undertaken within an ultraclean operating theatre. Procedures were performed and received by 12 anaesthetic trainees. The upper airway was topically anaesthetised with lidocaine and participants were not sedated. An optical particle sizer continuously sampled aerosol. Passage of the bronchoscope through the vocal cords generated similar peak median (IQR [range]) aerosol concentrations to coughing, 1020 (645-1245 [120-48,948]) vs. 1460 (390-2506 [40-12,280]) particles.l-1 respectively, p = 0.266. Coughs evoked when lidocaine was sprayed on the vocal cords generated 91,700 (41,907-166,774 [390-557,817]) particles.l-1 which was significantly greater than volitional coughs (p < 0.001). For 38 nasendoscopies in 12 participants, the aerosol concentrations were relatively low, 180 (120-525 [0-9552]) particles.l-1 , however, five nasendoscopies generated peak aerosol concentrations greater than a volitional cough. Awake tracheal intubation and nasendoscopy can generate high concentrations of respiratory aerosol. Specific risks are associated with lidocaine spray of the larynx, instrumentation of the vocal cords, procedural coughing and deep breaths. Given the proximity of practitioners to patient-generated aerosol, airborne infection control precautions are appropriate when undertaking awake upper airway endoscopy (including awake tracheal intubation, nasendoscopy and bronchoscopy) if respirable pathogens cannot be confidently excluded.

来源:Anaesthesia,查看原文。版权归原作者所有。

引用

  1. Shrimpton AJ, O'Farrell G, Howes HM, et al. A quantitative evaluation of aerosol generation during awake tracheal intubation. Anaesthesia. 2023. PMID: 36710390. doi:10.1111/anae.15968

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。