使用实时传感器网络测量手术室及周边区域的气溶胶
该研究在五个手术室中部署低成本光电传感器,测量和分析人为生成气溶胶的稀释与泄漏,评估真实条件下的空气处理和扩散。研究发现不同手术室间差异显著,设备靠近通风口会影响气流和气溶胶消除模式,但手术室空气交换稳健,延长等待时间可能不必要,且手术室外邻近区域未见显著气溶胶浓度。
要点
- 手术室间气溶胶清除存在显著差异,设备摆放影响气流。
- 手术室空气交换稳健,延长手术间周转等待时间可能不必要。
- 手术室外邻近区域未检测到显著气溶胶浓度。
核心结论
- 手术室空气交换稳健,延长等待时间可能不必要,且气溶胶不会显著扩散至手术室外邻近区域。
研究设计与样本
- 在五个手术室手术间期部署低成本光电传感器,测量并分析人为生成气溶胶的稀释和泄漏。
主要发现
- 不同手术室间气溶胶稀释和消除存在显著差异。
- 设备 placement 靠近通风口会影响气流,改变气溶胶运动和消除模式。
- 手术室外邻近区域未观察到显著气溶胶浓度。
临床怎么用
- 气溶胶扩散模型和速率可预测手术室及其他临床区域的气溶胶持久性,有助于量化风险。
- 手术室空气处理稳健,可能无需延长手术间周转前的等待时间。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
The ability to measure and track aerosols in the vicinity of patients with suspected or confirmed COVID-19 is highly desirable. At present, there is no way to measure and track, in real time, the sizes, dispersion and dilution/disappearance of aerosols that are generated by airway manipulations such as mask ventilation; tracheal intubation; bronchoscopy; dental and gastro-intestinal endoscopy procedures; or by vigorous breathing, coughing or exercise. We deployed low-cost photoelectric sensors in five operating theatres between surgical cases. We measured and analysed dilution and exfiltration of aerosols we generated to evaluate air handling and dispersion under real-world conditions. These data were used to develop a model of aerosol persistence. We found significant variation between different operating theatres. Equipment placement near air vents affects air flows, impacting aerosol movement and elimination patterns. Despite these impediments, air exchange in operating theatres is robust and prolonged fallow time before theatre turnover may not be necessary. Significant concentrations of aerosols are not seen in adjoining areas outside of the operating theatre. These models and dispersion rates can predict aerosol persistence in operating theatres and other clinical areas and potentially facilitate quantification of risk, with obvious and far-reaching implications for designing, evaluating and confirming air handling in non-medical environments.
引用
- Hecker JG, He J, Rochlin R, et al. Measuring aerosols in the operating theatre and beyond using a real-time sensor network. Anaesthesia. 2022. PMID: 36047649. doi:10.1111/anae.15842
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。