Macintosh与Airtraq喉镜声门视野与插管力关系
本研究基于既往临床研究数据,对14例先后使用Macintosh和Airtraq喉镜插管的患者,比较插管力与声门视野关系的斜率。结论:Airtraq插管力对声门视野的依赖较Macintosh更小。
要点
- Macintosh喉镜下声门视野越差,插管力越大。
- Airtraq喉镜插管力对声门视野的依赖小于Macintosh。
- 该结论基于14例患者同体配对数据。
核心结论
- 与Macintosh相比,Airtraq喉镜插管力对声门视野的依赖更小。
研究设计与样本
- 基于既往临床研究数据,对14例先后使用Macintosh和Airtraq喉镜插管的患者进行同体配对分析。
- 采用最小二乘线性回归和稳健回归比较插管力与声门视野关系的斜率。
主要发现
- Macintosh插管力(N)与声门视野(%)关系的斜率为-0.679(SE 0.147)。
- Airtraq对应斜率为-0.076(SE 0.246)。
临床怎么用
- 在声门视野不佳时,Airtraq可能比Macintosh更少依赖增加插管力来改善视野。
局限性
- 样本量仅14例,来自既往临床研究数据。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Because intubation-mediated cervical spine and spinal cord injury are likely determined by intubation force magnitude, understanding the determinants of intubation force magnitude is clinically relevant. With direct (Macintosh) laryngoscopy, when glottic view is less favorable, anesthesiologists apply greater force. We hypothesized that, when compared with direct (Macintosh) laryngoscopy, intubation force with an optical indirect laryngoscope (Airtraq) would be less dependent on glottic visualization. METHODS: Using data obtained in a prior clinical study, we tested whether the slope of the intubation force versus glottic view relationship differed between intubations performed in 14 patients who were intubated twice, once with a Macintosh and once with an Airtraq videolaryngoscope. Slopes were compared using least-squares linear regression and robust regression. RESULTS: The slope of the intubation force (N) versus glottic view (%) relationship with the Macintosh (-0.679 [standard error {SE}, 0.147]) was significantly more negative than that of the Airtraq (-0.076 [SE, 0.246]). The least-squares regression difference in slopes was -0.603 (SE, 0.287); P = .046. The robust regression difference in slopes was -0.747 (SE, 0.187); P = .0005. Thus, when compared with the Macintosh, intubation force magnitude with Airtraq laryngoscopy was less dependent on glottic visualization. CONCLUSIONS: Previously, we reported that intubation force with the Airtraq was less in magnitude compared with the Macintosh. Our current study adds that intubation force also is less dependent on glottic view with Airtraq compared with the Macintosh.
引用
- Hindman BJ, Dexter F, Gadomski BC, et al. Relationship Between Glottic View and Intubation Force During Macintosh and Airtraq Laryngoscopy and Intubation. Anesth Analg. 2022. PMID: 35551148. doi:10.1213/ANE.0000000000006082
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。