便携式负压装置对手术室呼出气溶胶的影响
该实验使用多分散气溶胶发生器雾化生理盐水产生气溶胶,评估便携式负压装置在距离气溶胶发射源10 cm和30 cm处对激光粒子计数器检测的颗粒计数的影响。结果显示,对于0.5 µm、0.7 µm和1.0 µm颗粒浓度,气溶胶显著减少(p<0.001)。
要点
- 便携式负压装置可显著减少0.5、0.7、1.0 µm气溶胶颗粒浓度。
- 无负压装置时0.5 µm气溶胶中位浓度为3128 particles/ft3/min。
- 负压装置距离10 cm时0.5 µm浓度较背景为-0.5 particles/ft3/min。
核心结论
- 对于0.5 µm、0.7 µm和1.0 µm颗粒浓度,便携式负压装置可显著减少气溶胶(p<0.001)。
研究设计与样本
- 实验使用多分散气溶胶发生器雾化生理盐水产生气溶胶,便携式负压装置置于距离气溶胶发射源10 cm和30 cm处,用激光粒子计数器检测颗粒计数。
主要发现
- 无便携式负压装置时,检测到的0.5 µm气溶胶中位浓度为3128(1533, 22832)particles/ft3/min。
- 负压装置距离10 cm时,0.5 µm气溶胶中位浓度与背景浓度相比为-0.5(-8, 8)particles/ft3/min。
- 负压装置距离30 cm时,0.5 µm气溶胶中位浓度与背景浓度相比为398(89, 1749)particles/ft3/min。
临床怎么用
- 在区域麻醉下自主呼吸患者的手术中,可考虑使用便携式负压装置靠近患者气道以减少气溶胶暴露。
局限性
- 需进一步实验评估手术室人员面对潜在感染患者时的安全性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: Spontaneously breathing patients undergoing procedures under regional anesthesia can expose operating room personnel to infectious agents. The use of localized negative pressure within proximity of a patient's airway is expected to reduce the amount of bioaerosols dispersed particularly for anesthesia staff who are frequently near the patient's airway. METHODS: In the experiment, aerosols were produced using a polydisperse aerosol generator with nebulized saline. A portable negative pressure unit was set up at set distances of 10 cm and 30 cm with the aim of reducing aerosol particle counts detected by a laser-based particle counter. RESULTS: Without the portable negative pressure unit, the median concentration of 0.5 µm aerosols detected was 3128 (1533, 22832) particles/ft3/min. With the portable negative pressure unit 10 cm and 30 cm from the site of aerosol emittance, the median concentration compared with background concentration was -0.5 (-8, 8) particles/ft3/min and 398 (89, 1749) particles/ft3/min, respectively. CONCLUSIONS: For particle concentrations of 0.5 µm, 0.7 µm, and 1.0 µm a significant amount of aerosol reduction was observed (p<0.001). Further experiments are warranted to assess the safety of staff when encountering a potentially infectious patient in the operating room.
引用
- Popovic M, Beathe J, Gbaje E, et al. Effect of portable negative pressure units on expelled aerosols in the operating room environment. Reg Anesth Pain Med. 2022. PMID: 35365549. doi:10.1136/rapm-2022-103489
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。