FreeStyle Libre Pro 瞬感动态血糖监测系统与床旁毛细血管血糖检测在基础-餐时胰岛素方案治疗的住院2型糖尿病患者中的比较

Comparison of the FreeStyle Libre Pro Flash Continuous Glucose Monitoring (CGM) System and Point-of-Care Capillary Glucose Testing in Hospitalized Patients With Type 2 Diabetes Treated With Basal-Bolus Insulin Regimen.
PubMed 文献 · Diabetes Care · 2020-07-08

前瞻性研究比较住院2型糖尿病患者中FreeStyle Libre Pro动态血糖监测(CGM)与床旁毛细血管血糖检测(POC)的表现。POC日均血糖显著高于CGM(188.9±37.3 vs 176.1±46.9 mg/dL,平均差12.8 mg/dL,95%CI 8.3-17.2),POC检出<70 mg/dL低血糖的患者比例低于CGM(14% vs 56%),<54 mg/dL(4.1% vs 36%),均P<0.001。CGM总体平均绝对相对误差(MARD)为14.8%,低血糖范围内准确性较低。

要点

  • POC日均血糖显著高于CGM,平均差12.8 mg/dL(95%CI 8.3-17.2 mg/dL)。
  • POC检出<70 mg/dL低血糖患者比例显著低于CGM(14% vs 56%,P<0.001)。
  • CGM总体MARD为14.8%,51-69 mg/dL范围MARD为28.0%。

核心结论

  • 与POC相比,FreeStyle Libre CGM显示更低的日均血糖和更高的低血糖事件检出率,尤其是夜间和持续性低血糖,在住院2型糖尿病患者中。
  • CGM在低血糖范围内的准确性较低。

研究设计与样本

  • 前瞻性研究,纳入入住普通内科和外科病房的成年2型糖尿病患者。
  • 患者住院期间接受POC(餐前和睡前)和CGM监测。

主要发现

  • POC日均血糖显著高于CGM:188.9±37.3 vs 176.1±46.9 mg/dL,估计平均差12.8 mg/dL(95%CI 8.3-17.2 mg/dL)。
  • POC检出血糖<70 mg/dL的患者比例为14%,CGM为56%;POC检出<54 mg/dL为4.1%,CGM为36%(均P<0.001)。
  • CGM夜间低血糖<54 mg/dL为26%,持续性低血糖<54 mg/dL为12%。
  • 总体MARD为14.8%,血糖70-250 mg/dL范围内MARD为11.4%-16.7%,51-69 mg/dL范围MARD为28.0%。
  • 血糖读数在±15%/15 mg/dL、±20%/20 mg/dL、±30%/30 mg/dL范围内的比例分别为62%、76%、91%;误差网格分析显示98.8%的血糖配对位于A区和B区。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: We compared the performance of the FreeStyle Libre Pro continuous glucose monitoring (CGM) and point-of-care capillary glucose testing (POC) among insulin-treated hospitalized patients with type 2 diabetes (T2D). RESEARCH DESIGN AND METHODS: This was a prospective study in adult patients with T2D admitted to general medicine and surgery wards. Patients were monitored with POC before meals and bedtime and with CGM during the hospital stay. Study end points included differences between POC and CGM in mean daily blood glucose (BG), hypoglycemia <70 and <54 mg/dL, and nocturnal hypoglycemia. We also calculated the mean absolute relative difference (MARD), ±15%/15 mg/dL, ±20%/20 mg/dL, and ±30%/30 mg/dL and error grid analysis between matched glucose pairs. RESULTS: Mean daily glucose was significantly higher by POC (188.9 ± 37.3 vs. 176.1 ± 46.9 mg/dL) with an estimated mean difference of 12.8 mg/dL (95% CI 8.3-17.2 mg/dL), and proportions of patients with glucose readings <70 mg/dL (14% vs. 56%) and <54 mg/dL (4.1% vs. 36%) detected by POC BG were significantly lower compared with CGM (all P < 0.001). Nocturnal and prolonged CGM hypoglycemia <54 mg/dL were 26% and 12%, respectively. The overall MARD was 14.8%, ranging between 11.4% and 16.7% for glucose values between 70 and 250 mg/dL and higher for 51-69 mg/dL (MARD 28.0%). The percentages of glucose readings within ±15%/15 mg/dL, ±20%/20 mg/dL, and ±30%/30 mg/dL were 62%, 76%, and 91%, respectively. Error grid analysis showed 98.8% of glucose pairs within zones A and B. CONCLUSIONS: Compared with POC, FreeStyle Libre CGM showed lower mean daily glucose and higher detection of hypoglycemic events, particularly nocturnal and prolonged hypoglycemia in hospitalized patients with T2D. CGM's accuracy was lower in the hypoglycemic range.

来源:Diabetes Care,查看原文。版权归原作者所有。

引用

  1. Galindo RJ, Migdal AL, Davis GM, et al. Comparison of the FreeStyle Libre Pro Flash Continuous Glucose Monitoring (CGM) System and Point-of-Care Capillary Glucose Testing in Hospitalized Patients With Type 2 Diabetes Treated With Basal-Bolus Insulin Regimen. Diabetes Care. 2020. PMID: 32641372. doi:10.2337/dc19-2073

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。