开发临床原型PTeye用于术中实时无标记甲状旁腺识别
研究将实验室搭建的近红外自发荧光(NIRAF)光谱系统转化为临床原型PTeye,并在甲状腺切除术/甲状旁腺切除术患者中评估。实验室系统在162例患者中准确率92.5%,PTeye在35例患者中准确率96.1%,且PTeye可在手术室照明下工作,仅需实验室系统25%的激光功率。
要点
- PTeye原型在术中识别甲状旁腺的准确率达96.1%。
- PTeye可在环境光下工作,激光功率仅为实验室系统的25%。
- 血液不影响NIRAF测量,出血术野不影响甲状旁腺识别。
核心结论
- 临床原型PTeye在无标记术中甲状旁腺识别中表现出高准确率。
- PTeye界面直观,可在环境光下指导识别甲状旁腺组织,是可靠易用的手术工具。
研究设计与样本
- 实验室搭建的光谱系统在162例接受甲状腺切除术和/或甲状旁腺切除术的患者中评估。
- 基于结果设计临床原型PTeye,并在35例患者中研究。
- 实验室系统与PTeye在每组20例患者中由单一用户同时比较。
主要发现
- 实验室系统引导甲状旁腺识别的总体准确率为92.5%。
- PTeye系统的准确率为96.1%。
- PTeye可在手术室灯保持开启时指导甲状旁腺识别,且仅需实验室设置所用激光功率的25%。
- 无论甲状腺或甲状旁腺病理如何,甲状旁腺的NIRAF强度均高于甲状腺和其他颈部组织。
- 血液似乎不影响两种系统获得的组织NIRAF测量。
临床怎么用
- PTeye可作为术中实时无标记识别甲状旁腺的辅助工具,尤其在环境光下和低激光功率条件下。
- 出血术野不影响NIRAF测量,PTeye可能适用于存在出血的手术场景。
局限性
- 比较研究每组仅20例患者,样本量有限。
- 研究为单中心、单一用户操作比较,可能影响普适性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Patients undergoing thyroidectomy may have inadvertent damage or removal of the parathyroid gland(s) due to difficulty in real-time parathyroid identification. Near-infrared autofluorescence (NIRAF) has been demonstrated as a label-free modality for intraoperative parathyroid identification with high accuracy. This study presents the translation of that approach into a user-friendly clinical prototype for rapid intraoperative guidance in parathyroid identification. METHODS: A laboratory (lab)-built spectroscopy system that measures NIRAF in tissue was evaluated for identifying parathyroid glands in vivo across 162 patients undergoing thyroidectomy and/or parathyroidectomy. Based on these results, a clinical prototype called PTeye was designed with a user-friendly interface and subsequently investigated in 35 patients. The performance of the lab-built system and the clinical prototype were concurrently compared side by side by a single user with 20 patients in each group. The influence of (i) intrapatient and interpatient variability of NIRAF in thyroid and parathyroid glands and (ii) thyroid and parathyroid pathology on intraoperative parathyroid identification were investigated. The effect of blood on NIRAF intensity of parathyroid and thyroid was tested ex vivo with the PTeye system to assess if a hemorrhagic surgical field would affect parathyroid identification. Accuracy of both systems were determined by correlating the acquired data with either visual confirmation by a surgeon for unexcised parathyroid glands or histology reports for excised parathyroid glands. RESULTS: The overall accuracy of the lab-built system in guiding parathyroid identification was 92.5%, while the PTeye system achieved an accuracy of 96.1%. Unlike the lab-built system, the PTeye could guide parathyroid identification even as the operating room lights remained on and required only 25% of the laser power used by the lab-built setup. Parathyroid glands had elevated NIRAF intensity compared to thyroid and other neck tissues, regardless of thyroid or parathyroid pathology. Blood did not seem to affect tissue NIRAF measurements obtained with both systems. CONCLUSION: In this study, the clinical prototype PTeye demonstrated high accuracy for label-free intraoperative parathyroid identification. The intuitive interface of the PTeye that can guide in identifying parathyroid tissue in the presence of ambient room lights suggests that it is a reliable and easy-to-use tool for surgical personnel.
引用
- Thomas G, McWade MA, Paras C, et al. Developing a Clinical Prototype to Guide Surgeons for Intraoperative Label-Free Identification of Parathyroid Glands in Real Time. Thyroid. 2018. PMID: 30084742. doi:10.1089/thy.2017.0716