连续血糖监测仪在住院儿童糖尿病管理中的准确性
该研究回顾性分析了Dexcom G6连续血糖监测系统在住院儿童糖尿病患者中的准确性,共纳入277次住院(来自202名2-18岁患者)的2904对CGM与床旁血糖配对数据,平均绝对相对差异为15.9%,96.6%的值落在Clarke误差网格分析的A区和B区。血清pH、钠和血尿素氮对CGM值无影响。作者认为CGM在住院儿童糖尿病患者中表现出可接受的准确性,应将CGM数据整合到医院电子记录中以优化管理。
要点
- 住院儿童糖尿病患者中Dexcom G6 CGM的MARD为15.9%,96.6%的值位于CEG分析A区和B区。
- 约62%的配对值落在15%/15 mg/dL范围内,74%在20%/20范围内,88%在30%/30范围内。
- 血清pH、钠和血尿素氮对CGM值或绝对相对差异无影响。
核心结论
- CGM在住院儿童糖尿病患者中表现出可接受的准确性,应将CGM数据整合到医院电子记录中以优化管理。
研究设计与样本
- 回顾性审查了2018年3月至2023年9月期间入住学术儿童医院的儿童糖尿病患者的Dexcom G6 CGM系统准确性。
- 通过交叉引用Dexcom Clarity数据库与内部住院数据库,识别所有有CGM数据的住院儿童。
- 从Clarity记录传感器葡萄糖读数,从电子病历记录床旁血糖、血尿素氮和pH值。
- 采用平均绝对相对差异和Clarke误差网格分析评估CGM准确性和临床可靠性。
主要发现
- 该期间共有3200次儿童糖尿病住院,其中277次(来自202名2-18岁患者)有相关CGM数据。
- 比较了2904对CGM与床旁血糖测量值,MARD为15.9%,96.6%的值位于CEG分析的A区和B区。
- 约62%的配对值落在15%/15 mg/dL范围内(取较大者),74%在20%/20范围内,88%在30%/30范围内。
- 线性回归分析显示血清pH、钠和血尿素氮对CGM值或绝对相对差异无影响。
临床怎么用
- 作者认为CGM数据应整合到医院电子记录中以优化管理。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: The adoption of continuous glucose monitors (CGMs) in inpatient settings in the pediatric population has been slow because of a scarcity of data on their reliability in hospitalized children. RESEARCH DESIGN AND METHODS: We retrospectively reviewed the accuracy of the Dexcom G6 CGM system in pediatric patients with diabetes admitted to our academic children's hospital from March 2018 to September 2023. We cross-referenced the Dexcom Clarity database against an internal database of inpatient admissions to identify all children with CGM data admitted to the hospital. We recorded sensor glucose readings from Clarity and values for point-of-care (POC) glucose, blood urea nitrogen (BUN), and pH from the electronic medical record. CGM accuracy and clinical reliability were measured by mean absolute relative difference (MARD) and Clarke error grid (CEG) analyses. RESULTS: There were 3,200 admissions of children with diabetes in this period, of which 277 (from 202 patients age 2-18 years) had associated CGM data. Paired CGM and POC measurements (n = 2,904) were compared, resulting in an MARD of 15.9%, with 96.6% of the values in zones A and B of the CEG analysis. Approximately 62% of paired values fell within a 15% or 15 mg/dL difference, whichever was larger (15%/15 mg/dL range), 74% within 20%/20, and 88% within 30%/30. Serum pH, sodium, and BUN had no impact on CGM values or absolute relative difference in linear regression analysis. CONCLUSIONS: CGMs demonstrated acceptable accuracy in hospitalized children with diabetes. CGM data should be integrated into hospital electronic records to optimize management.