在模拟家庭采样条件下儿童唾液类固醇的采集
研究比较三种唾液采集方法(Salivette皮质醇、被动流涎、SalivaBio)所测唾液糖皮质激素浓度,并评估其在模拟邮寄条件下的稳定性;同时在30名6岁以下儿童中评估SalivaBio和SaliPac的采集时间与照护者接受度。结果显示三种方法所测唾液皮质醇和可的松浓度无差异,在室温和4°C下稳定72小时,SaliPac耐受性良好且能在4分钟内采集足够唾液。
要点
- Salivette、被动流涎和SalivaBio三种方法采集的唾液皮质醇和可的松浓度无差异。
- 唾液皮质醇和可的松在室温及4°C下稳定72小时,反复冻融未致显著降解。
- 6岁以下儿童中SalivaBio和SaliPac耐受良好,4分钟内可采集足够唾液。
核心结论
- Salivette、被动流涎和SalivaBio采集的唾液皮质醇和可的松浓度相当,在模拟家庭采集条件下稳定。
- SaliPac是年幼儿童可接受的唾液采样装置。
研究设计与样本
- 六名健康成人使用Salivette Cortisol、被动流涎和SalivaBio在夜间、醒时和下午3点采集唾液样本,持续五天。
- 30名6岁以下儿童中评估SalivaBio和SaliPac采集1 mL唾液的时间及照护者接受度。
- 唾液样本在4°C、室温和50°C下储存24、48、72小时和1周以模拟邮寄条件,用质谱法测定皮质醇和可的松浓度。
主要发现
- 三种采集方法所测唾液糖皮质激素浓度无差异。
- 唾液皮质醇和可的松在室温和4°C下稳定72小时,反复冻融循环未导致显著降解。
- 6岁以下儿童中SalivaBio和SaliPac耐受良好,采集足够唾液用于唾液类固醇分析的时间少于4分钟。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
CONTEXT: Measurement of salivary glucocorticoids is an accepted method for testing adrenal function but there are few data on stability during home collection. Current salivary collection techniques require active participation or present a choking hazard and are unsuitable for young children. OBJECTIVE: We sought to compare different salivary collection methods; assess the stability of salivary glucocorticoids under conditions replicating home collection; and assess patient tolerability and caregiver acceptability of a salivary collection device for young children, a swab encased in an infant pacifier (SaliPac). METHODS: Six healthy adults collected salivary samples using a Salivette Cortisol, passive drool, and SalivaBio at night, waking, and 3 Pm for five days. Time to collect 1-mL saliva using the SalivaBio and SaliPac and caregiver acceptability were assessed in 30 children younger than 6 years. Saliva was stored at 4 °C, room temperature (RT), and 50 °C for 24, 48, 72 hours and 1 week to replicate potential postage conditions. Salivary cortisol and cortisone concentrations were measured by mass spectrometry. RESULTS: There was no difference in salivary glucocorticoid concentrations using the 3 collection methods. Salivary cortisol and cortisone were stable for 72 hours at RT and 4 °C, and repeated freeze-thaw cycles did not cause significant degradation. In children younger than 6 years the SalivaBio and SaliPac were well tolerated and collected sufficient saliva for salivary steroid analysis in less than 4 minutes. CONCLUSION: Salivette, passive drool, and SalivaBio collect samples with comparable salivary cortisol and cortisone concentrations, which are stable under conditions replicating home collection. SaliPac is an acceptable device for salivary sampling in young children.
引用
- Tonge JJ, Keevil BG, Craig JN, et al. Salivary Steroid Collection in Children Under Conditions Replicating Home Sampling. J Clin Endocrinol Metab. 2022. PMID: 35961299. doi:10.1210/clinem/dgac419
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。