文献速递复苏、休克与脓毒症

大型异质性心脏骤停数据集中心肺复苏质量分析的方法与框架

Methodology and framework for the analysis of cardiopulmonary resuscitation quality in large and heterogeneous cardiac arrest datasets.
PubMed 文献 · Resuscitation · 2021-09-10
一句话结论自动化方法可从多厂商除颤器文件与临床标注中统一提取CPR质量指标,全时段中位误差CCF<2%、CCR<1 min⁻¹、中断比<1.5%。

本研究分析2236例OHCA病例的多厂商除颤器数据,利用胸阻抗与按压深度信号自动识别胸外按压,并以设备时间戳和临床标注界定分析区间及自主循环恢复时段。以人工审核为金标准,自动计算的CCF、CCR和中断比全时段中位误差分别<2%、<1 min⁻¹和<1.5%,大误差(CCF和中断比>10%、CCR>10 min⁻¹)病例比例<10%,且较短子区间(如气道置入间期)误差更低。该方法为跨来源CPR质量数据的汇总分析提供了验证工具。

要点

  • 对2236例多厂商除颤器OHCA数据,自动提取的CCF全时段中位误差<2%,CCR<1 min⁻¹,中断比<1.5%。
  • CCF或中断比误差>10%、CCR误差>10 min⁻¹的大误差病例比例低于10%。
  • 较短分析子区间(如气道置入间期)的误差低于全时段分析。
  • 自动识别胸外按压基于胸阻抗与按压深度信号,分析区间由设备时间戳和临床标注设定。
  • 该方法支持多来源除颤器文件与临床标注的聚合与CPR质量分析。

适用人群与场景

  • 多厂商除颤器记录的院外心脏骤停(OHCA)病例,共2236例;不涉及急诊/ICU院内场景或院前特定病情阶段。

诊断标准与关键阈值

  • 原文未提及

处置流程与药物剂量

  • 原文未提及

红旗征与不能漏的情况

  • 原文未提及

研究设计与关键结果

  • 方法学验证研究:2236例OHCA病例、多除颤器型号与制造商;自动提取CPR质量指标以人工审核为金标准;全时段中位误差CCF<2%、CCR<1 min⁻¹、中断比<1.5%;大误差(CCF和中断比>10%、CCR>10 min⁻¹)病例比例<10%。

证据类型、等级与局限

  • 方法学验证研究(非RCT/队列/病例对照),以人工审核为金标准;局限:原文未明确报告证据等级,数据来自多厂商除颤器文件,未提及外部验证或前瞻性应用。

对床旁实践的改变

  • 原文未提及

指南或共识出处

  • 原文未提及

以上提炼依据论文摘要(全文未开放获取,或许可不允许翻译转载);全文请点「阅读原文」。

英文原文(摘要)· 点击展开

BACKGROUND: Out-of-hospital cardiac arrest (OHCA) data debriefing and clinical research often require the retrospective analysis of large datasets containing defibrillator files from different vendors and clinical annotations by the emergency medical services.

AIM: To introduce and evaluate a methodology to automatically extract cardiopulmonary resuscitation (CPR) quality data in a uniform and systematic way from OHCA datasets from multiple heterogeneous sources.

METHODS: A dataset of 2236 OHCA cases from multiple defibrillator models and manufacturers was analyzed. Chest compressions were automatically identified using the thoracic impedance and compression depth signals. Device event time-stamps and clinical annotations were used to set the start and end of the analysis interval, and to identify periods with spontaneous circulation. A manual audit of the automatic annotations was conducted and used as gold standard. Chest compression fraction (CCF), rate (CCR) and interruption ratio were computed as CPR quality variables. The unsigned error between the automated procedure and the gold standard was calculated.

RESULTS: Full-episode median errors below 2% in CCF, 1 min-1 in CCR, and 1.5% in interruption ratio, were measured for all signals and devices. The proportion of cases with large errors (>10% in CCF and interruption ratio, and >10 min-1 in CCR) was below 10%. Errors were lower for shorter sub-intervals of interest, like the airway insertion interval.

CONCLUSIONS: An automated methodology was validated to accurately compute CPR metrics in large and heterogeneous OHCA datasets. Automated processing of defibrillator files and the associated clinical annotations enables the aggregation and analysis of CPR data from multiple sources.

来源:Resuscitation,查看原文。版权归原作者所有。

参考来源

  1. Jaureguibeitia X, Aramendi E, Irusta U, et al. Methodology and framework for the analysis of cardiopulmonary resuscitation quality in large and heterogeneous cardiac arrest datasets. Resuscitation. 2021. PMID: 34509553. doi:10.1016/j.resuscitation.2021.09.005

供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。