文献速递复苏、休克与脓毒症

两种机械胸外按压装置复苏后标准化损伤评估:一项前瞻性随机大型动物试验

Standardized post-resuscitation damage assessment of two mechanical chest compression devices: a prospective randomized large animal trial.
PubMed 文献 · Scand J Trauma Resusc Emerg Med · 2021-06-05
一句话结论猪室颤模型中,LUCAS 2 装置较 Corpuls cpr 复苏成功率更高、胸部损伤更轻,提示不同机械按压装置可能显著影响复苏研究结局。

该研究在18头雄性德国长白猪中诱发心室颤动,随机分为 LUCAS™ 2(mCCD1)组与 Corpuls™ cpr(mCCD2)组行机械心肺复苏,高级生命支持最多10个循环,恢复自主循环(ROSC)者监测8小时。mCCD1组5头、mCCD2组1头实现ROSC(p=0.048),仅mCCD1组动物存活至监测结束(p<0.01)。mCCD1组复苏期间肺内分流更少(p=0.025)、正常通气血流比更高(p=0.017);mCCD2组胸部损伤显著更重(p=0.046)。作者认为不同机械按压装置可能显著影响复苏研究结局及试验间可比性。

要点

  • 猪室颤模型中,mCCD1(LUCAS™ 2)组5头 vs mCCD2(Corpuls™ cpr)组1头实现ROSC(p=0.048)。
  • 仅mCCD1组动物存活至8小时监测结束(p<0.01)。
  • mCCD1组复苏期间肺内分流更少(p=0.025)、正常通气血流比更高(p=0.017)。
  • mCCD2组胸部损伤评分显著更重(p=0.046)。
  • 作者认为不同机械按压装置可能显著影响复苏研究结局,并建议采用标准化复苏后损伤评估。

适用人群与场景

  • 18头雄性德国长白猪,麻醉并置管后诱发心室颤动,随机分入两组行机械心肺复苏。

诊断标准与关键阈值

  • 以心室颤动作为心脏骤停模型,机械心肺复苏启动,高级生命支持最多10个循环;实现ROSC者监测8小时。

处置流程与药物剂量

  • 机械心肺复苏分别采用 LUCAS™ 2(mCCD1)或 Corpuls™ cpr(mCCD2)装置。
  • 高级生命支持最多10个循环;实现ROSC的动物监测8小时。
  • 复苏期间进行通气血流比测量与血气分析;胸部损伤采用标准化损伤评分评估。

研究设计与关键结果

  • 研究类型:前瞻性随机大型动物试验;18头雄性德国长白猪。
  • mCCD1组5头 vs mCCD2组1头实现ROSC(p=0.048)。
  • 仅mCCD1组动物存活至监测结束(p<0.01)。
  • mCCD1组复苏期间肺内分流更少(p=0.025)、正常通气血流比更高(p=0.017)。
  • mCCD2组胸部损伤显著更重(p=0.046)。

证据类型、等级与局限

  • 证据类型:前瞻性随机大型动物试验(猪)。
  • 局限:作者指出需受控的人体与动物心肺复苏研究及标准化复苏后损伤评估来确认潜在危害。

对床旁实践的改变

  • 作者认为,在幼猪实验性心肺复苏中,LUCAS 2装置较Corpuls cpr复苏结局更优、胸部损伤更少。
  • 作者提醒,实验研究中不同机械按压装置可能显著影响复苏研究结局及不同试验间的可比性。
  • 作者建议,受控的人体与动物心肺复苏研究及标准化复苏后损伤评估有助于确认潜在危害。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: Mechanical chest compression devices are accepted alternatives for cardiopulmonary resuscitation (CPR) under specific circumstances. Current devices lack prospective and comparative data on their specific cardiovascular effects and potential for severe thoracic injuries.

OBJECTIVES: To compare CPR effectiveness and thoracic injuries of two mechanical chest compression devices in pigs.

STUDY DESIGN: Prospective randomised trial.

ANIMALS: Eighteen male German landrace pigs.

METHODS: Ventricular fibrillation was induced in anaesthetised and instrumented pigs and the animals were randomised into two intervention groups. Mechanical CPR was initiated by means of LUCAS™ 2 (mCCD1) or Corpuls™ cpr (mCCD2) device. Advanced life support was applied for a maximum of 10 cycles and animals achieving ROSC were monitored for 8 h. Ventilation/perfusion measurements were performed and blood gas analyses were taken. Thoracic injuries were assessed via a standardised damage score.

RESULTS: Five animals of the mCCD1 group and one animal of the mCCD2 group achieved ROSC (p = 0.048). Only the mCCD1 animals survived until the end of the monitoring period (p < 0.01). MCCD1 animals showed less pulmonary shunt (p = 0.025) and higher normal V/Q (p = 0.017) during CPR. MCCD2 animals showed significantly more severe thoracic injuries (p = 0.046).

CONCLUSION: The LUCAS 2 device shows superior resuscitation outcomes and less thoracic injuries compared to Corpuls cpr when used for experimental CPR in juvenile pigs. Researchers should be aware that different mCCDs for experimental studies may significantly influence the respective outcome of resuscitation studies and affect comparability of different trials. Controlled human and animal CPR studies and a standardised post-resuscitation injury evaluation could help to confirm potential hazards.

TRIAL REGISTRATION: Trial approval number: G16-1-042-E4.

来源:Scand J Trauma Resusc Emerg Med,查看原文。版权归原作者所有。

参考来源

  1. Ruemmler R, Stein J, Duenges B, et al. Standardized post-resuscitation damage assessment of two mechanical chest compression devices: a prospective randomized large animal trial. Scand J Trauma Resusc Emerg Med. 2021. PMID: 34090500. doi:10.1186/s13049-021-00892-4

供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。