无创逐搏每搏量测量在猪模型迷你液体挑战中判定前负荷反应性的初步研究

Noninvasive Beat-To-Beat Stroke Volume Measurements to Determine Preload Responsiveness During Mini-Fluid Challenge in a Swine Model: A Preliminary Study.
PubMed 文献 · Shock · 2021-01-01
一句话结论猪模型初步研究显示,电阻抗断层成像(EIT)可无创逐搏追踪迷你液体挑战中的每搏量变化并判定前负荷反应性,尚需进一步动物与临床研究验证。

该初步动物实验在5只麻醉机械通气的猪中,先移除每只动物总血量的25%至30%制造低血容量,再进行多次液体注射。EIT设备成功逐搏追踪了从低血容量、前负荷反应性到目标容量和容量过负荷不同容量状态下的每搏量变化。5只猪共50次100 mL液体注射(每只猪10次)中,前负荷反应性状态下前负荷反应性数值高达32.3%,而容量过负荷状态下低至-4.9%。EIT与有创脉搏轮廓分析(PCA)测得的每搏量偏倚为0 mL,一致性界限为±3.6 mL(范围17.6 mL至51.0 mL)。作者认为EIT能够在迷你液体挑战中测量逐搏每搏量变化并判定前负荷反应性,但需要进一步的动物和临床研究来证明EIT方法作为液体管理新工具的可行性。

要点

  • 在5只猪中,EIT成功逐搏追踪从低血容量、前负荷反应性到目标容量和容量过负荷不同容量状态下的每搏量变化。
  • 5只猪共50次100 mL液体注射(每只猪10次),前负荷反应性状态下前负荷反应性数值高达32.3%,容量过负荷状态下低至-4.9%。
  • EIT与有创脉搏轮廓分析(PCA)测得的每搏量偏倚为0 mL,一致性界限为±3.6 mL,范围17.6 mL至51.0 mL。
  • 作者认为EIT能够在迷你液体挑战中测量逐搏每搏量变化并判定前负荷反应性,但需要进一步动物和临床研究证明其作为液体管理新工具的可行性。

适用人群与场景

  • 5只麻醉并机械通气的猪,先移除每只动物总血量的25%至30%制造低血容量,再进行多次液体注射。

处置流程与药物剂量

  • 每只猪共进行10次100 mL液体注射,5只猪合计50次100 mL液体注射。

研究设计与关键结果

  • 研究类型为动物实验,5只猪麻醉机械通气,移除总血量的25%至30%后进行多次液体注射,共50次100 mL液体注射(每只猪10次)。
  • EIT设备成功逐搏追踪从低血容量、前负荷反应性到目标容量和容量过负荷不同容量状态下的每搏量变化。
  • 前负荷反应性状态下前负荷反应性数值高达32.3%,容量过负荷状态下低至-4.9%。
  • EIT与有创脉搏轮廓分析(PCA)测得的每搏量偏倚为0 mL,一致性界限为±3.6 mL,范围17.6 mL至51.0 mL。

证据类型、等级与局限

  • 证据类型:动物实验(初步研究)。
  • 局限:作者指出需要进一步的动物和临床研究来证明EIT方法作为液体管理新工具的可行性。

对床旁实践的改变

  • 作者认为EIT能够在迷你液体挑战中测量逐搏每搏量变化并判定前负荷反应性。
  • 作者指出需要进一步的动物和临床研究来证明EIT方法作为液体管理新工具的可行性。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

Cardiac output (CO) is an important parameter in fluid management decisions for treating hemodynamically unstable patients in intensive care unit. The gold standard for CO measurements is the thermodilution method, which is an invasive procedure with intermittent results. Recently, electrical impedance tomography (EIT) has emerged as a new method for noninvasive measurements of stroke volume (SV). The objectives of this paper are to compare EIT with an invasive pulse contour analysis (PCA) method in measuring SV during mini-fluid challenge in animals and determine preload responsiveness with EIT. Five pigs were anesthetized and mechanically ventilated. After removing 25% to 30% of the total blood from each animal, multiple fluid injections were conducted. The EIT device successfully tracked changes in SV beat-to-beat during varying volume states, i.e., from hypovolemia and preload responsiveness to target volume and volume overload. From a total of 50 100-mL fluid injections on five pigs (10 injections per pig), the preload responsiveness value was as large as 32.3% in the preload responsiveness state while in the volume overload state it was as low as -4.9%. The bias of the measured SV data using EIT and PCA was 0 mL, and the limits of agreement were ±3.6 mL in the range of 17.6 mL to 51.0 mL. The results of the animal experiments suggested that EIT is capable of measuring beat-to-beat SV changes during mini-fluid challenge and determine preload responsiveness. Further animal and clinical studies will be needed to demonstrate the feasibility of the EIT method as a new tool for fluid management.

来源:Shock,查看原文。版权归原作者所有。

参考来源

  1. Ko RE, Jang GY, Chung CR, et al. Noninvasive Beat-To-Beat Stroke Volume Measurements to Determine Preload Responsiveness During Mini-Fluid Challenge in a Swine Model: A Preliminary Study. Shock. 2021. PMID: 33534400. doi:10.1097/SHK.0000000000001739

供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。