文献速递复苏、休克与脓毒症心脑血管急症

Analyze Whilst Compressing算法在CPR中检测心室颤动的性能评估:自动体外除颤器的比较研究

Analyze Whilst Compressing algorithm for detection of ventricular fibrillation during CPR: A comparative performance evaluation for automated external defibrillators.
PubMed 文献 · Resuscitation · 2021-01-30
一句话结论AWC算法在胸外按压中分析心律,对不可除颤心律避免中断按压,对VF患者缩短电击前暂停,但需34.4%的病例进行按压中断再确认。

本研究提出用于AED的Analyze Whilst Compressing(AWC)算法组合,旨在减少电击前暂停和早期停止胸外按压以治疗再颤动。研究纳入2916例OHCA患者的16057条ECG条带,按患者分为训练集和验证集。验证结果显示,标准分析阶段VF敏感性98.3%、不可除颤心律特异性>99%;AWC阶段第二步再确认后敏感性92.1%、特异性>99%。AWC对66-83%的不可除颤心律不中断按压,为92.1%的VF患者最小化电击前暂停,但34.4%的病例需要中断按压再确认。

要点

  • AWC算法在胸外按压存在时进行第一步分析,若检测到可除颤心律则需在无按压下进行第二步再确认(分析时长5秒)。
  • 验证集中,标准分析阶段(按压中断5秒后分析)VF敏感性98.3%、不可除颤心律特异性>99%。
  • AWC阶段第二步再确认后VF敏感性92.1%、特异性>99%,性能与其它AED算法相似,符合标准推荐目标。
  • AWC对66-83%的不可除颤心律不中断胸外按压,为92.1%的VF患者最小化电击前暂停。
  • AWC需中断按压再确认的比例为34.4%,再确认是其它报告算法(25.7-100%)的常见局限。

适用人群与场景

  • 接受AED治疗的院外心脏骤停(OHCA)患者,数据来自2916例患者。

诊断标准与关键阈值

  • 原文未提及诊断标准的具体阈值;仅描述AWC算法步骤:第一步在按压存在时分析,若检测到可除颤心律则第二步在无按压下再确认(分析时长5秒)。

处置流程与药物剂量

  • 第一步:在胸外按压存在时运行AWC Step1分析心律。
  • 第二步:若Step1检测到可除颤心律,则中断按压进行Step2再确认,分析时长5秒。
  • 标准分析阶段:每两分钟中断按压后运行5秒的传统电击建议分析。

红旗征与不能漏的情况

  • 原文未提及

研究设计与关键结果

  • 研究类型为算法性能评估,使用16057条ECG条带来自2916例OHCA患者,按患者分为训练集(8559条,1604例)和验证集(7498条,1312例)。
  • 验证集中,标准分析阶段VF敏感性98.3%、不可除颤心律特异性>99%;AWC阶段第二步再确认后敏感性92.1%、特异性>99%。
  • AWC对66-83%的不可除颤心律不中断胸外按压,为92.1%的VF患者最小化电击前暂停。
  • AWC需中断按压再确认的比例为34.4%;其它报告算法再确认比例为25.7-100%。

证据类型、等级与局限

  • 证据类型为算法比较性能评估研究;原文未给出证据等级或推荐强度;局限包括需再确认比例较高(34.4%),且为基于特定AED设备(DEFIGARD TOUCH7)的回顾性数据验证。

对床旁实践的改变

  • 对于不可除颤心律(占66-83%),AWC可避免中断胸外按压,可能改善CPR质量。
  • 对于VF患者,AWC可在92.1%的病例中最小化电击前暂停,但需注意34.4%的病例仍需中断按压进行再确认。
  • 适用与不适用人群原文未明确说明;常见陷阱是再确认步骤可能导致按压中断,需权衡。

指南或共识出处

  • 原文未提及

以上提炼依据论文摘要(全文未开放获取,或许可不允许翻译转载);全文请点「阅读原文」。

英文原文(摘要)· 点击展开

OBJECTIVE: The aim of this study was to present new combination of algorithms for rhythm analysis during cardiopulmonary resuscitation (CPR) in automated external defibrillators (AED), called Analyze Whilst Compressing (AWC), designed for decreasing pre-shock pause and early stopping of chest compressions (CC) for treating refibrillation.

METHODS: Two stages for AED rhythm analysis were presented, namely, "Standard Analysis Stage" (conventional shock-advisory analysis run over 5 s after CC interruption every two minutes) and "AWC Stage" (two-step sequential analysis process during CPR). AWC steps were run in presence of CC (Step1), and if shockable rhythm was detected then a reconfirmation step was run in absence of CC (Step2, analysis duration 5 s).

RESULTS: In total 16,057 ECG strips from 2916 out-of-hospital cardiac arrest (OHCA) patients treated with AEDs (DEFIGARD TOUCH7, Schiller Médical, France) were subjected patient-wise to AWC training (8559 strips, 1604 patients) and validation (7498 strips, 1312 patients). Considering validation results, "Standard Analysis Stage" presented ventricular fibrillation (VF) sensitivity Se = 98.3% and non-shockable rhythm specificity Sp>99%; "AWC Stage" decision after Step2 reconfirmation achieved Se = 92.1%, Sp>99%.

CONCLUSION: AWC presented similar performances to other AED algorithms during CPR, fulfilling performance goals recommended by standards. AWC provided advances in the challenge for improving CPR quality by: (i) not interrupting chest compressions for prevalent part of non-shockable rhythms (66-83%); (ii) minimizing pre-shock pause for 92.1% of VF patients. AWC required hands-off reconfirmation in 34.4% of cases. Reconfirmation was also common limitation of other reported algorithms (25.7-100%) although following different protocols for triggering chest compression resumption and shock delivery.