文献速递复苏、休克与脓毒症创伤与烧伤

无动脉线时床旁实时管理和滴定部分复苏性血管内球囊阻断主动脉:压力可靠,血流不可靠!

Real-time bedside management and titration of partial resuscitative endovascular balloon occlusion of the aorta without an arterial line: Good for pressure, not for flow!
PubMed 文献 · J Trauma Acute Care Surg · 2021-01-01
一句话结论猪失血性休克模型显示,COMPASS装置可无动脉线准确监测pREBOA近远端压力,但远端MAP不能可靠反映主动脉血流,基于远端MAP的滴定不可靠。

该研究在10头失血性休克猪中,使用COMPASS手持数字压力装置替代动脉线监测pREBOA近端和远端压力,并由对动脉线和主动脉血流设盲的操作者仅依据COMPASS远端MAP滴定球囊放气以目标主动脉血流0.7 L/min。结果显示COMPASS与动脉线压力相关性极强(远端r=0.979,近端r=0.989,p<0.01),但远端COMPASS MAP与主动脉血流仅弱相关(r=0.47,p<0.0001),多数操作未能达到目标血流。提示无动脉线时压力监测可靠,但以远端MAP指导血流滴定不可靠。

要点

  • 在猪失血性休克模型中,COMPASS装置测得的近端和远端MAP与动脉线压力高度相关(近端r=0.989,远端r=0.979,p<0.01)。
  • 远端COMPASS MAP与主动脉血流仅呈弱相关(r=0.47,p<0.0001),原文认为无临床意义,多数操作未达到目标主动脉血流0.7 L/min。
  • 仅依据远端MAP滴定pREBOA球囊放气以目标主动脉血流0.7 L/min,未能可靠达到目标血流。
  • 10头猪存活时间10–120分钟,平均50分钟,1头存活至120分钟。
  • 无动脉线时COMPASS可可靠监测pREBOA近远端压力,但不能替代血流监测来滴定部分血流。

适用人群与场景

  • 标准化失血性休克猪模型(10头),用于评估pREBOA置入和血流滴定方法。
  • 原文未提及人类适用人群、排除人群或具体急诊/ICU场景。

诊断标准与关键阈值

  • 目标主动脉血流设为0.7 L/min,由对动脉线压力和主动脉血流设盲的操作者仅使用COMPASS装置滴定球囊放气来实现。

处置流程与药物剂量

  • 第一步:在标准化失血性休克猪中置入pREBOA,充气球囊控制出血15分钟。
  • 第二步:放气球囊,操作者仅依据COMPASS装置读数(对动脉线压力和主动脉血流设盲)尝试达到目标主动脉血流0.7 L/min。
  • 原文未提及药物剂量、给药途径、频次或疗程。

红旗征与不能漏的情况

  • 原文未提及

研究设计与关键结果

  • 研究类型:动物实验(猪失血性休克模型),样本量10头,单中心(原文未明确中心数)。
  • 干预:使用COMPASS手持数字压力装置监测pREBOA近端和远端压力,并仅依据COMPASS远端MAP滴定球囊放气以目标主动脉血流0.7 L/min;对照:动脉线压力监测和主动脉/肠系膜上动脉血流探头。
  • 主要终点结果:远端MAP与远端COMPASS MAP强相关(r=0.979,p<0.01);近端动脉线与近端COMPASS强相关(r=0.989,p<0.01);远端COMPASS MAP与主动脉血流弱相关(r=0.47,p<0.0001),原文认为无临床意义,多数操作未达到目标血流。
  • 关键次要终点:10头猪存活时间10–120分钟,平均50分钟,1头存活至120分钟。
  • 不良事件:原文未提及。

证据类型、等级与局限

  • 证据类型:动物实验(猪模型),非RCT。
  • 证据等级或推荐强度:原文未提及。
  • 局限:样本量小(10头猪)、动物模型、单中心(原文未明确)、结果需进一步研究验证;原文指出需要进一步研究寻找合适的代理指标来靶向特定主动脉血流水平。

对床旁实践的改变

  • 在无动脉线条件下,COMPASS装置可用于pREBOA近端和远端压力监测,压力读数可靠。
  • 不应仅依据远端MAP来滴定pREBOA部分血流,因为远端MAP与主动脉血流相关性弱,多数情况下不能达到目标血流0.7 L/min。
  • 该结果来自猪失血性休克模型,能否推广至人类急诊/ICU实践尚需进一步研究;原文未提供人类适用或不适用人群。

指南或共识出处

  • 原文未提及

以上提炼依据论文摘要(全文未开放获取,或许可不允许翻译转载);全文请点「阅读原文」。

英文原文(摘要)· 点击展开

BACKGROUND: Partial resuscitative endovascular balloon occlusion of the aorta (pREBOA) attempts to minimize ischemia/reperfusion injury while controlling hemorrhage. There are little data on optimal methods to evaluate and titrate partial flow, which typically requires invasive arterial line monitoring. We sought to examine the use of a miniaturized handheld digital pressure device (COMPASS; Mirador Biomedical, Seattle, WA) for pREBOA placement and titration of flow.

METHODS: Ten swine underwent standardized hemorrhagic shock. Carotid and iliac pressures were monitored with both arterial line and COMPASS devices, and flow was monitored by aortic and superior mesenteric artery flow probes. Partial resuscitative endovascular balloon occlusion of the aorta was inflated to control hemorrhage for 15 minutes before being deflated to try targeting aortic flow of 0.7 L/min (using only the COMPASS device) by an operator blinded to the arterial line pressures and aortic flow. Correlations between COMPASS and proximal/distal arterial line were evaluated, as well as actual aortic flow.

RESULTS: There was strong correlation between the distal mean arterial pressure (MAP) and the distal COMPASS MAP (r = 0.979, p < 0.01), as well as between the proximal arterial line and the proximal COMPASS on the pREBOA (r = 0.989, p < 0.01). There was a significant but weaker correlation between the distal compass MAP reading and aortic flow (r = 0.47, p < 0.0001), although it was not clinically significant and predicted flow was not achieved in a majority of the procedures. Of 10 pigs, survival times ranged from 10 to 120 minutes, with a mean survival of 50 minutes, and 1 pig surviving to 120 minutes.

CONCLUSION: Highly reliable pressure monitoring is achieved proximally and distally without arterial lines using the COMPASS device on the pREBOA. Despite accurate readings, distal MAPs were a poor indicator of aortic flow, and titration based upon distal MAPs did not provide reliable results. Further investigation will be required to find a suitable proxy for targeting specific aortic flow levels using pREBOA.

来源:J Trauma Acute Care Surg,查看原文。版权归原作者所有。