文献速递复苏、休克与脓毒症重症综合管理与镇静镇痛

除颤器文件软件自动标注:是否已可常规应用?

Software annotation of defibrillator files: Ready for prime time?
PubMed 文献 · Resuscitation · 2020-12-31
一句话结论软件自动标注测按压频率准确,但测胸外按压分数(CCF)与人工标注差异有临床意义,不宜直接替代人工审核。

该回顾性观察研究纳入100例院外心脏骤停(OHCA)接受CPR的患者,比较软件自动标注、完全人工标注与有限人工标注(仅标记CPR开始/结束与ROSC,不标记每次按压)对胸外按压分数(CCF)和按压频率的测量。结果显示三种方法测得的病例平均按压频率无显著差异(108.1–108.6次/分),ICC>0.90;但CCF在软件标注为0.64±0.19,完全人工标注为0.86±0.07,有限人工标注为0.81±0.10,软件与人工之间差异有临床意义,而完全人工与有限人工标注一致性良好至优秀。提示软件标注可用于按压频率质控,但CCF需人工复核。

要点

  • 软件自动标注测病例平均按压频率与人工标注无显著差异(108.1–108.6次/分),ICC>0.90,可用于频率质控。
  • 软件标注测CCF为0.64±0.19,完全人工标注为0.86±0.07,两者差异有临床意义,不能直接等同。
  • 有限人工标注(仅标CPR起止与ROSC)与完全人工标注CCF相近(0.81±0.10 vs 0.86±0.07),ICC 0.69。
  • 按分钟epoch分析CCF时,完全人工与有限人工标注ICC为0.83,一致性良好至优秀。
  • 若用除颤器文件做CCF质量改进,需人工审核或至少有限人工标注校正软件结果。

适用人群与场景

  • 院外心脏骤停(OHCA)接受CPR的患者,从首次CPR开始至除颤器移除期间的生物阻抗波形。

诊断标准与关键阈值

  • 原文未提及

处置流程与药物剂量

  • 原文未提及

红旗征与不能漏的情况

  • 原文未提及

研究设计与关键结果

  • 研究类型:回顾性观察研究,单中心(原文未明确中心数),样本量100例OHCA接受CPR的患者。
  • 干预/比较:软件自动标注 vs 完全人工标注 vs 有限人工标注(仅标记CPR开始/结束与ROSC,不标记胸外按压)除颤器文件。
  • 主要终点:病例平均按压频率三种方法无显著差异(108.1–108.6次/分),ICC>0.90。
  • 主要终点:病例平均CCF软件标注0.64±0.19,完全人工标注0.86±0.07,有限人工标注0.81±0.10;软件与人工差异有临床意义。
  • 完全人工 vs 有限人工标注CCF的ICC为0.69;按分钟epoch分析ICC为0.83。

证据类型、等级与局限

  • 证据类型:回顾性观察研究(原文未给出证据等级或推荐强度)。
  • 局限:回顾性、观察性设计,样本量100例,单中心(原文未明确),软件与人工标注差异可能影响质控解读。

对床旁实践的改变

  • 除颤器文件软件自动标注可用于胸外按压频率的快速质控,但CCF测量不能直接替代人工标注,需人工审核或有限人工标注校正。
  • 在质量改进项目中,若仅关注按压频率,软件标注可节省人力;若关注CCF,建议采用有限人工标注(标记CPR起止与ROSC)以提高准确性。
  • 原文未提及适用与不适用人群的明确界限,临床决策需结合具体情况。

指南或共识出处

  • 原文未提及

以上提炼依据论文摘要(全文未开放获取,或许可不允许翻译转载);全文请点「阅读原文」。

英文原文(摘要)· 点击展开

BACKGROUND: High-quality chest compressions are associated with improved outcomes after cardiac arrest. Defibrillators record important information about chest compressions during cardiopulmonary resuscitation (CPR) and can be used in quality-improvement programs. Defibrillator review software can automatically annotate files and measure chest compression metrics. However, evidence is limited regarding the accuracy of such measurements.

OBJECTIVE: To compare chest compression fraction (CCF) and rate measurements made with software annotation vs. manual annotation vs. limited manual annotation of defibrillator files recorded during out-of-hospital cardiac arrest (OHCA) CPR.

METHODS: This was a retrospective, observational study of 100 patients who had CPR for OHCA. We assessed chest compression bioimpedance waveforms from the time of initial CPR until defibrillator removal. A reviewer revised software annotations in two ways: completely manual annotations and limited manual annotations, which marked the beginning and end of CPR and ROSC, but not chest compressions. Measurements were compared for CCF and rate using intraclass correlation coefficient (ICC) analysis.

RESULTS: Case mean rate showed no significant difference between the methods (108.1-108.6 compressions per minute) and ICC was excellent (>0.90). The case mean (±SD) CCF for software, manual, and limited manual annotation was 0.64 ± 0.19, 0.86 ± 0.07, and 0.81 ± 0.10, respectively. The ICC for manual vs. limited manual annotation of CCF was 0.69 while for individual minute epochs it was 0.83.

CONCLUSION: Software annotation performed very well for chest compression rate. For CCF, the difference between manual and software annotation measurements was clinically important, while manual vs. limited manual annotation were similar with an ICC that was good-to-excellent.

来源:Resuscitation,查看原文。版权归原作者所有。

参考来源

  1. Gupta V, Schmicker RH, Owens P, et al. Software annotation of defibrillator files: Ready for prime time? Resuscitation. 2020. PMID: 33388365. doi:10.1016/j.resuscitation.2020.12.019

供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。