文献速递心脑血管急症

磁心图对疑似急性冠脉综合征患者的诊断准确性

Diagnostic accuracy of the magnetocardiograph for patients with suspected acute coronary syndrome.
PubMed 文献 · Emerg Med J · 2020-10-13
一句话结论VitalScan磁心图目前无法准确排除急性冠脉综合征,尚不适用于临床实践。

该研究为前瞻性队列研究,纳入756例疑似急性冠脉综合征成人,分析680例(训练集293例,验证集387例),其中96例(14%)确诊急性冠脉综合征。训练集曲线下面积:磁心图0.66、改良曼彻斯特急性冠脉综合征评分0.64、两者联合0.70;验证集分别为0.56、0.69、0.64。验证集在排除阈值下磁心图敏感度0.89、特异度0.15;联合策略敏感度0.85、特异度0.30。作者结论为磁心图目前不能准确排除急性冠脉综合征,尚不适用于临床实践,需进一步开发研究。

要点

  • 验证集(n=387)中磁心图排除急性冠脉综合征阈值的敏感度0.89、特异度0.15,无法可靠排除。
  • 验证集磁心图曲线下面积仅0.56(95%CI 0.48至0.64),联合改良曼彻斯特评分也仅0.64(95%CI 0.56至0.72)。
  • 训练集磁心图联合改良曼彻斯特评分曲线下面积0.70(95%CI 0.63至0.77),但验证集降至0.64,未见稳定增益。
  • 作者结论:VitalScan磁心图目前不能准确排除急性冠脉综合征,尚不适用于临床实践,需进一步开发研究。

适用人群与场景

  • 成人疑似急性冠脉综合征患者,前瞻性队列纳入756例,最终分析680例(训练集293例,验证集387例)。

研究设计与关键结果

  • 前瞻性队列研究评估磁心图对成人疑似急性冠脉综合征的诊断准确性,参考标准由独立裁决委员会基于30天检查与事件确定。
  • 共纳入756例,分析680例(训练集293例,验证集387例),其中96例(14%)确诊急性冠脉综合征。
  • 训练集曲线下面积:磁心图0.66(95%CI 0.58至0.74)、改良曼彻斯特评分0.64(95%CI 0.54至0.73)、两者联合0.70(95%CI 0.63至0.77)。
  • 训练集磁心图在达到可接受排除敏感度(>0.98)的阈值下,特异度为0.16(95%CI 0.12至0.21)。
  • 验证集(n=387)曲线下面积:磁心图0.56(95%CI 0.48至0.64)、改良曼彻斯特评分0.69(95%CI 0.61至0.77)、两者联合0.64(95%CI 0.56至0.72)。
  • 验证集磁心图在排除阈值下敏感度0.89(95%CI 0.77至0.95)、特异度0.15(95%CI 0.12至0.20)。
  • 验证集磁心图联合改良曼彻斯特评分敏感度0.85(95%CI 0.73至0.92)、特异度0.30(95%CI 0.25至0.35)。

证据类型、等级与局限

  • 证据类型:前瞻性队列诊断准确性研究。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者结论认为VitalScan磁心图目前不能准确排除急性冠脉综合征,尚不适用于临床实践,需进一步开发研究。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: We aimed to estimate the diagnostic accuracy of the VitalScan magnetocardiograph (MCG) for suspected acute coronary syndrome (ACS).

METHODS: We undertook a prospective cohort study evaluating the diagnostic accuracy of the MCG in adults with suspected ACS. The reference standard of ACS was determined by an independent adjudication committee based on 30-day investigations and events. The cohort was split into a training sample, to derive the MCG algorithm and an algorithm combining MCG with a modified Manchester Acute Coronary Syndrome (MACS) clinical probability score, and a validation sample, to estimate diagnostic accuracy.

RESULTS: We recruited 756 participants and analysed data from 680 (293 training, 387 validation), of whom 96 (14%) had ACS. In the training sample, the respective area under the receiver operating characteristic (AUROC) curves were the following: MCG 0.66 (95% CI 0.58 to 0.74), MACS 0.64 (95% CI 0.54 to 0.73) and MCG+MACS 0.70 (95% CI 0.63 to 0.77). MCG specificity was 0.16 (95% CI 0.12 to 0.21) at the threshold achieving acceptable sensitivity for rule-out (>0.98). In the validation sample (n=387), the respective AUROCs were the following: MCG 0.56 (95% CI 0.48 to 0.64), MACS 0.69 (95% CI 0.61 to 0.77) and MCG+MACS 0.64 (95% CI 0.56 to 0.72). MCG sensitivity was 0.89 (95% CI 0.77 to 0.95) and specificity 0.15 (95% CI 0.12 to 0.20) at the rule-out threshold. MCG+MACS sensitivity was 0.85 (95% CI 0.73 to 0.92) and specificity 0.30 (95% CI 0.25 to 0.35).

CONCLUSION: The VitalScan MCG is currently unable to accurately rule out ACS and is not yet ready for use in clinical practice. Further developmental research is required.

来源:Emerg Med J,查看原文。版权归原作者所有。

参考来源

  1. Goodacre S, Walters SJ, Qayyum H, et al. Diagnostic accuracy of the magnetocardiograph for patients with suspected acute coronary syndrome. Emerg Med J. 2020. PMID: 33051274. doi:10.1136/emermed-2020-210396

供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。