巴黎大区公共自动体外除颤器记录质量不足:一项描述性研究

Insufficient quality of public automated external defibrillator recordings in the greater Paris area, a descriptive study.
PubMed 文献 · Emerg Med J · 2020-09-02
一句话结论巴黎大区院外心脏骤停中公共自动体外除颤器记录质量异质性大且不完整,约半数不适当决策因关键变量缺失无法分析。

该描述性研究收集了2017年1月1日至2019年4月31日巴黎大区8629例院外心脏骤停中部分患者使用的公共自动体外除颤器文件,最终258例患者接受公共自动体外除颤器救助,回收182份记录。记录来自12家制造商、需8种传输方式和16种软件提取,质量高度异质:2%无法解读,98%可解读记录中仅43%含胸阻抗曲线、34%含放电强度、8%含患者姓名,日期和时间准确率分别为68%和48%。公共自动体外除颤器对可电击心律的灵敏度为87.6%(95%CI 83.7%~91.0%),特异度为99.5%(99.5%~99.5%)。作者认为记录采集是重大挑战,分析受异质性和质量不足影响,需改进记录质量。

要点

  • 巴黎大区258例接受公共自动体外除颤器救助的院外心脏骤停患者中回收182份记录,来自12家制造商。
  • 98%可解读记录中仅43%含胸阻抗曲线、34%含放电强度、8%含患者姓名。
  • 记录日期和时间准确率分别为68%和48%,2%记录完全无法解读。
  • 公共自动体外除颤器对可电击心律灵敏度87.6%(95%CI 83.7%~91.0%),特异度99.5%(99.5%~99.5%)。
  • 关键变量缺失导致约半数公共自动体外除颤器不适当决策无法分析。

适用人群与场景

  • 2017年1月1日至2019年4月31日巴黎大区发生的8629例院外心脏骤停中,部分患者使用了公共自动体外除颤器,最终258例患者接受公共自动体外除颤器救助。

研究设计与关键结果

  • 研究为描述性研究,收集巴黎大区院外心脏骤停中使用的公共自动体外除颤器文件,最终258例患者接受救助,回收182份记录。
  • 记录来自12家制造商,数据提取需8种不同传输方式和16种软件程序,记录质量高度异质。
  • 2%的记录质量极差无法解读;98%可解读记录中,仅43%包含胸阻抗曲线,34%包含放电强度,8%包含患者姓名。
  • 日期和时间分别在68%和48%的记录中准确。
  • 公共自动体外除颤器对可电击心律的灵敏度为87.6%(95%CI 83.7%~91.0%),特异度为99.5%(99.5%~99.5%),阳性预测值98.2%(96.4%~99.0%),阴性预测值96.4%(95.3%~96.8%)。
  • 重要变量缺失导致约半数公共自动体外除颤器不适当决策无法分析。

证据类型、等级与局限

  • 证据类型:描述性研究。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为公共自动体外除颤器记录的采集是重大挑战,其分析受异质性和质量不足(维护记录、患者详情和日志不完整)影响。
  • 作者认为自动体外除颤器记录目前是追踪公共自动体外除颤器性能和旁观者实践的最相关资源,记录质量需要改进。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: Western countries report a significant increase in the proportion of patients who experience out-of-hospital cardiac arrests (OHCAs) and benefit from a public automated external defibrillator (pAED) before the arrival of rescue teams. However, recordings of devices recovered after resuscitation are of variable quality. Analysis of these data may inform decisions of whether to implement an internal defibrillator for survivors, and provide useful information about the performance of pAED algorithms and the actions of bystanders.

OBJECTIVE: To investigate the quality of the information recorded by pAEDs during OHCAs in the Paris area.

METHODS: pAED files used for some of the 8629 OHCAs that occurred in the greater Paris area between 1 January 2017 and 31 April 2019 on the day of the arrest were collected. The presence and accuracy of 23 factors required to interpret the recording was noted, including readability of the ECG, the presence of an impedance curve and the accuracy of the date and time. The recordings were analysed to assess the diagnostic and therapeutic performance of the pAEDs used.

RESULTS: A total of 258 patients with an OHCA received assistance from a pAED, and 182 recordings were recovered. The pAEDs were made by 12 different manufacturers. Data extraction required eight different transmission modes and 16 software programmes; recordings were of highly heterogeneous quality. Two per cent of the recordings were of such poor quality that they were not interpretable. Among the 98% remaining, only 43% included a thoracic impedance curve, 34% the intensity of the shocks delivered and 8% the patient name. The date and time were accurate in 68% and 48% of recordings, respectively. The pAEDs had 87.6% (95% CI 83.7% to 91.0%) sensitivity and 99.5% (99.5% to 99.5%) specificity for defibrillating shockable rhythms (positive predictive value 98.2% (96.4% to 99.0%), negative predictive value 96.4% (95.3% to 96.8%)). The absence of important variables prevented the analysis of approximately half of the inappropriate decisions made by pAEDs.

CONCLUSION: Collection of pAED recordings is a major challenge. Their analysis is compromised by heterogeneity and poor quality (incomplete maintenance records, patient details and logs). AED recordings are currently the most relevant resource to track pAED performance and bystander practices. The quality of these recordings needs to improve.

来源:Emerg Med J,查看原文。版权归原作者所有。

参考来源

  1. Derkenne C, Jost D, Haruel PA, et al. Insufficient quality of public automated external defibrillator recordings in the greater Paris area, a descriptive study. Emerg Med J. 2020. PMID: 32878960. doi:10.1136/emermed-2019-209371

供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。