食管球囊压力-容积曲线用于辅助机械通气时食管压校准的可行性

[Feasibility of using esophageal balloon pressure-volume curve to calibrate esophageal pressure during assisted mechanical ventilation].
PubMed 文献 · Zhonghua Wei Zhong Bing Ji Jiu Yi Xue · 2020-01-01
一句话结论辅助机械通气时,用小几何容积食管球囊(2.8 mL)和球囊压力-容积曲线校准食管压可行,简化法仅需在0.5、1.0、1.5 mL球囊容积下监测食管压即可评估食管壁弹性并校准食管压。

该前瞻性研究纳入2017年6月至2019年1月首都医科大学附属北京天坛医院重症医学科30例术后经气管插管行压力支持通气且需食管压监测、呼吸稳定的患者。用2.8 mL小几何容积食管球囊,以0.5 mL递增至2.5 mL行球囊容积试验,记录呼气末和吸气末食管压,识别压力-容积曲线中间线性段,其容积范围为球囊工作容积,斜率为食管壁弹性,呼气末与吸气末食管压差最大处球囊容积为最佳球囊容积,食管壁弹性与最佳球囊容积的乘积为食管壁回缩压,校准食管压为食管压与食管壁回缩压之差。所有患者呼气末和吸气末压力-容积曲线均存在中间线性段,工作容积下校准食管压不随球囊充气而升高,与最佳球囊容积下校准食管压一致性良好,平均差异及95%CI为-0.02(-1.50~1.50)cmH2O。简化法估算的食管壁弹性和校准食管压与标准法一致性良好,平均差异及95%CI分别为-0.2(-1.0~0.6)cmH2O/mL和0.2(-1.1~1.4)cmH2O。作者认为,辅助机械通气时用小几何容积食管球囊监测食管压并用球囊压力-容积曲线校准食管压可行,简化法可用于简化临床应用。

要点

  • 30例术后经气管插管行压力支持通气、呼吸稳定且需食管压监测的患者中,呼气末和吸气末压力-容积曲线均存在中间线性段。
  • 工作容积下校准食管压与最佳球囊容积下校准食管压一致性良好,平均差异及95%CI为-0.02(-1.50~1.50)cmH2O。
  • 简化法估算的食管壁弹性与标准法平均差异及95%CI为-0.2(-1.0~0.6)cmH2O/mL。
  • 简化法校准的食管压与标准法平均差异及95%CI为0.2(-1.1~1.4)cmH2O。
  • 作者认为,仅监测球囊容积0.5、1.0、1.5 mL时的食管压即可评估食管壁弹性并校准食管压,可用于简化临床应用。

适用人群与场景

  • 2017年6月至2019年1月首都医科大学附属北京天坛医院重症医学科收治的术后患者,经气管插管行压力支持通气、需食管压监测且呼吸稳定。

诊断标准与关键阈值

  • 以食管球囊压力-容积曲线中呼气末曲线呈线性相关的中间段为中间线性段,其容积范围为球囊工作容积,斜率为食管壁弹性。
  • 呼气末与吸气末食管压差最大的球囊容积为最佳球囊容积。
  • 食管壁弹性与最佳球囊容积的乘积为食管壁回缩压,校准食管压为食管压与食管壁回缩压之差。

处置流程与药物剂量

  • 用几何容积2.8 mL的食管球囊行食管压监测,以0.5 mL递增间歇充气至2.5 mL行球囊容积试验,记录呼气末和吸气末食管压,获得呼气末和吸气末压力-容积曲线。
  • 识别呼气末压力-容积曲线的中间线性段,确定球囊工作容积、食管壁弹性和最佳球囊容积,计算食管壁回缩压并校准食管压。
  • 简化法:基于患者球囊工作容积均位于0.5~1.5 mL,以0.5 mL与1.5 mL球囊容积下呼气末食管压之差除以1.0 mL估算食管壁弹性,并用该弹性校准0.5~1.5 mL范围内的食管压。

研究设计与关键结果

  • 前瞻性研究纳入30例术后经气管插管行压力支持通气、呼吸稳定且需食管压监测的患者,所有患者呼气末和吸气末压力-容积曲线均存在中间线性段。
  • 工作容积下校准食管压不随球囊充气而升高,与最佳球囊容积下校准食管压一致性良好,平均差异及95%CI为-0.02(-1.50~1.50)cmH2O(1 cmH2O=0.098 kPa)。
  • 简化法估算的食管壁弹性与标准法平均差异及95%CI为-0.2(-1.0~0.6)cmH2O/mL,简化法校准的食管压与标准法平均差异及95%CI为0.2(-1.1~1.4)cmH2O。

证据类型、等级与局限

  • 证据类型:前瞻性研究。
  • 作者结论:辅助机械通气时用小几何容积食管球囊监测食管压并用球囊压力-容积曲线校准食管压可行,简化法可用于简化临床应用。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为,辅助机械通气时可用小几何容积食管球囊监测食管压,并用球囊压力-容积曲线校准食管压。
  • 作者认为,简化法仅需监测球囊容积0.5、1.0、1.5 mL时的食管压即可评估食管壁弹性并校准食管压,可用于简化临床应用。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

OBJECTIVE: To evaluate the feasibility of esophageal pressure (Pes) calibration by the esophageal balloon pressure-volume (P-V) curve during assisted mechanical ventilation.

METHODS: A prospective study was conducted. The postoperative patients admitted to intensive care unit of Beijing Tiantan Hospital Affiliated to Capital Medical University from June 2017 to January 2019 who needed pressure support ventilation by tracheal intubation and Pes monitoring with stable breath were enrolled. The Pes monitoring was performed by the esophageal balloon with a small geometric volume (2.8 mL). (1) Balloon volume tests of esophageal balloon were performed by inflating intermittently 0.5 mL increments up to 2.5 mL, the end-expiratory and end-inspiratory Pes were recorded to obtain end-expiratory and end-inspiratory P-V curves. The intermediate section in end-expiatory P-V curve that showed linear correlation was identified (as intermediate linear section), whose volume range was balloon working volume (Vwork) and slope was esophageal wall elastance (Ees), the balloon volume with the largest difference between end-expiratory and end-inspiratory Pes was the best balloon volume (Vbest), and the product of Ees and Vbest was esophageal wall recoil pressure reacting to balloon filling. To minimize the effect of esophageal wall on Pes, the calibrated Pes was the difference of Pes and esophageal wall recoil pressure. The consistency of calibrated Pes obtained by balloon volume at Vbest and other Vwork were analyzed. (2) For the convenience of clinical application, a simplified method was introduced to calibrate Pes. Based on all Vwork of patients located in 0.5-1.5 mL, the difference of end-expiratory Pes between balloon volume at 0.5 mL and 1.5 mL divided by 1.0 mL was used to estimate Ees, and the Pes among 0.5-1.5 mL was calibrated by Ees obtained by the simple method. The consistency of calibrated Pes obtained by the simple method and standard method were observed.

RESULTS: Totally 30 patients were enrolled, all end-expiratory and end-inspiratory P-V curves existed the intermediate linear section, the calibrated Pes at Vwork did not increase with the balloon being inflated and had a good consistency with the calibrated Pes at Vbest, mean difference and 95% confidence interval (95%CI) was -0.02 (-1.50-1.50) cmH2O (1 cmH2O = 0.098 kPa). The Ees and calibrated Pes estimated by the simple method had a good agreement with the standard method, mean difference and 95%CI was -0.2 (-1.0-0.6) cmH2O/mL and 0.2 (-1.1-1.4) cmH2O, respectively.

CONCLUSIONS: During assisted mechanical ventilation, the use of a small geometric volume esophageal balloon to monitor Pes and balloon P-V curve to calibrate Pes is feasible. The simple method can be used for simplifying clinical application, that's only by monitoring Pes at balloon volume at 0.5, 1.0 and 1.5 mL to evaluate the Ees and calibrate Pes.

来源:Zhonghua Wei Zhong Bing Ji Jiu Yi Xue,查看原文。版权归原作者所有。

参考来源

  1. Sun X, Chen G, Yang Y, et al. [Feasibility of using esophageal balloon pressure-volume curve to calibrate esophageal pressure during assisted mechanical ventilation]. Zhonghua Wei Zhong Bing Ji Jiu Yi Xue. 2020. PMID: 32788014. doi:10.3760/cma.j.cn121430-20200413-00282

供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。