最小漏气法测定人工气道套囊压力的临床研究

[A clinical study on the determination of cuff pressure in artificial airway by minimum air leakage method].
PubMed 文献 · Zhonghua Wei Zhong Bing Ji Jiu Yi Xue · 2020-01-01
一句话结论最小漏气法较套囊压力表法能使气管插管套囊压力更高、呼吸机漏气量更少,并降低呼吸机相关性肺炎发生率(4% vs 10%)。

该前瞻性随机对照研究纳入2015年12月至2019年6月天津市第五中心医院重症医学科100例气管插管患者,随机分为最小漏气法组(实验组,50例)和套囊压力表法组(对照组,50例,套囊压力维持25-30 cmH2O)。结果显示实验组在充气0、4、8小时的套囊压力均显著高于对照组,漏气量更小;实验组呼吸机相关性肺炎发生率显著低于对照组(4% vs 10%,P<0.05),其他气道并发症发生率两组无统计学差异。作者认为最小漏气法使套囊压力更准确、漏气更少、安全有效,值得临床推广。

要点

  • 最小漏气法组充气0、4、8小时套囊压力均高于压力表法组(33.72±9.14 vs 25.68±5.26 mmHg、30.54±7.81 vs 24.35±4.93 mmHg、26.57±5.64 vs 22.42±4.14 mmHg,均P<0.05)。
  • 最小漏气法组呼吸机漏气量小于压力表法组(0小时25.57±8.51 vs 34.65±9.47 mL,4小时40.54±8.51 vs 60.34±7.85 mL,均P<0.05)。
  • 最小漏气法组呼吸机相关性肺炎发生率显著低于压力表法组(4% vs 10%,P<0.05)。
  • 两组气道黏膜水肿、溃疡、气管食管瘘、声嘶、咳嗽、咽痛、气管软化、套囊破裂发生率无统计学差异(均P>0.05)。
  • 两组套囊压力均随充气时间延长而下降,呼吸机漏气量随时间延长而增加。

适用人群与场景

  • 2015年12月至2019年6月天津市第五中心医院重症医学科收治的气管插管患者100例,男53例、女47例,年龄23-87岁,平均(68.53±8.46)岁,插管时间1-16天。
  • 所有患者插管成功后取平卧位、床头抬高30度。

处置流程与药物剂量

  • 实验组采用最小漏气法,并用套囊压力表获取套囊压力。
  • 对照组套囊压力维持于25-30 cmH2O(1 cmH2O=0.098 kPa)。
  • 分别于充气开始时及充气后4小时、8小时比较两组套囊压力和呼吸机漏气量

研究设计与关键结果

  • 前瞻性随机对照研究,纳入100例气管插管患者,随机分为最小漏气法组(实验组,50例)和套囊压力表法组(对照组,50例)。
  • 充气后4小时和8小时两组套囊压力均低于首次充气时,呼吸机漏气量随时间延长逐渐增加。
  • 实验组各时间点套囊压力均显著高于对照组(mmHg:0小时33.72±9.14 vs 25.68±5.26,4小时30.54±7.81 vs 24.35±4.93,8小时26.57±5.64 vs 22.42±4.14,均P<0.05)。
  • 实验组呼吸机漏气量小于对照组(mL:0小时25.57±8.51 vs 34.65±9.47,4小时40.54±8.51 vs 60.34±7.85,均P<0.05)。
  • 实验组呼吸机相关性肺炎发生率显著低于对照组(4% vs 10%,P<0.05)。
  • 两组其他气道并发症发生率无统计学差异(气道黏膜水肿14% vs 12%,溃疡8% vs 6%,气管食管瘘0% vs 0%,声嘶4% vs 6%,咳嗽30% vs 34%,咽痛28% vs 32%,气管软化0% vs 0%,套囊破裂10% vs 8%,均P>0.05)。

证据类型、等级与局限

  • 证据类型:前瞻性随机对照研究。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为最小漏气法使临床获得的气管插管套囊压力更准确,漏气更少,安全有效,值得临床推广。
  • 作者认为最佳套囊压力对预防呼吸机相关性肺炎和减少气道并发症非常重要。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

OBJECTIVE: To compare the cuff pressure and leakage volume and the related complications of filling the tracheal tube cuff by minimum air leakage method and cuff pressure manometer method after endotracheal intubation, so as to provide theoretical basis for patients who was intubated to obtain appropriate cuff pressure.

METHODS: A prospective randomized controlled study was conducted. 100 patients admitted to the department of critical care medicine of the Fifth Center Hospital in Tianjin from December 2015 to June 2019 were enrolled. According to the random number table method, the patients were divided into the experimental group and control group, with 50 patients in each group. After successful endotracheal intubation, all patients were placed in a supine position with the head of the bed raised by 30 degree angle. The experimental group used the minimum air leakage method, and used the cuff pressure manometer to obtain the cuff pressure. In the control group, cuff pressure was maintained at 25-30 cmH2O (1 cmH2O = 0.098 kPa). Parameters such as cuff pressure and ventilator leakage volume at the beginning and 4 hours, 8 hours after the inflation were compared between the two groups, as well as the incidence of ventilation-associated pneumonia (VAP) and airway complications after extubation.

RESULTS: Among the 100 cases, 53 were males and 47 were females. The age ranged from 23 to 87 years old, with an average of (68.53±8.46) years old. The intubation time ranged from 1 to 16 days. (1) At 4 hours and 8 hours after inflation, the cuff pressures of the two groups were lower than that of the first time of inflation, and the air leakage of the ventilator increased gradually with the extension of time. Compared with the control group, cuff pressures at each time point in the experimental group were significantly higher than those in the control group [mmHg (1 mmHg = 0.133 kPa): 33.72±9.14 vs. 25.68±5.26 at 0 hour, 30.54±7.81 vs. 24.35±4.93 at 4 hours, 26.57±5.64 vs. 22.42±4.14 at 8 hours, all P < 0.05], and ventilator leakage volumes were smaller than those in the control group (mL: 25.57±8.51 vs. 34.65±9.47 at 0 hour, 40.54±8.51 vs. 60.34±7.85 at 4 hours, both P < 0.05). (2) The incidence of VAP in the experimental group was significantly lower than that in the control group (4% vs. 10%, P < 0.05). There was no statistically significant difference in the incidence of other airway complications between the experimental group and control group (airway mucosal edema: 14% vs. 12%, ulcer: 8% vs. 6%, tracheal esophageal fistula: 0% vs. 0%, hoarseness: 4% vs. 6%, cough: 30% vs. 34%, sore throat: 28% vs. 32%, tracheal softening: 0% vs. 0%, cuff rupture: 10% vs. 8%, all P > 0.05).

CONCLUSIONS: The optimal cuff pressure is very important for preventing VAP and reducing airway complications. The minimum air leakage method makes the clinical obtained endotracheal intubation cuff pressure more accurately, with less air leakage, safe and effective, and it is worthy of clinical promotion.

来源:Zhonghua Wei Zhong Bing Ji Jiu Yi Xue,查看原文。版权归原作者所有。