声门上气道装置在尸体心肺复苏中产生负胸腔内压的能力不一:一项前瞻性交叉研究

Supraglottic airway devices variably develop negative intrathoracic pressures: A prospective cross-over study of cardiopulmonary resuscitation in human cadavers.
PubMed 文献 · Resuscitation · 2020-01-18
一句话结论人体尸体模型显示,不同声门上气道装置在复苏减压期产生负胸腔内压的能力存在差异,且需阻抗阈值装置才能显著产生负压。

该前瞻性交叉研究在七具近期死亡、体型各异的人体尸体上,比较气管插管与5种声门上气道装置在常规手动、自动及主动加压-减压心肺复苏中(含或不含阻抗阈值装置、仰卧或头高位)的气道压(作为胸腔内压的替代指标)。主要结局为各组减压期气道压的比较。结果显示,所有心肺复苏方法中均需阻抗阈值装置才能产生显著更低的负胸腔内压;不同声门上气道装置支持负胸腔内压的能力存在差异。作者认为,在优化心脏骤停结局时应强烈考虑声门上气道装置的差异,因为部分装置在多种心肺复苏方法和装置下不能稳定形成密封或负胸腔内压。

要点

  • 在人体尸体模型中,所有心肺复苏方法(常规手动、自动、主动加压-减压)均需阻抗阈值装置才能产生显著更低的负胸腔内压。
  • 5种声门上气道装置支持负胸腔内压的能力存在差异,部分装置在多种心肺复苏方法和装置下不能稳定形成密封或负胸腔内压。
  • 气管插管与声门上气道装置的气道压比较在七具近期死亡、体型各异的人体尸体上进行,装置顺序随机。
  • 作者认为,在优化心脏骤停结局时应强烈考虑声门上气道装置的差异。

适用人群与场景

  • 七具近期死亡、体型各异的人体尸体,用于前瞻性交叉设计研究。

研究设计与关键结果

  • 研究类型:前瞻性交叉设计,在七具近期死亡的人体尸体上进行;干预:气管插管与5种声门上气道装置(随机顺序),在仰卧和头高位下,分别用常规手动、自动及主动加压-减压心肺复苏,含或不含阻抗阈值装置;主要结局:各组减压期气道压比较。
  • 结果:所有心肺复苏方法中均需阻抗阈值装置才能产生显著更低的负胸腔内压;不同声门上气道装置支持负胸腔内压的能力存在差异。

证据类型、等级与局限

  • 证据类型:前瞻性交叉研究,人体尸体模型。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为,在优化心脏骤停结局时应强烈考虑声门上气道装置的差异,因为部分声门上气道装置在多种心肺复苏方法和装置下不能稳定形成密封或负胸腔内压。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

AIM OF THE STUDY: Negative intrathoracic pressure (ITP) during the decompression phase of cardiopulmonary resuscitation (CPR) is essential to refill the heart, increase cardiac output, maintain cerebral and coronary perfusion pressures, and improve survival. In order to generate negative ITP, an airway seal is necessary. We tested the hypothesis that some supraglottic airway (SGA) devices do not seal the airway as well the standard endotracheal tube (ETT).

METHODS: Airway pressures (AP) were measured as a surrogate for ITP in seven recently deceased human cadavers of varying body habitus. Conventional manual, automated, and active compression-decompression CPR were performed with and without an impedance threshold device (ITD) in supine and Head Up positions. Positive pressure ventilation was delivered by an ETT and 5 SGA devices tested in a randomized order in this prospective cross-over designed study. The primary outcome was comparisons of decompression AP between all groups.

RESULTS: An ITD was required to generate significantly lower negative ITP during the decompression phase of all methods of CPR. SGAs varied in their ability to support negative ITP.

CONCLUSION: In a human cadaver model, the ability to generate negative intrathoracic pressures varied with different SGAs and an ITD regardless of the body position or CPR method. Differences in SGAs devices should be strongly considered when trying to optimize cardiac arrest outcomes, as some SGAs do not consistently develop a seal or negative intrathoracic pressure with multiple different CPR methods and devices.

来源:Resuscitation,查看原文。版权归原作者所有。