TEG 6s与TEG 5000分析仪在创伤患者凝血评估中的比较

A comparison between the TEG 6s and TEG 5000 analyzers to assess coagulation in trauma patients.
PubMed 文献 · J Trauma Acute Care Surg · 2020-01-01
一句话结论多中心观察性研究显示,TEG 6s在创伤患者中与TEG 5000高度可靠相关,且设备内重复性相当或更优。

本研究为在美国12个一级和二级创伤中心开展的多中心非侵入性观察性方法比较研究,纳入475例符合全额或有限创伤团队标准的成人创伤患者,入院后1小时内采血,比较TEG 6s与TEG 5000在CK.R、CFF.MA、CK.LY30、CRT.MA和CK.MA参数上的一致性。结果显示各参数回归斜率接近1、线性相关系数高(ρ=0.86–0.95),提示两设备间可靠性强;TEG 6s的设备内重复性整体优于TEG 5000,尤其在CFF.MA和CK.LY30上。作者认为TEG 6s在创伤患者中高度可靠,与TEG 5000密切 related,设备内重复性等效或更优。

要点

  • 在475例成人创伤患者中,TEG 6s与TEG 5000各参数线性相关強(ρ=0.86–0.95),回归斜率接近1。
  • CFF.MA的相关系数最高(ρ=0.95,ß=0.99),CK.LY30为ρ=0.91、ß=1.01,CK.R为ρ=0.9、ß=1.05。
  • TEG 6s的CRT.MA与TEG 5000的CK.MA比较:ß=1.06、ρ=0.86;与TEG 5000的CRT.MA比较:ß=0.93、ρ=0.93。
  • TEG 6s的设备内重复性整体优于TEG 5000,尤其在CFF.MA和CK.LY30参数上。
  • 作者认为TEG 6s在创伤患者中高度可靠,鉴于其可在救治现场使用的潜在优势,两设备间一致性的确认是诊断检测的重要进展。

适用人群与场景

  • 成人创伤患者,符合全额或有限创伤团队标准,来自美国12个一级和二级创伤中心。
  • 入院后1小时内采血,用于方法比较。

研究设计与关键结果

  • 多中心非侵入性观察性方法比较研究,纳入475例成人创伤患者,男性68.6%,中位年龄49岁。
  • TEG 6s与TEG 5000参数比较:CK.R ß=1.05、ρ=0.9;CFF.MA ß=0.99、ρ=0.95;CK.LY30 ß=1.01、ρ=0.91;TEG 6s的CRT.MA vs TEG 5000的CK.MA ß=1.06、ρ=0.86;TEG 6s的CRT.MA vs TEG 5000的CRT.MA ß=0.93、ρ=0.93。
  • 设备内重复性整体TEG 6s优于TEG 5000,尤其在CFF.MA和CK.LY30参数上。

证据类型、等级与局限

  • 诊断试验,证据等级II级。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为TEG 6s在创伤患者中高度可靠,与TEG 5000密切 related,设备内重复性等效或更优。
  • 作者认为鉴于TEG 6s在救治现场使用的潜在优势,确认两设备间一致性是诊断检测的重要进展。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: Trauma-induced coagulopathy is a major driver of mortality following severe injury. Viscoelastic goal-directed resuscitation can reduce mortality after injury. The TEG 5000 system is widely used for viscoelastic testing. However, the TEG 6s system incorporates newer technology, with encouraging results in cardiovascular interventions. The purpose of this study was to validate the TEG 6s system for use in trauma patients.

METHODS: Multicenter noninvasive observational study for method comparison conducted at 12 US Levels I and II trauma centers. Agreement between the TEG 6s and TEG 5000 systems was examined using citrated kaolin reaction time (CK.R), citrated functional fibrinogen maximum amplitude (CFF.MA), citrated kaolin percent clot lysis at 30 minutes (CK.LY30), citrated RapidTEG maximum amplitude (CRT.MA), and citrated kaolin maximum amplitude (CK.MA) parameters in adults meeting full or limited trauma team criteria. Blood was drawn ≤1 hour after admission. Assays were repeated in duplicate. Reliability (TEG 5000 vs. TEG 6s analyzers) and repeatability (interdevice comparison) was quantified. Linear regression was used to define the relationship between TEG 6s and TEG 5000 devices.

RESULTS: A total of 475 patients were enrolled. The cohort was predominantly male (68.6%) with a median age of 49 years. Regression line slope estimates (ß) and linear correlation estimates (p) were as follows: CK.R (ß = 1.05, ρ = 0.9), CFF.MA (ß = 0.99, ρ = 0.95), CK.LY30 (ß = 1.01, ρ = 0.91), CRT.MA (TEG 6s) versus CK.MA (TEG 5000) (ß = 1.06, ρ = 0.86) as well as versus CRT.MA (TEG 5000) (ß = 0.93, ρ = 0.93), indicating strong reliability between the devices. Overall, within-device repeatability was better for TEG 6s versus TEG 5000, particularly for CFF.MA and CK.LY30.

CONCLUSION: The TEG 6s device appears to be highly reliable for use in trauma patients, with close correlation to the TEG 5000 device and equivalent/improved within-device reliability. Given the potential advantages of using the TEG 6s device at the site of care, confirmation of agreement between the devices represents an important advance in diagnostic testing.

LEVEL OF EVIDENCE: Diagnostic test, level II.

来源:J Trauma Acute Care Surg,查看原文。版权归原作者所有。

参考来源

  1. Neal MD, Moore EE, Walsh M, et al. A comparison between the TEG 6s and TEG 5000 analyzers to assess coagulation in trauma patients. J Trauma Acute Care Surg. 2020. PMID: 31738314. doi:10.1097/TA.0000000000002545

供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。