文献速递重症综合管理与镇静镇痛复苏、休克与脓毒症

连续无创动脉压监测评估机械通气危重患者血流动力学状态的临床研究

[A clinical study of the evaluation of hemodynamic status in mechanically ventilated critically ill patients by continuous non-invasive arterial pressure monitor].
PubMed 文献 · Zhonghua Wei Zhong Bing Ji Jiu Yi Xue · 2019-01-01
一句话结论在机械通气危重患者中,连续无创动脉压监测(CNAP)与脉搏指示连续心排血量监测(PiCCO)在测量心指数和脉压变异率上具有良好一致性,但平均动脉压和体循环血管阻力指数的一致性较差,解读需谨慎。

该前瞻性观察自身对照研究纳入2018年6月至2019年3月福建省立医院重症医学科收治的38例需要血流动力学监测的机械通气危重患者,最终分析32例。入ICU后即刻置入PiCCO导管,于0小时和24小时分别用热稀释法测得平均动脉压(MAPPiCCO)、心指数(CIPiCCO)、脉压变异率(PPVPiCCO)和体循环血管阻力指数(SVRIPiCCO),同时用CNAP测得相应指标(MAPCNAP、CICNAP、PPVCNAP、SVRICNAP),所有测量重复三次取平均值。结果显示,CNAP与PiCCO测得的CI和PPV无显著差异且一致性良好,但CNAP测得的MAP和SVRI显著高于PiCCO,一致性差。结论认为CNAP监测机械通气危重患者的CI和PPV与PiCCO可比,而MAP和SVRI结果可能不准确,应正确谨慎解读。

要点

  • 在机械通气危重患者中,CNAP与PiCCO测得的CI和PPV无显著差异,且Bland-Altman分析显示一致性良好,平均差异分别为1.2 mL×s-1×m-2和0.5%。
  • CNAP测得的MAP和SVRI显著高于PiCCO(MAP 65.6±9.4 mmHg vs 60.1±9.2 mmHg,SVRI 206.2±53.9 kPa×s×L-1×m-2 vs 179.5±57.8 kPa×s×L-1×m-2,均P<0.01),一致性差。
  • CNAP与PiCCO测得的MAP、CI、PPV和SVRI均呈显著正相关(r值分别为0.624、0.864、0.835、0.655,均P<0.05)。
  • CNAP监测CI和PPV与PiCCO可比,但MAP和SVRI结果可能不准确,临床解读需谨慎。

适用人群与场景

  • 纳入标准:2018年6月至2019年3月福建省立医院重症医学科第四科室收治的机械通气且需要血流动力学监测的危重患者。
  • 排除标准:24小时内死亡、24小时内放弃治疗、心房颤动、因技术问题数据丢失的患者。

研究设计与关键结果

  • 研究类型:前瞻性观察自身对照研究;最终分析32例,其中女性12例、男性20例,年龄26~84岁,平均(66.8±19.1)岁,BMI(23.7±3.9)kg/m2,APACHE II评分19.5±5.3,SOFA评分9.7±4.1。
  • 干预与对照:入ICU后即刻置入PiCCO导管,于0小时和24小时用热稀释法测量MAPPiCCO、CIPiCCO、PPVPiCCO、SVRIPiCCO,同时用CNAP测量MAPCNAP、CICNAP、PPVCNAP、SVRICNAP,所有测量重复三次取平均值。
  • 主要结果:CNAP与PiCCO测得的CI和PPV无显著差异[CI(mL×s-1×m-2):59.8±12.6 vs 58.5±14.2;PPV:(14.7±6.8)% vs(14.0±6.8)%,均P>0.05];CNAP测得的MAP和SVRI显著高于PiCCO[MAP(mmHg):65.6±9.4 vs 60.1±9.2;SVRI(kPa×s×L-1×m-2):206.2±53.9 vs 179.5±57.8,均P<0.01]。
  • 相关性分析:CNAP与PiCCO测得的MAP、CI、PPV和SVRI均呈显著正相关(r值分别为0.624、0.864、0.835、0.655,均P<0.05)。
  • 一致性分析:Bland-Altman分析显示CNAP与PiCCO测量CI和PPV的一致性良好,平均差异分别为1.2 mL×s-1×m-2和0.5%,95%CI分别为-12.8~15.3 mL×s-1×m-2和-7.1%~8.2%;而MAP和SVRI的一致性差,平均差异分别为5.5 mmHg和26.8 kPa×s×L-1×m-2,95%CI分别为-10.4~21.3 mmHg和-64.5~118.0 kPa×s×L-1×m-2。

证据类型、等级与局限

  • 证据类型:前瞻性观察自身对照研究。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 在机械通气危重患者中,CNAP监测的CI和PPV与PiCCO可比,可用于无创监测CI和PPV。
  • CNAP监测的MAP和SVRI结果可能不准确,应正确谨慎解读。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

OBJECTIVE: To evaluate the difference and correlation between continuous non-invasive arterial pressure (CNAP) monitor and pulse indicated continuous cardiac output (PiCCO) monitor on determination of hemodynamic parameters in mechanically ventilated critically ill patients, and to assess the feasibility of non-invasive monitoring of hemodynamics with CNAP.

METHODS: A prospective observation self-control study was conducted.The critically ill patients with mechanical ventilation who needed hemodynamics monitoring, and admitted to the fourth department of intensive care unit (ICU) of Fujian Provincial Hospital from June 2018 to March 2019 were enrolled. PiCCO catheter were inserted immediately after admission, the hemodynamic indexes were measured by thermodilution method, and mean arterial pressure (MAPPiCCO), cardiac index (CIPiCCO), pulse pressure variation rate (PPVPiCCO) and systemic vascular resistance index (SVRIPiCCO) were obtained at 0 hour and 24 hours respectively. Meanwhile, the above indexes (MAPCNAP, CICNAP, PPVCNAP and SVRICNAP) were measured with CNAP. All measurements were repeated thrice and average values were reported. The differences in above parameters between the two methods were evaluated. Pearson test was used for the correlation analysis and Bland-Altman analysis method was used for consistency test.

RESULTS: Thirty-eight patients were enrolled into this study. One patient died within 24 hours was excluded, 2 patients were excluded due to withdrawing treatment within 24 hours, 2 patients were excluded because of atrial fibrillation, and 1 patient's data was lost due to technical problems. Thus, data from 32 patients were available for final analysis. There were 12 females and 20 males, aging 26-84 years old with the mean of (66.8±19.1) years old, body mass index (BMI) of (23.7±3.9) kg/m2, acute physiology and chronic health evaluation II (APACHE II) score of 19.5±5.3, sepsis-related organ failure assessment (SOFA) score of 9.7±4.1. There were no significant differences in CI or PPV between CNAP and PiCCO groups [CI (mL×s-1×m-2): 59.8±12.6 vs. 58.5±14.2, PPV: (14.7±6.8)% vs. (14.0±6.8)%, both P > 0.05]. MAP and SVRI measured by CNAP were significantly higher than those measured by PiCCO [MAP (mmHg, 1 mmHg = 0.133 kPa): 65.6±9.4 vs. 60.1±9.2, SVRI (kPa×s×L-1×m-2): 206.2±53.9 vs. 179.5±57.8, both P < 0.01]. The correlation analysis showed that MAP, CI, PPV and SVRI measured by the two methods were significantly positively correlated (r value was 0.624, 0.864, 0.835 and 0.655 respectively, all P < 0.05). Bland-Altman analysis showed that CNAP and PiCCO had a good consistency for the measurement of CI and PPV, the average differences were 1.2 mL×s-1×m-2 and 0.5% respectively, while the 95% confidence interval (95%CI) were -12.8-15.3 mL×s-1×m-2 and -7.1%-8.2% respectively. However, the consistency of MAP and SVRI measured by those two methods was poor, the average differences were 5.5 mmHg and 26.8 kPa×s×L-1×m-2 respectively, while the 95%CI was -10.4-21.3 mmHg and -64.5-118.0 kPa×s×L-1×m-2 respectively.

CONCLUSIONS: CNAP was comparable with PiCCO when monitoring CI and PPV in mechanically ventilated critically ill patients; while the results of MAP and SVRI might be inaccurate, which should be interpreted correctly and carefully.

来源:Zhonghua Wei Zhong Bing Ji Jiu Yi Xue,查看原文。版权归原作者所有。

参考来源

  1. Xue Y, Wu W, Chen M, et al. [A clinical study of the evaluation of hemodynamic status in mechanically ventilated critically ill patients by continuous non-invasive arterial pressure monitor]. Zhonghua Wei Zhong Bing Ji Jiu Yi Xue. 2019. PMID: 31771720. doi:10.3760/cma.j.issn.2095-4352.2019.10.010